OBJECTIVE: This study compared the primary stability of two commercially available acetabular components from the same manufacturer, which differ only in geometry; a hemispherical and a peripherally enhanced design (peripheral self-locking (PSL)). The objective was to determine whether altered geometry resulted in better primary stability. METHODS: Acetabular components were seated with 0.8 mm to 2 mm interference fits in reamed polyethylene bone substrate of two different densities (0.22 g/cm(3) and 0.45 g/cm(3)). The primary stability of each component design was investigated by measuring the peak failure load during uniaxial pull-out and tangential lever-out tests. RESULTS: There was no statistically significant difference in seating force (p = 0.104) or primary stability (pull-out p = 0.171, lever-out p = 0.087) of the two components in the low-density substrate. Similarly, in the high-density substrate, there was no statistically significant difference in the peak pull-out force (p = 0.154) or lever-out moment (p = 0.574) between the designs. However, the PSL component required a significantly higher seating force than the hemispherical cup in the high-density bone analogue (p = 0.006). CONCLUSIONS: Higher seating forces associated with the PSL design may result in inadequate seating and increased risk of component malpositioning or acetabular fracture in the intra-operative setting in high-density bone stock. Our results, if translated clinically, suggest that a purely hemispherical geometry may have an advantage over a peripherally enhanced geometry in high density bone stock. Cite this article: Bone Joint Res 2013;2:264-9.
OBJECTIVE: This study compared the primary stability of two commercially available acetabular components from the same manufacturer, which differ only in geometry; a hemispherical and a peripherally enhanced design (peripheral self-locking (PSL)). The objective was to determine whether altered geometry resulted in better primary stability. METHODS: Acetabular components were seated with 0.8 mm to 2 mm interference fits in reamed polyethylene bone substrate of two different densities (0.22 g/cm(3) and 0.45 g/cm(3)). The primary stability of each component design was investigated by measuring the peak failure load during uniaxial pull-out and tangential lever-out tests. RESULTS: There was no statistically significant difference in seating force (p = 0.104) or primary stability (pull-out p = 0.171, lever-out p = 0.087) of the two components in the low-density substrate. Similarly, in the high-density substrate, there was no statistically significant difference in the peak pull-out force (p = 0.154) or lever-out moment (p = 0.574) between the designs. However, the PSL component required a significantly higher seating force than the hemispherical cup in the high-density bone analogue (p = 0.006). CONCLUSIONS: Higher seating forces associated with the PSL design may result in inadequate seating and increased risk of component malpositioning or acetabular fracture in the intra-operative setting in high-density bone stock. Our results, if translated clinically, suggest that a purely hemispherical geometry may have an advantage over a peripherally enhanced geometry in high density bone stock. Cite this article: Bone Joint Res 2013;2:264-9.
To compare the primary stability of two commercially available
acetabular components produced by the same manufacturer, differing
only in geometryIt was hypothesised that there would be no differences in primary
stability achieved by the two designs in a controlled bench-top
modelAlthough the primary stability achieved by both cups was found
to be similar, the force required to seat the PSL design in high
density substrate was 25% higher than that of the purely hemispherical
componentClinically, higher seating forces are harder to generate and
increase the risk of inadequate component seating, malpositioning
and acetabular fractures, which may compromise the outcome of the
operationIf our results translate clinically, the hemispherical geometry
is superior in higher density bone as it achieves the same initial
stability but requires a lower insertional forceStrengths: comparison was made between components that only varied
in geometry using a controlled reproducible test set-upLimitations: the use of a dry synthetic bone substrate rather
than real bone and not testing torsion stability or combined mechanisms
Introduction
Uncemented and hybrid total hip replacements (THRs) have become
increasingly popular over the last decade. Data from the National
Joint Registry shows that the use of uncemented THRs overtook cemented
arthroplasties in England and Wales for the first time in 2009.[1] In 2011, 32.6% of
primary hip replacements were cemented, 44.7% uncemented and 17.4%
hybrids.[1]Uncemented components require excellent primary stability as
micromotion at the bone–prosthesis interface inhibits osseointegration
and impairs long-term component stability.[2-5] Initial
primary stability is achieved by the principle of press-fit, where
an oversized component is inserted into an undersized reamed cavity.
