CT dose and image quality in the last three scanner generations.

Andreas Christe1, Johannes Heverhagen1, Christoph Ozdoba1, Christian Weisstanner1, Stefan Ulzheimer1, Lukas Ebner1.   

Abstract

AIM: To compare the computed tomography (CT) dose and image quality with the filtered back projection against the iterative reconstruction and CT with a minimal electronic noise detector.
METHODS: A lung phantom (Chest Phantom N1 by Kyoto Kagaku) was scanned with 3 different CT scanners: the Somatom Sensation, the Definition Flash and the Definition Edge (all from Siemens, Erlangen, Germany). The scan parameters were identical to the Siemens presetting for THORAX ROUTINE (scan length 35 cm and FOV 33 cm). Nine different exposition levels were examined (reference mAs/peek voltage): 100/120, 100/100, 100/80, 50/120, 50/100, 50/80, 25/120, 25/100 and 25 mAs/80 kVp. Images from the SOMATOM Sensation were reconstructed using classic filtered back projection. Iterative reconstruction (SAFIRE, level 3) was performed for the two other scanners. A Stellar detector was used with the Somatom Definition Edge. The CT doses were represented by the dose length products (DLPs) (mGycm) provided by the scanners. Signal, contrast, noise and subjective image quality were recorded by two different radiologists with 10 and 3 years of experience in chest CT radiology. To determine the average dose reduction between two scanners, the integral of the dose difference was calculated from the lowest to the highest noise level.
RESULTS: When using iterative reconstruction (IR) instead of filtered back projection (FBP), the average dose reduction was 30%, 52% and 80% for bone, soft tissue and air, respectively, for the same image quality (P < 0.0001). The recently introduced Stellar detector (Sd) lowered the radiation dose by an additional 27%, 54% and 70% for bone, soft tissue and air, respectively (P < 0.0001). The benefit of dose reduction was larger at lower dose levels. With the same radiation dose, an average of 34% (22%-37%) and 25% (13%-46%) more contrast to noise was achieved by changing from FBP to IR and from IR to Sd, respectively. For the same contrast to noise level, an average of 59% (46%-71%) and 51% (38%-68%) dose reduction was produced for IR and Sd, respectively. For the same subjective image quality, the dose could be reduced by 25% (2%-42%) and 44% (33%-54%) using IR and Sd, respectively.
CONCLUSION: This study showed an average dose reduction between 27% and 70% for the new Stellar detector, which is equivalent to using IR instead of FBP.

Keywords:  Computed tomography detector; Computed tomography image quality; Computed tomography signal to noise; Dose reduction; Image noise; Low dose computed tomography

Year:  2013        PMID: 24349646      PMCID: PMC3856334          DOI: 10.4329/wjr.v5.i11.421

Source DB:  PubMed          Journal:  World J Radiol        ISSN: 1949-8470


  32 in total

1.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Authors:  M Gies; W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  CT patterns of fungal pulmonary infections of the lung: comparison of standard-dose and simulated low-dose CT.

Authors:  Andreas Christe; Margaret C Lin; Andrew C Yen; Rich L Hallett; Kingshuk Roychoudhury; Florian Schmitzberger; Dominik Fleischmann; Ann N Leung; Geoffrey D Rubin; Geoffry D Rubin; Peter Vock; Justus E Roos
Journal:  Eur J Radiol       Date:  2011-08-10       Impact factor: 3.528

3.  Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation.

Authors:  Tom H Mulkens; Patrick Bellinck; Michel Baeyaert; Dirk Ghysen; Xavier Van Dijck; Elvier Mussen; Caroline Venstermans; Jean-Luc Termote
Journal:  Radiology       Date:  2005-08-26       Impact factor: 11.105

Review 4.  Principles of CT: radiation dose and image quality.

Authors:  Lee W Goldman
Journal:  J Nucl Med Technol       Date:  2007-11-15

5.  The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments.

Authors:  K D Hopper; S H King; M E Lobell; T R TenHave; J S Weaver
Journal:  Radiology       Date:  1997-12       Impact factor: 11.105

6.  Radioprotection to the eye during CT scanning.

Authors:  K D Hopper; J D Neuman; S H King; A R Kunselman
Journal:  AJNR Am J Neuroradiol       Date:  2001 Jun-Jul       Impact factor: 3.825

7.  Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements.

