Literature DB >> 24146175

Primary stability of the Fitmore stem: biomechanical comparison.

Wojciech Pepke1, Jan Nadorf, Volker Ewerbeck, Marcus R Streit, Stefan Kinkel, Tobias Gotterbarm, Michael W Maier, J Philippe Kretzer.   

Abstract

PURPOSE: After clinical introduction of the Fitmore stem (Zimmer), we noticed the formation of cortical hypertrophies in a few cases. We questioned whether (1) the primary stability or (2) load transfer of the Fitmore stem differs from other stems unassociated with the formation of hypertrophies. We compared the Fitmore stem to the well-established CLS stem.
METHODS: Four Fitmore and four CLS stems were implanted in eight synthetic femurs. A cyclic torque around the stem axis and a mediolateral cyclic torque were applied. Micromotions between stems and femurs were measured to classify the specific rotational implant stability and to analyse the bending behaviour of the stem.
RESULTS: No statistical differences were found between the two stem designs with respect to their rotational stability (p = 0.82). For both stems, a proximal fixation was found. However, for the mediolateral bending behavior, we observed a significantly (p < 0.01) higher flexibility of the CLS stem compared to the Fitmore stem.
CONCLUSION: Hip stem implantation may induce remodelling of the periprosthetic bone structure. Considering the proximal fixation of both stems, rotational stability of the Fitmore® stem might not be a plausible explanation for clinically observed formation of hypertrophies. However, bending results support our hypothesis that the CLS stem presumably closely follows the bending of the bone, whereas the shorter Fitmore stem acts more rigidly. Stem rigidity and flexibility needs to be considered, as they may influence the load transfer at the implant-bone interface and thus possibly affect bone remodelling processes.

Entities:  

Mesh:

Year:  2013        PMID: 24146175      PMCID: PMC3936089          DOI: 10.1007/s00264-013-2138-4

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  28 in total

1.  Changes in bone mineral density in the proximal femur after cementless total hip arthroplasty. A five-year longitudinal study.

Authors:  W Brodner; P Bitzan; F Lomoschitz; P Krepler; R Jankovsky; S Lehr; F Kainberger; F Gottsauner-Wolf
Journal:  J Bone Joint Surg Br       Date:  2004-01

2.  The effect of multifilaments and monofilaments on cementless femoral revision hip components: an experimental study.

Authors:  Eike Jakubowitz; Stefan Kinkel; Jan Nadorf; Christian Heisel; J Philippe Kretzer; Marc N Thomsen
Journal:  Clin Biomech (Bristol, Avon)       Date:  2010-12-13       Impact factor: 2.063

3.  Bone remodeling characteristics of a short-stemmed total hip replacement.

Authors:  Hsiang-Ho Chen; Bernard F Morrey; Kai-Nan An; Zong-Ping Luo
Journal:  J Arthroplasty       Date:  2008-10-09       Impact factor: 4.757

4.  [Dynamic force measurement in implantation of total endoprostheses of the hip joint].

Authors:  U Schmidbauer; T Brendel; K G Kunze; M Nietert; H Ecke
Journal:  Unfallchirurgie       Date:  1993-02

5.  Fit of the uncemented femoral component and the use of cement influence the strain transfer the femoral cortex.

Authors:  M Jasty; D O O'Connor; R M Henshaw; T P Harrigan; W H Harris
Journal:  J Orthop Res       Date:  1994-09       Impact factor: 3.494

6.  A conservative femoral replacement for total hip arthroplasty. A prospective study.

Authors:  B F Morrey; R A Adams; M Kessler
Journal:  J Bone Joint Surg Br       Date:  2000-09

7.  Bone remodeling around stable uncemented titanium stems during the second decade after total hip arthroplasty: a DXA study at 12 and 17 years.

Authors:  C Merle; M R Streit; C Volz; M Pritsch; T Gotterbarm; P R Aldinger
Journal:  Osteoporos Int       Date:  2010-11-23       Impact factor: 4.507

8.  ESB Research Award 1992. The mechanism of bone remodeling and resorption around press-fitted THA stems.

Authors:  B Van Rietbergen; R Huiskes; H Weinans; D R Sumner; T M Turner; J O Galante
Journal:  J Biomech       Date:  1993 Apr-May       Impact factor: 2.712

9.  Short stems for total hip arthroplasty: initial experience with the Fitmore stem.

Authors:  K Gustke
Journal:  J Bone Joint Surg Br       Date:  2012-11

10.  High survival in young patients using a second generation uncemented total hip replacement.

Authors:  Marcus R Streit; Kerstin Schröder; Matthias Körber; Christian Merle; Tobias Gotterbarm; Volker Ewerbeck; Peter R Aldinger
Journal:  Int Orthop       Date:  2011-11-24       Impact factor: 3.075

View more
  7 in total

1.  Can the metaphyseal anchored Metha short stem safely be revised with a standard CLS stem? A biomechanical analysis.

Authors:  Shuang G Yan; Matthias Woiczinski; Tobias F Schmidutz; Patrick Weber; Alexander C Paulus; Arnd Steinbrück; Volkmar Jansson; Florian Schmidutz
Journal:  Int Orthop       Date:  2017-05-09       Impact factor: 3.075

2.  Length of clinically proven cemented hip stems: state of the art or subject to improvement?

Authors:  Moussa Hamadouche; Alexander Jahnke; Caroline Scemama; Bernd Alexander Ishaque; Markus Rickert; Luc Kerboull; Eike Jakubowitz
Journal:  Int Orthop       Date:  2014-09-20       Impact factor: 3.075

3.  Primary rotational stability of various megaprostheses in a biomechanical sawbone model with proximal femoral defects extending to the isthmus.

Authors:  Stefan Kinkel; Jan Nadorf; Jan Dennis Graage; Eike Jakubowitz; Jan Philippe Kretzer
Journal:  PLoS One       Date:  2015-06-01       Impact factor: 3.240

4.  Cortical hypertrophy with a short, curved uncemented hip stem does not have any clinical impact during early follow-up.

Authors:  Michael W Maier; Marcus R Streit; Moritz M Innmann; Marlis Krüger; Jan Nadorf; J Philippe Kretzer; Volker Ewerbeck; Tobias Gotterbarm
Journal:  BMC Musculoskelet Disord       Date:  2015-12-01       Impact factor: 2.362

5.  Comparison of radiographic changes in rectangular curved short stem with thin versus thick porous coating for cementless total hip arthroplasty: a retrospective study with a propensity score matching.

Authors:  Yutaro Munakata; Yujiro Kuramitsu; Yutaka Usui; Ken Okazaki
Journal:  J Orthop Surg Res       Date:  2021-04-13       Impact factor: 2.359

6.  Promising medium-term results of anterior approach with an anatomical short stem in primary hip arthroplasty.

Authors:  Fabrizio Rivera; Alessandro Bardelli; Andrea Giolitti
Journal:  J Orthop Traumatol       Date:  2021-03-06

7.  Do radiological findings around the Fitmore stem change over time? : radiolucency around the short hip stem disappears within a few years: a retrospective study.

Authors:  Hideki Fujii; Tetsuo Hayama; Toshiomi Abe; Motoi Takahashi; Ayano Amagami; Yohei Matsushita; Takuya Otani; Mitsuru Saito
Journal:  Bone Jt Open       Date:  2022-01
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.