The most common geometric design used to achieve the press-fit effect has
been the hemispherical component. However, clinical concerns have
arisen over the generation of excessive peri-prosthetic strains,
which may be associated with subsequent fractures, and polar gaps,
which may serve as route for wear particles or obviate bony ingrowth.[6-9] Consequently, other acetabular geometries
have been introduced. One of the most widely available is the peripherally enhanced
design, which in essence comprises hemispherical geometry with a
wider diameter at the peripheral edge. In contrast to the purely
hemispherical geometry, the peripherally enhanced design is typically
inserted into a cavity reamed to the same diameter as the component’s nominal
dimension (‘line-to-line’), meaning that the overall press-fit is
provided entirely by the enhanced periphery.Although several studies have evaluated the primary stability
of different acetabular components, these studies have typically
compared designs that vary not only in geometry, but also in material
properties, wall thickness, and surface finish.[10-15] Consequently, any differences in primary
stability seen in these studies cannot be attributed solely to differences
in the geometry of the components.The aim of the current study was to compare the primary stability
of two commercially-available acetabular components produced by
the same manufacturer that differ only in their geometry. It was
hypothesised that there would be no differences in the primary stability
achieved by the two geometries in a controlled bench-top model.
Materials and Methods
Acetabular components
We evaluated hemispherical and peripheral self-locking (PSL)
acetabular component designs of the Trident hip system (Stryker,
Newbury, United Kingdom). The PSL component was identical to the
hemispherical design, except for a 1.8 mm enhanced peripheral structure.
Both components had a core diameter of 52 mm, were manufactured
from the same titanium alloy and had the same hydroxyapatite-treated surface.
One of each design was used to test each density (0.22 g/cm3 and
0.45 g/cm3) and each
mode of failure (pull-out and lever-out), necessitating a total
of four of each design.
Synthetic bone substrate
Polyethylene foam (Pedilin Foam; Otto Bock, Bock, Austria) was
used as a synthetic bone substrate. Synthetic bone substrate has
been shown to lower interspecimen variability compared with post-mortem material,
while providing a similar compressive modulus to that of cadaveric
bone.[10,16,17] Cylindrical blocks (radius 51 mm
and height 50 mm) were prepared in densities of 0.22 g/cm3 and
0.45 g/cm3 in an effort to model low- and high-density
bone stock, respectively.[10] The
mean compressive modulus for low- and high-density substrate blocks (77 MPa
(sd 12)) and 211 MPa (sd 8), respectively) encompassed
the range of values reported for cancellous bone stock (approximately
50 MPa to 450 MPa).[18]Test blocks were subsequently mounted on a lathe and reamed 0.5
mm beyond the equator of the acetabular reamer to a depth of 26.5
mm. A standard depth gauge (resolution 0.1 mm) was used to verify
the depth of the cavity. Reaming parameters for each acetabular
geometry in low and high density substrate are shown in Table I. Consistent
with previous studies,[10,19] lower press-fit
values were used to ensure full seating of the acetabular components
in the high density substrate. Our own pilot work also demonstrated
that forces in excess of 10 kN were required to insert the PSL component
in the high-density substrate when press-fits of 2 mm and 1.8 mm were
initially used.[20] Such
magnitude of force was above the maximum rated capacity of our materials
testing machine and would be challenging and potentially damaging
to achieve manually in the intra-operative setting. Therefore the
chosen press-fit values were modified as described above.Reaming parameters used for hemispherical
and peripheral self-locking (PSL) acetabular component designs in
low- (0.22 g/cm3) and high-density (0.45 g/cm3)
substrate
Seating
Acetabular components were inserted into the reamed test blocks
using a custom introducing assembly, which provided a uniformly
distributed load to each acetabular component through the original
manufacturer’s component-to-inserter locking mechanism (Fig. 1).
A uniaxial materials testing machine (Instron 5800R, Instron, High
Wycombe, United Kingdom), fitted with a 10 kN load cell (accuracy,
0.1% full scale), was used to seat the two acetabular designs at a linear rate of 2.5 mm/s.
Acetabular components were considered to be seated when: 1) the
displacement of the component was equivalent to the measured depth
of the reamed cavity; and 2) the entire hydroxyapatite-coated area
of the component was engaged within the bone substrate, such that
only the inner sleeve of the acetabular component was visible.Photograph showing the peripheral self-locking
(PSL) Trident acetabular component (Stryker, Newbury, United Kingdom) mounted
on the brass rod connecting assembly used to secure the component
to the load cell.
Primary stability
Primary stability of the two geometries was evaluated using two
modes of failure: direct distraction (pull-out tests) and sagittal/coronal
rotation (lever-out tests). Ten repetitions were performed for each
mode of failure, in both high- and low-density bone substrate. One acetabular
component was used for each set of ten tests. Therefore a total
of 80 tests of primary stability were conducted, 40 each for the
hemispherical and PSL designs, using four hemispherical and four
PSL components.
