Literature DB >> 24146436

Development of a Cost-Effective Modular Pixelated NaI(Tl) Detector for Clinical SPECT Applications.

Mike Rozler1, Haoning Liang, Wei Chang.   

Abstract

A new pixelated detector for high-resolution clinical SPECT applications was designed and tested. The modular detector is based on a scintillator block comprised of 2.75×2.75×10 mm3 NaI(Tl) pixels and decoded by an array of 51 mm diameter single-anode PMTs. Several configurations, utilizing two types of PMTs, were evaluated using a collimated beam source to measure positioning accuracy directly. Good pixel separation was observed, with correct pixel identification ranging from 60 to 72% averaged over the entire area of the modules, depending on the PMT type and configuration. This translates to a significant improvement in positioning accuracy compared to continuous slab detectors of the same thickness, along with effective reduction of "dead" space at the edges. The observed 10% average energy resolution compares well to continuous slab detectors. The combined performance demonstrates the suitability of pixelated detectors decoded with a relatively small number of medium-sized PMTs as a cost-effective approach for high resolution clinical SPECT applications, in particular those involving curved detector geometries.

Entities:  

Keywords:  Biomedical imaging; gamma camera; single photon emission computed tomography

Year:  2012        PMID: 24146436      PMCID: PMC3800469          DOI: 10.1109/TNS.2012.2210908

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  8 in total

1.  C-SPECT - a Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential.

Authors:  Wei Chang; Caesar E Ordonez; Haoning Liang; Yusheng Li; Jingai Liu
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-06       Impact factor: 1.679

2.  Recent advances in SPECT imaging.

Authors:  Mark T Madsen
Journal:  J Nucl Med       Date:  2007-04       Impact factor: 10.057

3.  The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements.

Authors:  Raymond R Raylman; Stan Majewski; Mark F Smith; James Proffitt; William Hammond; Amarnath Srinivasan; John McKisson; Vladimir Popov; Andrew Weisenberger; Clifford O Judy; Brian Kross; Srikanth Ramasubramanian; Larry E Banta; Paul E Kinahan; Kyle Champley
Journal:  Phys Med Biol       Date:  2008-01-10       Impact factor: 3.609

4.  An evaluation of a two-dimensional array detector for high resolution PET.

Authors:  M Dahlbom; E J Hoffman
Journal:  IEEE Trans Med Imaging       Date:  1988       Impact factor: 10.048

5.  Importance of consideration of radiation doses from cardiac imaging procedures and risks of cancer.

Authors:  George A Beller
Journal:  J Nucl Cardiol       Date:  2010 Jan-Feb       Impact factor: 5.952

6.  A new apparatus for brain imaging: four-head rotating gamma camera single-photon emission computed tomograph.

Authors:  K Kimura; K Hashikawa; H Etani; A Uehara; T Kozuka; H Moriwaki; Y Isaka; M Matsumoto; T Kamada; H Moriyama
Journal:  J Nucl Med       Date:  1990-05       Impact factor: 10.057

7.  Maximum-Likelihood Methods for Processing Signals From Gamma-Ray Detectors.

Authors:  Harrison H Barrett; William C J Hunter; Brian William Miller; Stephen K Moore; Yichun Chen; Lars R Furenlid
Journal:  IEEE Trans Nucl Sci       Date:  2009-06-01       Impact factor: 1.679

8.  Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras.

Authors:  Fabio P Esteves; Paolo Raggi; Russell D Folks; Zohar Keidar; J Wells Askew; Shmuel Rispler; Michael K O'Connor; Liudmilla Verdes; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2009-08-18       Impact factor: 5.952

  8 in total