Variability of mechanical torque-limiting devices in clinical service at a US dental school.

Michael S McCracken1, Lillian Mitchell, Rashmi Hegde, Mahendra D Mavalli.   

Abstract

PURPOSE: The purpose of this study was to measure the variability of torque produced by a population of mechanical torque-limiting devices in clinical service in a US dental school. The torque-limiting devices were divided into two categories according to their mode of action: toggle-type and beam wrenches. Proper action of these devices is essential for calibrated delivery of preload to implant prosthetic screws.
MATERIALS AND METHODS: Seventeen torque-limiting devices (35 Ncm) were obtained from graduate prosthodontic, predoctoral, and faculty practice clinics. Nine of these were toggle-type devices, and eight were beam-type wrenches. Torque from each wrench was measured using an MGT electronic torque meter. Wrenches were tested in two modes, slow (over 4 seconds) and fast (over 1 second).
RESULTS: Toggle-type torque wrenches produced a mean (+/- SD) torque of 38.1 +/- 16.0 Ncm; beam-type wrenches produced 32.8 +/- 1.1 Ncm. These results were not significantly different. When tested in fast mode (1 second), toggle-type wrenches produced 28.0 +/- 9.6 Ncm; in the slow mode (4 seconds) they produced significantly more force, 36.6 +/- 14.0 Ncm (p < 0.001). Beam-type wrenches produced 33.2 +/- 1.1 Ncm and 32.8 +/- 1.1 Ncm in fast and slow modes, respectively.
CONCLUSIONS: Both types of wrenches tested were capable of producing accurate torque values; however, variability was higher in the toggle-type group. Some toggle-type torque wrenches in clinical service delivered unacceptably high torque values. It is recommended that clinicians calibrate toggle-type wrenches frequently. Torque wrenches should be activated slowly, over 4 seconds, when using a correctly calibrated toggle-type wrench.

Entities:  

Mesh:

Year:  2009        PMID: 19765196     DOI: 10.1111/j.1532-849X.2009.00524.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  7 in total

1.  The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayyaz; Mahdi Kadkhodazadeh
Journal:  J Periodontal Implant Sci       Date:  2013-10-29       Impact factor: 2.614

2.  Experimental study on the factors affecting torque of beam-type implant torque wrenches.

Authors:  Hiroki Shiba; Yuji Sato; Junichi Furuya; Tokiko Osawa; Akio Isobe; Myu Hayashi; Noboru Kitagawa
Journal:  BMC Oral Health       Date:  2021-07-15       Impact factor: 2.757

3.  The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants.

Authors:  Ali Fayaz; Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ghassem Ansari
Journal:  Dent Res J (Isfahan)       Date:  2014-01

4.  The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems.

Authors:  Mehmet Ali Erdem; Burcin Karatasli; Onur Dinçer Kose; Taha Emre Kose; Erhan Çene; Serhan Aydın Aya; Abdulkadir Burak Cankaya
Journal:  Biomed Res Int       Date:  2017-11-07       Impact factor: 3.411

5.  The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants.

Authors:  Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayaz; Minoo Mahshid
Journal:  J Dent (Tehran)       Date:  2013-01-31

6.  The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Ali Fayaz; Seyed Jalil Sadr; Friedrich Lampert; Maziar Mir
Journal:  Clin Cosmet Investig Dent       Date:  2012-07-31

7.  Effects of Aging Torque Controllers on Screw Tightening Force and Bacterial Micro-Leakage on the Implant-Abutment Complex.

Authors:  Yousef Jiries; Tamar Brosh; Shlomo Matalon; Vladimir Perlis; Zeev Ormianer
Journal:  Materials (Basel)       Date:  2022-01-14       Impact factor: 3.623