Pull-out test
Following seating, the uniaxial materials testing machine was
used to distract acetabular components directly from the reamed
cavity at a rate of 2 mm/s.
Lever-out test
Following seating, the introducing assembly was disconnected
from the load cell and reconnected to a universal attachment via
a steel cable and adjustable pulley system. The effective moment
arm for the lever-out test was defined as the measured distance
from the universal joint to the substrate surface (Fig. 2). Components
were subsequently distracted at a rate of 10 mm/s.[10]Diagrammatic illustration of the moment
arm determined for ‘lever-out’ testing. A steel cable, low friction
pulley and universal connection are used to apply a lever-out moment
to the introducing assembly. The moment arm was defined by the distance of
the universal joint from the surface of the synthetic block.
Data reduction and statistical analysis
Force and displacement data were sampled at 100 Hz during seating and
failure testing. Seating force was defined as the peak force generated
during component insertion, while pull-out force was defined as
the maximum force during direct distraction. The peak lever-out
moment was calculated by multiplying the peak lever-out force by
the respective moment arm.The SPSS statistical package v16.0 for Windows (SPSS Inc., Chicago,
Illinois) was used for all statistical procedures. Underlying assumptions
of normality and equality of variance were assessed using the Kolmogorov–Smirnoff
and Levene’s tests, respectively. As press-fit parameters differed between
high- and low-density bone substrate, separate analyses were conducted for each. Between-component
differences in peak seating force, pull-out force and lever-out moment
for each test were evaluated using independent sample t-tests.
An alpha level of 0.05 was used for all tests of significance (significance
at p < 0.05).
Results
Seating forces
There was no statistically significant difference in the mean
seating force between the hemispherical and PSL components in the
low-density substrate (4078 N (2384 to 6743) versus 4649
N (3166 to 6850); p = 0.104). In high density substrate, however,
the mean seating force for the PSL geometry was 25% higher compared
with the hemispherical design (7858 N (4752 to 13 756) versus 6264
N (3897 to 9200)) (Fig. 3). This difference was statistically significant
(p = 0.006).Seating force for hemispherical and
peripheral self-locking (PSL) implant designs in low- (0.22 g/cm3)
and high-density (0.45 g/cm3) substrate. The error bars
denote the range of values (* p = 0.006).The primary stability of the two designs is shown in Table II.
There were no statistically significant differences in mean pull-out
force or lever-out moment between the two geometries in either low-
or high-density substrate (all p ≥ 0.087).Primary stability of hemispherical
and peripheral self-locking (PSL) acetabular component designs in
low- (0.22 g/cm3) and high-density (0.45 g/cm3)
substrate* independent samples t-test
Discussion
The primary stability of uncemented acetabular components is
paramount for biological fixation. While several factors have been
suggested to influence this initial stability,[11,21] this is the first study, to our
knowledge, to directly compare the primary stability of two commercially
available acetabular component designs that differ only in geometry.
Consistent with our initial hypothesis, we observed that the geometry
of the acetabular component had minimal effect on primary stability during pull-out and lever-out
testing in both high- and low-density substrates, provided that
satisfactory seating had been achieved. However, contrary to our
original hypothesis, we also observed that the peripherally enhanced
geometry of the PSL component required higher seating forces – approximately
25% higher than those required by the hemispherical design. Thus,
in our controlled bench-top model, acetabular geometry appears to
have a profound effect on the force required to seat components.
This is similar to the senior author’s (MS) intra-operative observations.From a surgical point of view, it is important that acetabular
components can be inserted accurately and without the use of excessive
force. The higher force required to seat the PSL component in high-density bone
may be difficult to achieve, may lead to incomplete seating and/or
and component malpositioning and also may increase the risk of peri-operative
fractures. This finding is of particular importance, given that
many orthopaedic surgeons prefer to use uncemented arthroplasty
in younger patients who are more likely to have higher bone density.
