Atsushi Kameyama1, Akiko Haruyama, Masako Asami, Toshiyuki Takahashi.
Abstract
The purpose of this study was to determine any discrepancies in the outputs of five commercial dental radiometers and also to evaluate the accuracy of these devices using a laboratory-grade spectroradiometer. The power densities of 12 different curing light sources were repeatedly measured for a total of five times using each radiometer in a random order. The emission spectra of all of the curing light sources were also measured using the spectroradiometer, and the integral value of each spectrum was calculated to determine the genuine power densities, which were then compared to the displayed power densities measured by the dental radiometers. The displayed values of power density were various and were dependent on the brand of radiometer, and this may be because each radiometer has a different wavelength sensitivity. These results cast doubt upon the accuracy of commercially available dental radiometers.Entities:
Mesh:
Year: 2013 PMID: 24222744 PMCID: PMC3810486 DOI: 10.1155/2013/647941
Source DB: PubMed Journal: ScientificWorldJournal ISSN: 1537-744X
Curing light sources used in this study.
| Light-curing unit | Manufacturer | Type | Light guide | Curing mode |
|---|---|---|---|---|
| D-Lux 2000 (JB1646) | Dentrade Dental Supply & Trade, Osaka, Japan | QTH | Straight Turbo (11/8 mm) | High |
| Jetlite 3000 (9040763) | J. Morita USA, Mason Irvine, CA, USA | QTH | Straight | High |
| Demi Plus (760012733) | Kerr Hawe, Middleton, WI, USA | Single-peak LED | Straight Turbo (11/7 mm) | — |
| Pencure (SN13444) | J. Morita, Tokyo, Japan | Single-peak LED | Turbo (11/8 mm) | — |
| G-Light Prima (00156) | GC, Tokyo, Japan | Dual-wave LED | Normal | |
| Bluephase G2 (209409) | Ivoclar Vivadent, Schaan, Liechtenstein | Dual-wave LED | Straight | High |
Radiometers investigated in this study.
| Radiometer | Manufacturer | Type | Measurable wavelength | Measurable output intensity |
|---|---|---|---|---|
| USR-45 (—) | Ushio, Tokyo, Japan | Labo-grade | 200–800 nm | — |
| Bluephase Meter (005386) | Ivoclar Vivadent, Schaan, Liechtenstein | Digital | 380–515 nm | 300–2500 mW/cm2 |
| Cure Rite (7781) | Dentsply Caulk, Milford, DE, USA | Digital | 400–525 nm | <2000 mW/cm2 |
| Curing Light Meter 105 (4237/2009) | Rolence Enterprise, Chingli, Taiwan | Analogue | 400–500 nm | <2500 mW/cm2 |
| Model 100 Optilux Radiometer (145975) | SDS Kerr, Middleton, WI, USA | Analogue | 400–500 nm | <1000 mW/cm2 |
| Model L.E.D. Radiometer (79310406) | SDS Kerr, Middleton, WI, USA | Analogue | 400–500 nm | <2000 mW/cm2 |
Figure 1Spectral distributions of the 12 curing light sources used in this study.
Power density calculated from integral value of spectra at 200–800 nm, 380–525 nm, and 430–490 nm (n = 5, mW/cm2).
| Light-curing unit | Light guide/mode | 200–800 nm | 380–525 nm | 430–490 nm |
|---|---|---|---|---|
| D-Lux 2000 | Straight guide | 790 ± 54 | 728 ± 41 | 509 ± 28 |
| (100%) | (92.2%) | (64.4%) | ||
| Turbo guide | 1450 ± 56 | 1336 ± 51 | 940 ± 37 | |
| (100%) | (92.1%) | (64.8%) | ||
|
| ||||
| Jetlite 3000 | — | 1016 ± 57 | 966 ± 56 | 613 ± 32 |
| (100%) | (95.0%) | (60.3%) | ||
|
| ||||
| Demi Plus | Straight guide | 837 ± 6 | 811 ± 5 | 781 ± 5 |
| (100%) | (96.9%) | (93.3%) | ||
| Turbo guide | 1997 ± 58 | 1947 ± 56 | 1881 ± 55 | |
| (100%) | (97.5%) | (94.1%) | ||
|
| ||||
| Pencure | — | 1275 ± 18 | 1236 ± 17 | 1139 ± 16 |
| (100%) | (96.9%) | (89.3%) | ||
|
| ||||
| G-Light Prima | Normal | 2121 ± 68 | 2068 ± 66 | 1922 ± 58 |
| (100%) | (97.5%) | (90.6%) | ||
| F5 | 3334 ± 85 | 3238 ± 82 | 2991 ± 75 | |
| (100%) | (97.1%) | (89.7%) | ||
| PH | 2169 ± 100 | 2112 ± 100 | 1615 ± 101 | |
| (100%) | (97.3%) | (74.4%) | ||
| PL | 51 ± 3 | 42 ± 3 | 0.9 ± 0 | |
| (100%) | (82.3%) | (1.8%) | ||
|
| ||||
| Bluephase G2 | High | 1528 ± 27 | 1489 ± 25 | 1215 ± 29 |
| (100%) | (97.4%) | (79.5%) | ||
| Low | 839 ± 5 | 815 ± 7 | 563 ± 18 | |
| (100%) | (97.1%) | (67.1%) | ||
Displayed light output (Mean ± S.D.; mW/cm2) and percentage of the reduction from USR-45 (%) (n = 5).