Authors:  W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

Review 8.  Techniques and applications of automatic tube current modulation for CT.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Bernhard Schmidt; Bryan L Westerman; Hugh T Morgan; Sanjay Saini
Journal:  Radiology       Date:  2004-10-21       Impact factor: 11.105

Review 9.  ATM, ATR and DNA-PK: initiators of the cellular genotoxic stress responses.

Authors:  Jun Yang; Yingnian Yu; Hope E Hamrick; Penelope J Duerksen-Hughes
Journal:  Carcinogenesis       Date:  2003-08-14       Impact factor: 4.944

10.  Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose.

Authors:  Sebastian T Schindera; Patricia Graca; Michael A Patak; Susanne Abderhalden; Gabriel von Allmen; Peter Vock; Zsolt Szucs-Farkas
Journal:  Invest Radiol       Date:  2009-10       Impact factor: 6.016

View more
  9 in total

Review 1.  Advanced imaging techniques for chronic pancreatitis.

Authors:  Anushri Parakh; Temel Tirkes
Journal:  Abdom Radiol (NY)       Date:  2020-05

2.  A Practice Quality Improvement Project: Reducing Dose of Routine Chest CT Imaging in a Busy Clinical Practice.

Authors:  Edwin A Takahashi; Marc D Kohli; Shawn D Teague
Journal:  J Digit Imaging       Date:  2016-10       Impact factor: 4.056

3.  Motion preservation in type II odontoid fractures using temporary pedicle screw fixation: a preliminary study.

Authors:  Bin Han; Fangcai Li; Gang Chen; Hao Li; Qixin Chen
Journal:  Eur Spine J       Date:  2014-12-28       Impact factor: 3.134

4.  Low-dose biplanar radiography can be used in children and adolescents to accurately assess femoral and tibial torsion and greatly reduce irradiation.

Authors:  Olivier Meyrignac; Ramiro Moreno; Christiane Baunin; Julie Vial; Franck Accadbled; Agnes Sommet; Jérome Sales de Gauzy; Nicolas Sans
Journal:  Eur Radiol       Date:  2014-12-23       Impact factor: 5.315

5.  Preoperative planning of calcium deposit removal in calcifying tendinitis of the rotator cuff - possible contribution of computed tomography, ultrasound and conventional X-Ray.

Authors:  Kaywan Izadpanah; Martin Jaeger; Dirk Maier; Norbert P Südkamp; Peter Ogon
Journal:  BMC Musculoskelet Disord       Date:  2014-11-20       Impact factor: 2.362

6.  CARE Dose 4D combined with sinogram-affirmed iterative reconstruction improved the image quality and reduced the radiation dose in low dose CT of the small intestine.

Authors:  Lin Wang; Shenchu Gong; Jushun Yang; Jie Zhou; Jing Xiao; Jin-Hua Gu; Hong Yang; Jianfeng Zhu; Bosheng He
Journal:  J Appl Clin Med Phys       Date:  2018-12-03       Impact factor: 2.102

7.  Impact of ROI Size on the Accuracy of Noise Measurement in CT on Computational and ACR Phantoms.

Authors:  Choirul Anam; Pandji Triadyaksa; Ariij Naufal; Zaenal Arifin; Zaenul Muhlisin; Evi Setiawati; Wahyu Setia Budi
Journal:  J Biomed Phys Eng       Date:  2022-08-01

8.  Feasible Dose Reduction in Routine Chest Computed Tomography Maintaining Constant Image Quality Using the Last Three Scanner Generations: From Filtered Back Projection to Sinogram-affirmed Iterative Reconstruction and Impact of the Novel Fully Integrated Detector Design Minimizing Electronic Noise.

Authors:  Lukas Ebner; Felix Knobloch; Adrian Huber; Julia Landau; Daniel Ott; Johannes T Heverhagen; Andreas Christe
Journal:  J Clin Imaging Sci       Date:  2014-07-31

9.  Influence of exposure parameters and iterative reconstruction on automatic airway segmentation and analysis on MDCT-An ex vivo phantom study.

Authors:  Patricia Leutz-Schmidt; Oliver Weinheimer; Bertram J Jobst; Julien Dinkel; Jürgen Biederer; Hans-Ulrich Kauczor; Michael U Puderbach; Mark O Wielpütz
Journal:  PLoS One       Date:  2017-08-02       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.