Accuracy of the implant torque wrench following time in clinical service.

J Gutierrez1, J I Nicholls, W J Libman, T J Butson.   

Abstract

Thirty-five implant torque wrenches were tested for torque delivery accuracy. Seven of these drivers had a 10-Ncm value, 12 had a 20-Ncm value, 10 had a 30-Ncm value, and six had a 35-Ncm value. All of the wrenches had been in clinical service for a minimum of 1 month or a maximum of 3 years and were assumed by the clinicians to be fully effective in delivering the required torque values. Torque delivery accuracy was determined using a Tohnichi torque wrench calibrated to +/- 1% of the scale value. All torque wrenches had been subjected to steam sterilization following clinical use. Data collected on all wrenches included (1) age since purchase, (2) number of sterilization cycles to which each wrench had been subjected during the use period, and (3) 10 sequential torque readings using the Tohnichi wrench as the torque reading device. The results of this study showed that there was no significant correlation of delivered torque with (1) age or (2) number of sterilization cycles. Furthermore, in each torque category there were torques both above and below the control values. For the 10-Ncm torque wrenches, the largest value was 455.0% greater than the control, the largest values were 41.6% greater for 20-Ncm wrenches, 17.0% greater for the 30-Ncm torque wrenches, and 58.6% greater for the 35-Ncm wrenches. Corrosion of the spring in the handle of the torque wrench was found to be the reason for the 455.0% value.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9495178

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  8 in total

1.  The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayyaz; Mahdi Kadkhodazadeh
Journal:  J Periodontal Implant Sci       Date:  2013-10-29       Impact factor: 2.614

2.  Experimental study on the factors affecting torque of beam-type implant torque wrenches.

Authors:  Hiroki Shiba; Yuji Sato; Junichi Furuya; Tokiko Osawa; Akio Isobe; Myu Hayashi; Noboru Kitagawa
Journal:  BMC Oral Health       Date:  2021-07-15       Impact factor: 2.757

3.  The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants.

Authors:  Ali Fayaz; Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ghassem Ansari
Journal:  Dent Res J (Isfahan)       Date:  2014-01

4.  Evaluation of stability of interface between CCM (Co-Cr-Mo) UCLA abutment and external hex implant.

Authors:  Ki-Joon Yoon; Young-Bum Park; Hyunmin Choi; Youngsung Cho; Jae-Hoon Lee; Keun-Woo Lee
Journal:  J Adv Prosthodont       Date:  2016-12-15       Impact factor: 1.904

5.  The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems.

Authors:  Mehmet Ali Erdem; Burcin Karatasli; Onur Dinçer Kose; Taha Emre Kose; Erhan Çene; Serhan Aydın Aya; Abdulkadir Burak Cankaya
Journal:  Biomed Res Int       Date:  2017-11-07       Impact factor: 3.411

6.  The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants.

Authors:  Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayaz; Minoo Mahshid
Journal:  J Dent (Tehran)       Date:  2013-01-31

7.  Evaluation of the accuracy and related factors of the mechanical torque-limiting device for dental implants.

Authors:  Mahmood Kazemi; Ahmad Rohanian; Abbas Monzavi; Mohammad Sadegh Nazari
Journal:  J Dent (Tehran)       Date:  2013-03-31

8.  The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Ali Fayaz; Seyed Jalil Sadr; Friedrich Lampert; Maziar Mir
Journal:  Clin Cosmet Investig Dent       Date:  2012-07-31
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.