A study on dosimetric properties of electronic portal imaging device and its use as a quality assurance tool in Volumetric Modulated Arc Therapy.

Prabakar Sukumar1, Sriram Padmanaban1, Prakash Jeevanandam1, S A Syam Kumar1, Vivekanandan Nagarajan1.   

Abstract

AIM: In this study, the dosimetric properties of the electronic portal imaging device were examined and the quality assurance testing of Volumetric Modulated Arc Therapy was performed.
BACKGROUND: RapidArc involves the variable dose rate, leaf speed and the gantry rotation. The imager was studied for the effects like dose, dose rate, field size, leaf speed and sag during gantry rotation.
MATERIALS AND METHODS: A Varian RapidArc machine equipped with 120 multileaf collimator and amorphous silicon detector was used for the study. The characteristics that are variable in RapidArc treatment were studied for the portal imager. The accuracy of a dynamic multileaf collimator position at different gantry angles and during gantry rotation was examined using the picket fence test. The control of the dose rate and gantry speed was verified using a test field irradiating seven strips of the same dose with different dose rate and gantry speeds. The control over leaf speed during arc was verified by irradiating four strips of different leaf speeds with the same dose in each strip. To verify the results, the RapidArc test procedure was compared with the X-Omat film and verified for a period of 6 weeks using EPID.
RESULTS: The effect of gantry rotation on leaf accuracy was minimal. The dose in segments showed good agreement with mean deviation of 0.8% for dose rate control and 1.09% for leaf speed control over different gantry speeds.
CONCLUSION: The results provided a precise control of gantry speed, dose rate and leaf speeds during RapidArc delivery and were consistent over 6 weeks.

Entities:  

Keywords:  Dose rate; EPID; Leaf speed; Picket fence; Quality assurance; RapidArc

Year:  2011        PMID: 24376989      PMCID: PMC3863247          DOI: 10.1016/j.rpor.2011.08.001

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  12 in total

1.  Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device.

Authors:  K L Pasma; M L Dirkx; M Kroonwijk; A G Visser; B J Heijmen
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy.

Authors:  Nesrin Dogan; Leonid B Leybovich; Anil Sethi
Journal:  Phys Med Biol       Date:  2002-11-21       Impact factor: 3.609

3.  Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.

Authors:  Peter B Greer; Carmen C Popescu
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

4.  Evaluation of film and thermoluminescent dosimetry of high-energy electron beams in heterogeneous phantoms.

Authors:  E el-Khatib; J Antolak; J Scrimger
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

5.  Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.

Authors:  L N McDermott; M Wendling; B van Asselen; J Stroom; J J Sonke; M van Herk; B J Mijnheer
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

6.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

7.  Commissioning and quality assurance of RapidArc radiotherapy delivery system.

Authors:  C Clifton Ling; Pengpeng Zhang; Yves Archambault; Jiri Bocanek; Grace Tang; Thomas Losasso
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-10-01       Impact factor: 7.038

8.  Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT.

Authors:  B M C McCurdy; P B Greer
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

9.  Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device.

Authors:  Peter B Greer
Journal:  Med Phys       Date:  2007-10       Impact factor: 4.071

10.  Using the frame averaging of aS500 EPID for IMRT verification.

Authors:  J Chang; C C Ling
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

View more
  7 in total

1.  The importance of accurate treatment planning, delivery, and dose verification.

Authors:  Julian Malicki
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

2.  Exit fluence analysis using portal dosimetry in volumetric modulated arc therapy.

Authors:  Prabakar Sukumar; Sriram Padmanaban; Dhanabalan Rajasekaran; Muniyappan Kannan; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2012-07-15

3.  Comparison of individual and composite field analysis using array detector for Intensity Modulated Radiotherapy dose verification.

Authors:  Sathiyan Saminathan; Varatharaj Chandraraj; C H Sridhar; Ravikumar Manickam
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-19

4.  A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy.

Authors:  Dhanabalan Rajasekaran; Prakash Jeevanandam; Prabakar Sukumar; Arulpandiyan Ranganathan; Samdevakumar Johnjothi; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2014-09-06

5.  Commissioning and Acceptance Testing of the existing linear accelerator upgraded to volumetric modulated arc therapy.

Authors:  Ekambaram Varadharajan; Velayudham Ramasubramanian
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-06

6.  A study on conventional IMRT and RapidArc treatment planning techniques for head and neck cancers.

Authors:  S A Syam Kumar; Nagarajan Vivekanandan; Padmanaban Sriram
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-03

7.  Evaluation of application of EPID for rapid QC testing of linear accelerator.

Authors:  Shahla Ebrahimi Moghadam; Shahrokh Nasseri; Seyedeh Somayeh Seyedi; Hamid Gholamhosseinian; Mehdi Momennezhad
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-13
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.