The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices.

Minoo Mahshid1, Aboulfazl Saboury, Seyed Jalil Sadr, Ali Fayyaz, Mahdi Kadkhodazadeh.   

Abstract

PURPOSE: This study aimed to assess the combined effect of dismantling before sterilization and aging on the accuracy (±10% of the target torque) of spring-style mechanical torque devices (S-S MTDs).
METHODS: Twenty new S-SMTDs from two different manufacturers (Nobel Biocare and Straumann: 10 of each type) were selected and divided into two groups, namely, case (group A) and control (group B). For sterilization, 100 cycles of autoclaving were performed in 100 sequences. In each sequence, 10 repetitions of peak torque values were registered for aging. To measure and assess the output of each device, a Tohnichi torque gauge was used (P<0.05).
RESULTS: Before steam sterilization, all of the tested devices stayed within 10% of their target values. After 100 cycles of steam sterilization and aging with or without dismantling of the devices, the Nobel Biocare devices stayed within 10% of their target torque. In the Straumann devices, despite the significant difference between the peak torque and target torque values, the absolute error values stayed within 10% of their target torque.
CONCLUSION: Within the limitations of this study, there was no significant difference between the mean and absolute value of error between Nobel Biocare and Straumann S-S MTDs.

Entities:  

Keywords:  Dental implants; Safety; Sterilization; Torque

Year:  2013        PMID: 24236244      PMCID: PMC3825989          DOI: 10.5051/jpis.2013.43.5.221

Source DB:  PubMed          Journal:  J Periodontal Implant Sci        ISSN: 2093-2278            Impact factor:   2.614


  15 in total

1.  Accuracy of mechanical torque-limiting devices for implants.

Authors:  Jon P Standlee; Angelo A Caputo; Ming-ya J Chwu; Tao T Sun
Journal:  Int J Oral Maxillofac Implants       Date:  2002 Mar-Apr       Impact factor: 2.804

2.  Accuracy of mechanical torque devices for implants used in Brazilian dental offices.

Authors:  Gildo Coelho Santos; Sheila Pestana Passos; Maria Jacinta Moraes Coelho Santos
Journal:  Int J Prosthodont       Date:  2011 Jan-Feb       Impact factor: 1.681

3.  Accuracy of a manual torque application device for morse-taper implants: a technical note.

Authors:  Murat C Cehreli; Kivang Akça; Ergin Tönük
Journal:  Int J Oral Maxillofac Implants       Date:  2004 Sep-Oct       Impact factor: 2.804

4.  Probabilistic analysis of preload in the abutment screw of a dental implant complex.

Authors:  Teja Guda; Thomas A Ross; Lisa A Lang; Harry R Millwater
Journal:  J Prosthet Dent       Date:  2008-09       Impact factor: 3.426

Review 5.  A systematic review of the survival and complication rates of implant supported fixed dental prostheses with cantilever extensions after an observation period of at least 5 years.

Authors:  Marco Aglietta; Vincenzo Iorio Siciliano; Marcel Zwahlen; Urs Brägger; Bjarni E Pjetursson; Niklaus P Lang; Giovanni E Salvi
Journal:  Clin Oral Implants Res       Date:  2009-05       Impact factor: 5.977

6.  Variability of mechanical torque-limiting devices in clinical service at a US dental school.

Authors:  Michael S McCracken; Lillian Mitchell; Rashmi Hegde; Mahendra D Mavalli
Journal:  J Prosthodont       Date:  2009-09-17       Impact factor: 2.752

7.  Elongation and preload stress in dental implant abutment screws.

Authors:  J E Haack; R L Sakaguchi; T Sun; J P Coffey
Journal:  Int J Oral Maxillofac Implants       Date:  1995 Sep-Oct       Impact factor: 2.804

8.  Accuracy of the implant torque wrench following time in clinical service.

Authors:  J Gutierrez; J I Nicholls; W J Libman; T J Butson
Journal:  Int J Prosthodont       Date:  1997 Nov-Dec       Impact factor: 1.681

Review 9.  Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs).

Authors:  Bjarni E Pjetursson; Urs Brägger; Niklaus P Lang; Marcel Zwahlen
Journal:  Clin Oral Implants Res       Date:  2007-06       Impact factor: 5.977

10.  The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Ali Fayaz; Seyed Jalil Sadr; Friedrich Lampert; Maziar Mir
Journal:  Clin Cosmet Investig Dent       Date:  2012-07-31
View more
  2 in total

1.  The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants.

Authors:  Ali Fayaz; Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ghassem Ansari
Journal:  Dent Res J (Isfahan)       Date:  2014-01

2.  Effects of Aging Torque Controllers on Screw Tightening Force and Bacterial Micro-Leakage on the Implant-Abutment Complex.

Authors:  Yousef Jiries; Tamar Brosh; Shlomo Matalon; Vladimir Perlis; Zeev Ormianer
Journal:  Materials (Basel)       Date:  2022-01-14       Impact factor: 3.623

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.