Accuracy of mechanical torque devices for implants used in Brazilian dental offices.

Gildo Coelho Santos1, Sheila Pestana Passos, Maria Jacinta Moraes Coelho Santos.   

Abstract

The purpose of this study was to determine the accuracy of mechanical torque devices in delivering target torque values in dental offices in Salvador, Brazil. A team of researchers visited 16 dental offices, and the clinicians applied torque values (20 and 32 Ncm) to electronic torque controllers. Five repetitions were completed at each torque value and data were collected. When 20 Ncm of torque was used, 62.5% of measured values were accurate (within 10% of the target value). For 32 Ncm, however, only 37.5% of these values were achieved. Several of the tested mechanical torque devices were inaccurate.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21210000

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  4 in total

1.  The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayyaz; Mahdi Kadkhodazadeh
Journal:  J Periodontal Implant Sci       Date:  2013-10-29       Impact factor: 2.614

2.  The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants.

Authors:  Ali Fayaz; Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ghassem Ansari
Journal:  Dent Res J (Isfahan)       Date:  2014-01

3.  The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants.

Authors:  Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayaz; Minoo Mahshid
Journal:  J Dent (Tehran)       Date:  2013-01-31

4.  The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Ali Fayaz; Seyed Jalil Sadr; Friedrich Lampert; Maziar Mir
Journal:  Clin Cosmet Investig Dent       Date:  2012-07-31