The effect of multifilaments and monofilaments on cementless femoral revision hip components: an experimental study.

Eike Jakubowitz1, Stefan Kinkel, Jan Nadorf, Christian Heisel, J Philippe Kretzer, Marc N Thomsen.   

Abstract

BACKGROUND: Cerclage wires are widely used in revision hip surgery to reattach the lid of a femoral osteotomy. The present study compared the influence of multifilaments and monofilaments on primary stability of revision hip stems with different fixation principles.
METHODS: A standardized extended proximal femoral osteotomy was performed in the anterior cortex of 6 synthetic femora. We used a high-resolution measuring device to explore spatial micromovements of a diaphyseal and a metaphyseal fixating revision stem. Both of these were implanted in 3 femora. The specimens were measured again after consecutive restabilization of osteotomies with multifilaments and monofilaments. The movement graphs generated defined relative micromovements between stems and bones and the stabilizing effect of the two wire systems compared.
FINDINGS: Both multifilaments and monofilaments effected a major reduction of relative micromovements for both fixation principles. There were no differences in relative movements between the multifilament and monofilament treatments for the diaphyseal fixating stem. Yet for the metaphyseal fixating stem a significantly better restabilization was observed with multifilaments.
INTERPRETATION: Both multifilaments and monofilaments can support the revision hip stem in bridging the extended proximal femoral osteotomy. Yet, which wiring system should be chosen depends on the fixation principle of the revision stem. Multifilaments seem to be advantageous when used with metaphyseal fixating stems. However, the use of multifilaments with diaphyseal fixating components should be reconsidered as this might constrict the periosteal vascularity.
Copyright © 2010 Elsevier Ltd. All rights reserved.

Mesh:

Substances:

Year:  2010        PMID: 21146907     DOI: 10.1016/j.clinbiomech.2010.11.004

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  9 in total

1.  Influence of stem design on the primary stability of megaprostheses of the proximal femur.

Authors:  Stefan Kinkel; Jan Dennis Graage; Jan Philippe Kretzer; Eike Jakubowitz; Jan Nadorf
Journal:  Int Orthop       Date:  2013-08-18       Impact factor: 3.075

2.  Fixation pattern of conical and cylindrical modular revision hip stems in different size bone defects.

Authors:  Stefan Kinkel; Jan Nadorf; Marc N Thomsen; Christian Heisel; Alexander Jahnke; Jan P Kretzer; Eike Jakubowitz
Journal:  Int Orthop       Date:  2015-07-09       Impact factor: 3.075

3.  [The biomechanics of screws, cerclage wire and cerclage cable].

Authors:  C Schröder; M Woiczinski; S Utzschneider; M Kraxenberger; P Weber; V Jansson
Journal:  Orthopade       Date:  2013-05       Impact factor: 1.087

Review 4.  Past and present of the use of cerclage wires in orthopedics.

Authors:  Andrea Angelini; Concetto Battiato
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-09-04

5.  Length of clinically proven cemented hip stems: state of the art or subject to improvement?

Authors:  Moussa Hamadouche; Alexander Jahnke; Caroline Scemama; Bernd Alexander Ishaque; Markus Rickert; Luc Kerboull; Eike Jakubowitz
Journal:  Int Orthop       Date:  2014-09-20       Impact factor: 3.075

6.  Primary stability of the Fitmore stem: biomechanical comparison.

Authors:  Wojciech Pepke; Jan Nadorf; Volker Ewerbeck; Marcus R Streit; Stefan Kinkel; Tobias Gotterbarm; Michael W Maier; J Philippe Kretzer
Journal:  Int Orthop       Date:  2013-10-22       Impact factor: 3.075

7.  Prospects of implant with locking plate in fixation of subtrochanteric fracture: experimental demonstration of its potential benefits on synthetic femur model with supportive hierarchical nonlinear hyperelastic finite element analysis.

Authors:  Mohammed Hadi Latifi; Kunalan Ganthel; Shanmugam Rukmanikanthan; Azura Mansor; Tunku Kamarul; Mehmet Bilgen
Journal:  Biomed Eng Online       Date:  2012-04-30       Impact factor: 2.819

8.  Primary rotational stability of various megaprostheses in a biomechanical sawbone model with proximal femoral defects extending to the isthmus.

Authors:  Stefan Kinkel; Jan Nadorf; Jan Dennis Graage; Eike Jakubowitz; Jan Philippe Kretzer
Journal:  PLoS One       Date:  2015-06-01       Impact factor: 3.240

9.  Tibial revision knee arthroplasty with metaphyseal sleeves: The effect of stems on implant fixation and bone flexibility.

Authors:  Jan Nadorf; Stefan Kinkel; Simone Gantz; Eike Jakubowitz; J Philippe Kretzer
Journal:  PLoS One       Date:  2017-05-08       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.