| USR-45 | Bluephase Meter | Cure Rite | Curing Light Meter 105 | Model 100 Optilux | Model L.E.D. Radiometer | |
|---|---|---|---|---|---|---|
| D-Lux 2000 | 728 ± 41a | 618 ± 31b | 426 ± 27c | 306 ± 26d | 616 ± 32b | 442 ± 35c |
| (Straight) | (100%) | (84.9%) | (58.5%) | (42.0%) | (84.6%) | (60.7%) |
| D-Lux 2000 | 1336 ± 51a | 1316 ± 94b | 682 ± 64d | 690 ± 42d | 980 ± 27c | 710 ± 42d |
| (Turbo) | (100%) | (98.5%) | (51.0%) | (51.6%) | (73.4%) | (53.1%) |
| Jetlite 3000 | 966 ± 56a | 730 ± 25b | 543 ± 42c | 456 ± 13d | 688 ± 48b | 526 ± 37cd |
| (100%) | (75.6%) | (56.2%) | (47.2%) | (71.2%) | (54.5%) | |
| Demi Plus | 811 ± 5a | 684 ± 25b | 538 ± 8d | 492 ± 11e | 790 ± 11a | 600 ± 0c |
| (Straignt) | (100%) | (84.3%) | (66.3%) | (60.7%) | (97.4%) | (74.0%) |
| Demi Plus | 1947 ± 56a | 1756 ± 83b | 1395 ± 21c | 1680 ± 45b | 1000 ± 0∗d | 1410 ± 22c |
| (Turbo) | (100%) | (90.2%) | (71.6%) | (86.3%) | (51.4%) | (72.4%) |
| Pencure | 1236 ± 17a | 868 ± 26d | 917 ± 17c | 830 ± 35d | 1000 ± 0∗b | 850 ± 35d |
| (100%) | (70.2%) | (74.2%) | (67.1%) | (80.9%) | (68.8%) | |
| G-Light Prima | 2068 ± 66a | 1702 ± 72b | 2000 ± 0∗a | 880 ± 27e | 1000 ± 0∗d | 1130 ± 27c |
| (Normal) | (100%) | (82.3%) | (96.7%) | (42.6%) | (48.4%) | (54.6%) |
| G-Light Prima | 3238 ± 82a | 2672 ± 143b | 2000 ± 0∗d | 2500 ± 0∗c | 1000 ± 0∗e | 1950 ± 50d |
| (F5 mode) | (100%) | (82.5%) | (61.8%) | (77.2%) | (30.9%) | (60.2%) |
| G-Light Prima | 2112 ± 100a | 1496 ± 23 | 2000 ± 0∗b | 800 ± 0e | 1000 ± 0∗d | 1010 ± 22d |
| (PH mode) | (100%) | (70.8%) | (94.7%) | (37.9%) | (47.3%) | (47.8%) |
| G-Light Prima | 42 ± 3a | 0 ± 0b | 0 ± 0b | 0 ± 0b | 0 ± 0b | 0 ± 0b |
| (PL mode) | (100%) | (0%) | (0%) | (0%) | (0%) | (0%) |
| Bluephase G2 | 1489 ± 25a | 1312 ± 71b | 870 ± 3e | 970 ± 27d | 1000 ± 0∗cd | 1038 ± 16c |
| (High) | (100%) | (88.1%) | (58.4%) | (65.1%) | (67.2%) | (69.7%) |
| Bluephase G2 | 815 ± 7a | 668 ± 33b | 406 ± 3d | 400 ± 0d | 700 ± 0b | 550 ± 0b |
| (Low) | (100%) | (82.0%) | (49.8%) | (49.1%) | (85.9%) | (67.5%) |
Within the same row, different letters indicate groups that are statistically different (P < 0.05).
*Achieved to the highest limit of radiometer concerned.