Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose.

Sebastian T Schindera1, Patricia Graca, Michael A Patak, Susanne Abderhalden, Gabriel von Allmen, Peter Vock, Zsolt Szucs-Farkas.   

Abstract

OBJECTIVE: To compare image quality and radiation dose of thoracoabdominal computed tomography (CT) angiography at 80 and 100 kVp and to assess the feasibility of reducing contrast medium volume from 60 to 45 mL at 80 kVp.
MATERIALS AND METHODS: This retrospective study had institutional review board approval; informed consent was waived. Seventy-five patients who had undergone thoracoabdominal 64-section multidetector-row CT angiography were divided into 3 groups of 25 patients each. Patients of groups A (tube voltage, 100 kVp) and B (tube voltage, 80 kVp) received 60 mL of contrast medium at 4 mL/s. Patients of group C (tube voltage, 80 kVp) received 45 mL of contrast medium at 3 mL/s. Mean aortoiliac attenuation, image noise, and contrast-to-noise ratio were assessed. The measurement of radiation dose was based on the volume CT dose index. Three independent readers assessed the diagnostic image quality.
RESULTS: Mean aortoiliac attenuation for group B (621.1 +/- 90.5 HU) was significantly greater than for groups A and C (485.2 +/- 110.5 HU and 483.1 +/- 119.8 HU; respectively) (P < 0.001). Mean image noise was significantly higher for groups B and C than for group A (P < 0.05). The contrast-to-noise ratio did not significantly differ between the groups (group A, 35.0 +/- 13.8; group B, 31.7 +/- 10.1; group C, 27.3 +/- 11.5; P = 0.08). Mean volume CT dose index in groups B and C (5.2 +/- 0.4 mGy and 4.9 +/- 0.3 mGy, respectively) were reduced by 23.5% and 27.9%, respectively, compared with group A (6.8 +/- 0.8 mGy) (P < 0.001). The average overall diagnostic image quality for the 3 groups was graded as good or better. The score for group A was significantly higher than that for group C (P < 0.01), no difference was seen between group A and B (P = 0.92).
CONCLUSIONS: Reduction of tube voltage from 100 to 80 kVp for thoracoabdominal CT angiography significantly reduces radiation dose without compromising image quality. Reduction of contrast medium volume to 45 mL at 80 kVp resulted in lower but still diagnostically acceptable image quality.

Entities:  

Mesh:

Year:  2009        PMID: 19724236     DOI: 10.1097/RLI.0b013e3181acaf8a

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  23 in total

1.  Aortic CT angiography dose reduction: investigation of optimal noise index and iterative algorithm strength in combination with low kV.

Authors:  Osvaldo Rampado; Stefania Busso; Domenica Garabello; Ezio Marengo; Marco Valerio; Simona Capello; Simona Veglia; Ottavio Davini; Roberto Ropolo
Journal:  Radiol Med       Date:  2015-12-16       Impact factor: 3.469

Review 2.  Surveillance Imaging Following Endovascular Aneurysm Repair.

Authors:  Nirnimesh Pandey; Harold I Litt
Journal:  Semin Intervent Radiol       Date:  2015-09       Impact factor: 1.513

3.  CT dose and image quality in the last three scanner generations.

Authors:  Andreas Christe; Johannes Heverhagen; Christoph Ozdoba; Christian Weisstanner; Stefan Ulzheimer; Lukas Ebner
Journal:  World J Radiol       Date:  2013-11-28

4.  Low-tube-voltage (80 kVp) CT aortography using 320-row volume CT with adaptive iterative reconstruction: lower contrast medium and radiation dose.

Authors:  Chien-Ming Chen; Sung-Yu Chu; Ming-Yi Hsu; Ying-Lan Liao; Hui-Yu Tsai
Journal:  Eur Radiol       Date:  2013-10-02       Impact factor: 5.315

Review 5.  Complications of endovascular aneurysm repair of the thoracic and abdominal aorta: evaluation and management.

Authors:  Dania Daye; T Gregory Walker
Journal:  Cardiovasc Diagn Ther       Date:  2018-04

6.  An education and training programme for radiological institutes: impact on the reduction of the CT radiation dose.

Authors:  Sebastian T Schindera; Reto Treier; Gabriel von Allmen; Claude Nauer; Philipp R Trueb; Peter Vock; Zsolt Szucs-Farkas
Journal:  Eur Radiol       Date:  2011-05-31       Impact factor: 5.315

7.  Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose.

Authors:  Eun-Suk Cho; Tae-Sub Chung; Sung Jun Ahn; KyoungHoon Chong; Jang Hun Baek; Sang Hyun Suh
Journal:  Eur Radiol       Date:  2014-12-16       Impact factor: 5.315

8.  128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose.

Authors:  Vincenzo Russo; Monica Garattoni; Francesco Buia; Domenico Attinà; Luigi Lovato; Maurizio Zompatori
Journal:  Eur Radiol       Date:  2015-05-24       Impact factor: 5.315

9.  Low kV settings CT angiography (CTA) with low dose contrast medium volume protocol in the assessment of thoracic and abdominal aorta disease: a feasibility study.

Authors:  D Ippolito; C Talei Franzesi; D Fior; P A Bonaffini; O Minutolo; S Sironi
Journal:  Br J Radiol       Date:  2015-03-18       Impact factor: 3.039

10.  Automated attenuation-based tube voltage selection for body CTA: Performance evaluation of 192-slice dual-source CT.

Authors:  Anna Winklehner; Sonja Gordic; Eliane Lauk; Thomas Frauenfelder; Sebastian Leschka; Hatem Alkadhi; Daniela B Husarik
Journal:  Eur Radiol       Date:  2015-02-19       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.