Vivek Hegde1, Sameer Jadhav, Gayatri B Aher.
Abstract
AIM: The purpose of this study is to examine the intensity of light curing units and factors affecting it in dental offices.Entities:
Keywords: Output intensity; light curing units; light-emitting diode radiometer
Year: 2009 PMID: 20543916 PMCID: PMC2879716 DOI: 10.4103/0972-0707.57633
Source DB: PubMed Journal: J Conserv Dent ISSN: 0972-0707
Distribution of studied parameters
| Machines | ||
|---|---|---|
| LED (n=81) | QTH (n=119) | |
| No. of months used | 36.0 (4.0–78.0) | 36.0 (2.0–80.0) |
| No. of times battery/bulb changed | ||
| None | 58 (71.6) | 78 (65.5) |
| Once | 21 (25.9) | 36 (30.3) |
| Twice | 2 (2.5) | 5 (4.2) |
| Intensity by operator-I | ||
| <200 mW/cm2 | 15 (18.5) | 39 (32.8) |
| 200–400 | 58 (71.6) | 78 (65.5) |
| >400 | 8 (9.9) | 2 (1.7) |
| Intensity by operator-II | ||
| <200 mW/cm2 | 14 (17.3) | 37 (31.1) |
| 200-400 | 58 (71.6) | 80 (67.2) |
| >400 | 9 (11.1) | 2 (1.7) |
Values are median (minimum-maximum) values are n (%)
Association with duration of use (months)
| Machines | No. of months used | |||||
|---|---|---|---|---|---|---|
| <36 months | >36 months | LED | QTH | |||
| LED | QTH | LED | QTH | |||
| No. of times battery changed | ||||||
| None | 19 (82.6) | 21 (75.0) | 39 (67.2) | 57 (67.6) | 0.326 | 0.306 |
| Once | 4 (17.4) | 7 (25.0) | 7 (25.0) | 29 (31.9) | ||
| Twice | 0 | 0 | 2 (3.4) | 5 (5.5) | ||
| Intensity | ||||||
| 200 mW/cm2 | 1 (3.6) | 15 (25.9) | 38 (41.8) | 0.000 | 0.000 | |
| 200–400 | 15 (65.2) | 25 (89.3) | 43 (74.1) | 53 (58.2) | ||
| >400 | 8 (34.8) | 2 (7.1) | 0 | 0 | ||
Values are n (%), P values by Chi-square test
Association with number of times the battery or bulb is changed
| Machine | No. of times battery changed | |||||
|---|---|---|---|---|---|---|
| None | Once or twice | LED | QTH | |||
| LED | QTH | LED | QTH | |||
| Intensity | ||||||
| <200 mW/cm2 | 10 (17.2) | 31 (39.7) | 5 (21.7) | 8 (19.5) | 0.709 | 0.048 |
| 200–400 | 43 (74.1) | 48 (59.0) | 32 (78.0) | 53 (58.2) | ||
| >400 | 5 (8.6) | 1 (1.3) | 3 (13.0) | 1 (2.4) | ||
Agreement between two independent operators (intensity-light-emitting diode)
| Operator-II | ||||
|---|---|---|---|---|
| Intensity mW/cm2 | <200 | 200–400 | >400 | |
| Operator-I | <200 | 14 (93.3) | 1 (6.7) | 0 |
| 200–400 | 0 | 57(98.3) | 1(1.7) | |
| >400 | 0 | 0 | 8 (100.0) | |
Values are n (%), P value 5 0.000 by Chi-square test, suggesting strong agreement between two independent operators
Agreement between two independent operators (intensity - quartz-tungsten-halogen)
| Operator-II | ||||
|---|---|---|---|---|
| Intensity mW/cm2 | <200 | 200–400 | >400 | |
| Operator-I | <200 | 37 (94.9) | 2 (5.1) | 0 |
| 200–400 | 0 | 78 (100.0) | 0 | |
| >400 | 0 | 0 | 2 (100.0) | |
Values are n(%), P value = 0.000 by Chi-square test, suggesting strong agreement between two independent operators
Figure 1Proportion of faulty machines in use