Comparison of blood pressure monitoring by applanation tonometry and invasively assessed blood pressure in cardiological patients.

G Greiwe1, S Hoffmann2, L Herich3, M S Winkler2, C J Trepte2, C R Behem2, M Petzoldt2, D A Reuter4, S A Haas4.   

Abstract

The aim of this study was to evaluate the accuracy and precision of non-invasive continuous blood pressure measurement by applanation tonometry (AT) in awake or anaesthetised cardiological intensive care patients. Patients suffering from highly impaired left ventricular function atrial fibrillation or severe aortic valve stenosis were included into the study. Arterial blood pressure was recorded by applanation tonometry (T-Line 400, Tensys Medical®, USA) and an arterial line in awake or anaesthetised patients. Discrepancies in mean (MAP), systolic (SAP), and diastolic (DAP) arterial pressure between the two methods were assessed as bias, limits of agreement and percentage error respectively. In 31 patients a total of 27,900 measurements were analyzed. The concordance correlation coefficient was 0.23, 0.45 and 0.06 for MAP, SAP and DAP, respectively. For all patients bias for MAPAT compared to MAPAL was 14.96 mmHg (SAPAT 4.51 mmHg; DAPAT 19.12 mmHg) with limits of agreement for MAPAT of 46.25 and - 16.33 mm Hg (SAPAT 48.00 and - 38.98 mmHg; DAPAT 50.12 and - 11.89 mmHg). Percentage error for MAPAT was 56.8% (42.7% for SAPAT; 75.2% for DAPAT). We conclude that the AT method is not reliable in ICU patients with severe cardiac comorbidities.

Entities:  

Keywords:  Applanation tonometry; Blood pressure; Cardiological patients; Haemodynamic monitoring

Mesh:

Year:  2017        PMID: 29204771     DOI: 10.1007/s10877-017-0089-9

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  20 in total

1.  Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry.

Authors:  Ron Dueck; Oliver Goedje; Paul Clopton
Journal:  J Clin Monit Comput       Date:  2012-01-18       Impact factor: 2.502

2.  The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients.

Authors:  Bernd Saugel; Florian Fassio; Alexander Hapfelmeier; Agnes S Meidert; Roland M Schmid; Wolfgang Huber
Journal:  Intensive Care Med       Date:  2012-06-29       Impact factor: 17.440

3.  Using the Bland-Altman method to measure agreement with repeated measures.

Authors:  P S Myles; J Cui
Journal:  Br J Anaesth       Date:  2007-09       Impact factor: 9.166

4.  The concordance correlation coefficient for repeated measures estimated by variance components.

Authors:  Josep L Carrasco; Tonya S King; Vernon M Chinchilli
Journal:  J Biopharm Stat       Date:  2009       Impact factor: 1.051

5.  Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure.

Authors:  A S Meidert; W Huber; J N Müller; M Schöfthaler; A Hapfelmeier; N Langwieser; J Y Wagner; F Eyer; R M Schmid; B Saugel
Journal:  Br J Anaesth       Date:  2013-12-18       Impact factor: 9.166

6.  Radial artery cannulation in 1,000 patients: precautions and complications.

Authors:  M A Mandel; P J Dauchot
Journal:  J Hand Surg Am       Date:  1977-11       Impact factor: 2.230

7.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 9.  Radial artery cannulation and ischaemic damage: a review.

Authors:  R G Wilkins
Journal:  Anaesthesia       Date:  1985-09       Impact factor: 6.955

10.  Is applanation tonometry a reliable method for monitoring blood pressure in morbidly obese patients undergoing bariatric surgery?

Authors:  G Greiwe; P A Tariparast; C Behem; M Petzoldt; L Herich; C J Trepte; D A Reuter; S A Haas
Journal:  Br J Anaesth       Date:  2016-04-19       Impact factor: 9.166

View more
  3 in total

Review 1.  Journal of clinical monitoring and computing end of year summary 2018: hemodynamic monitoring and management.

Authors:  Bernd Saugel; Moritz Flick; Karim Bendjelid; Lester A H Critchley; Simon T Vistisen; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2019-03-07       Impact factor: 2.502

2.  Clinical Validation of a Soft Wireless Continuous Blood Pressure Sensor During Surgery.

Authors:  En-Fan Chou; Shin Yu Celia Cheung; Hailey Christine Maxwell; Nicholas Pham; Michelle Khine; Joseph Rinehart
Journal:  Front Digit Health       Date:  2021-07-22

Review 3.  Blood pressure monitoring techniques in the natural state of multi-scenes: A review.

Authors:  Ziyi Liu; Congcong Zhou; Hongwei Wang; Yong He
Journal:  Front Med (Lausanne)       Date:  2022-08-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.