Unsatisfactory seating of the acetabular component leads to incomplete
engagement and therefore small residual compressive forces that
may result in excessive micromotion and adversely affect primary
stability.[2,4,5] Malpositioning, including lateralisation
of the hip’s centre of rotation due to incomplete seating, can lead
to impingement and decreased range of movement,[22-24] a higher risk of dislocation,[25-27] and increased rates of wear.[28] Clinical data,
however, are paramount. We have highlighted that there are potential difficulties
in seating the PSL component in dense bone. This has been observed
clinically, with Nunag et al[29] reporting
20 of a series of 30 Trident PSL acetabular components (Stryker)
having incomplete seating. However, this series showed good clinical outcomes at
five years,[29] and
the Trident acetabular implant has good medium-term registry data,[1] although these data
do not distinguish between the two variants of the cup.A limitation of our study is that we quantified the primary stability
of the two acetabular geometries using discrete pull-out and lever-out
mechanisms in a dry synthetic bone substrate. Although the synthetic
material served as a representative model for evaluating the effects
of varying bone quality, we recognise that the substrate is unlikely
to have the same visco-elastic response as bone, despite possessing
a similar compressive modulus. The use of a synthetic substrate,
however, gives a consistent set of material properties for all testing,
allowing a more accurate comparison of the primary stability of
the different component geometries and it has been widely used in
previous studies.[10,13,14] While we also recognise that torsion
stability or combined mechanisms may arguably be more clinically
relevant predictors of stability,[10,13,30,31] the magnitude of pull-out forces
in our series (between 0.7 kN and 1.6 kN) is comparable to the range
of failure values reported in the literature (0.7 kN to 2 kN).[13,32] Similarly, the magnitude of the
lever-out moments resulting in failure in our current study (between
14 Nm and 40 Nm) is within the wide range reported in the literature
(5 Nm to 50 Nm).[10,12-14,32,33] Due to the availability
of implants, each acetabular component was re-used ten times, with a
new component for each density of substrate and mode of failure.
It was possible that repeated use could cause deformation of the
components and therefore changes in the subsequent seating forces
and stability. However, we observed no obvious deformation and sequential
analysis of the results did not show any trends (either increasing
or decreasing) with regards to seating force, pull-out force or
lever-out moment.The exact force needed to seat a component in an appropriately
reamed acetabular cavity in vivo is currently unknown.
Therefore, the seating forces generated in this study can only be
compared with corresponding values estimated from other in
vitro studies. Of the few studies that have quantified
seating forces with similar press-fit values (0 mm to 2 mm) and substrate
densities (0.2 g/cm3 and 0.5 g/cm3), four have
reported a range between 1.5 kN and 5 kN with a variety of uncemented
implants that included purely hemispherical as well as peripherally
enhanced designs.[10,11,15,30] The
range of seating forces generated in the current study (between
4 kN and 8 kN) is somewhat higher. It should be noted however that
while the cited studies have employed similar substrates (polyethylene
foam and polyurethane foam) and comparable press-fits, certain implant
designs were reported to be incompletely seated.[10] In comparison,
satisfactory seating was achieved in all the trials of this study
and may, in part, explain the relatively high forces required to seat
components. It should be also noted that although our method of
seating implants does not replicate the method of insertion used
in surgery (in which a series of impulsive blows are applied) it
is a controlled method of insertion that allows for accurate measurement
of the forces involved; it therefore enabled us to compare this important
parameter between the cup designs in a standardised fashion.
Conclusions
Although the primary stability achieved by both designs was found
to be similar, as assessed by pull-out and lever-out failure, the
force required to seat the PSL design in high density substrate
was 25% higher than that of the purely hemispherical implant. Clinically,
higher seating forces are harder to generate and increase the risk of
inadequate component seating, malpositioning and acetabular fractures,
which may compromise the outcome of the operation. If our results
translate clinically, the hemispherical geometry is superior in
higher density bone as it achieves the same initial stability but
requires a lower insertional force.
Table I
Reaming parameters used for hemispherical
and peripheral self-locking (PSL) acetabular component designs in
low- (0.22 g/cm3) and high-density (0.45 g/cm3)
substrate
Hemispherical (52 mm)
PSL (52 mm)
Low density (0.22 g/cm3)
Reaming diameter (mm)
50
52
Nominal interference fit (mm)
2
1.8
Reaming depth (mm)
26.5
26.5
High density (0.45 g/cm3)
Reaming diameter (mm)
51
53
Nominal interference fit (mm)
1
0.8
Reaming depth (mm)
26.5
26.5
Table II
Primary stability of hemispherical
and peripheral self-locking (PSL) acetabular component designs in
low- (0.22 g/cm3) and high-density (0.45 g/cm3)
substrate
Authors: Shantanu Patil; Arnie Bergula; Peter C Chen; Clifford W Colwell; Darryl D D'Lima Journal: J Bone Joint Surg Am Date: 2003 Impact factor: 5.284