Comparison of the Easyshade Compact and Advance in vitro and in vivo.

Andreas Zenthöfer1, Tomas Cabrera, Nicoleta Corcodel, Peter Rammelsberg, Alexander J Hassel.   

Abstract

OBJECTIVE: The objective of this study was to evaluate the repeatability, inter-examiner and inter-device reliability, and accuracy of the new VITA Easyshade Advance (ES-A) spectrophotometer in comparison with the Easyshade Compact (ES-C).
METHODS: For the in vitro study, six porcelain disks of random shade mixtures were fabricated from VITA ceramic powders. To analyze the repeatability of the ES-C and ES-A, readings were obtained by one examiner for one porcelain disk and one shade tab (VITA 3D-Master). Measurements were repeated by two examiners to evaluate inter-examiner reliability. To assess accuracy, porcelain disks and shade tabs were measured with the ES-C and ES-A and by means of a spectroradiometer (SR). ∆E values between the shade tabs and disks were calculated. Differences between devices were evaluated by use of univariate analysis. Reliability and accuracy of both devices were also tested in a clinical setting. The intra-class correlation coefficient (ICC) was calculated for all assessments of reliability.
RESULTS: Good agreement was observed for replicate measurements (ICC > 0.9), for inter-examiner reliability of the ES-A and ES-C (ICC > 0.9), and for inter-device agreement (ICC > 0.9). Univariate analysis revealed that ∆E values for the spectrophotometers were consistent (p = 0.783); spectroradiometrically measured ∆E was comparable, but the mean value was smaller than that for the spectrophotometers (p = 0.004). Clinical inter-device agreement was acceptable to high (ICC > 0.777). The accuracy of the ES-A was higher than that of the ES-C (p = 0.001).
CONCLUSION: Within the limitations of the experiment, the spectrophotometers tested enabled reliable and accurate measurement. In the clinical setting, however, the ES-A seems more accurate. CLINICAL RELEVANCE: ES-C and ES-A can be valuable tools for the determination of tooth colors. Because of their good reliability, they should especially be used for the assessment of tooth color changes over time.

Entities:  

Mesh:

Year:  2013        PMID: 24100636     DOI: 10.1007/s00784-013-1118-9

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  29 in total

1.  The repeatability of an intraoral dental colorimeter.

Authors:  Francis F Tung; Gary R Goldstein; Sungkoo Jang; Eugene Hittelman
Journal:  J Prosthet Dent       Date:  2002-12       Impact factor: 3.426

2.  Matching repeatability and interdevice agreement of 2 intraoral spectrophotometers.

Authors:  Aspasia Sarafianou; Phophi Kamposiora; George Papavasiliou; Helen Goula
Journal:  J Prosthet Dent       Date:  2012-03       Impact factor: 3.426

3.  Intraoral determination of the tolerance of dentists for perceptibility and acceptability of shade mismatch.

Authors:  R Duane Douglas; Tad J Steinhauer; Alvin G Wee
Journal:  J Prosthet Dent       Date:  2007-04       Impact factor: 3.426

4.  Coverage error of three conceptually different shade guide systems to vital unrestored dentition.

Authors:  Funda Bayindir; Shanglun Kuo; William M Johnston; Alvin G Wee
Journal:  J Prosthet Dent       Date:  2007-09       Impact factor: 3.426

5.  Reliability and accuracy of four dental shade-matching devices.

Authors:  Seungyee Kim-Pusateri; Jane D Brewer; Elaine L Davis; Alvin G Wee
Journal:  J Prosthet Dent       Date:  2009-03       Impact factor: 3.426

6.  The Ishihara test as a means of detecting and analyzing color vision.

Authors:  L H HARDY; G RAND; M C RITTLER
Journal:  J Gen Psychol       Date:  1947-01

7.  A comparison of new and conventional methods for quantification of tooth color.

Authors:  T P van der Burgt; J J ten Bosch; P C Borsboom; W J Kortsmit
Journal:  J Prosthet Dent       Date:  1990-02       Impact factor: 3.426

8.  A comparison between conventional visual and spectrophotometric methods for shade selection.

Authors:  Ahmad Judeh; Ahed Al-Wahadni
Journal:  Quintessence Int       Date:  2009-10       Impact factor: 1.677

9.  Reliability of shade selection using an intraoral spectrophotometer.

Authors:  Siegbert Witkowski; Nao-Daniel Yajima; Martin Wolkewitz; Jorge R Strub
Journal:  Clin Oral Investig       Date:  2011-08-10       Impact factor: 3.573

10.  New shade guide for evaluation of tooth whitening--colorimetric study.

Authors:  Rade D Paravina; William M Johnston; John M Powers
Journal:  J Esthet Restor Dent       Date:  2007       Impact factor: 2.843

View more
  9 in total

1.  In Vivo and in Vitro Evaluations of Repeatability and Accuracy of VITA Easyshade® Advance 4.0 Dental Shade-Matching Device.

Authors:  Dubravka Knezović; Davor Zlatarić; Iva Ž Illeš; Maja Alajbeg
Journal:  Acta Stomatol Croat       Date:  2015-06

2.  How reliable and accurate is the shade determination of premolars by spectrophotometry?

Authors:  Anna-Luisa Klotz; Yasamin Habibi; Alexander Jochen Hassel; Peter Rammelsberg; Andreas Zenthöfer
Journal:  Clin Oral Investig       Date:  2019-12-14       Impact factor: 3.573

3.  Final Colour of Ultratranslucent Multilayered Zirconia Veneers, Effect of Thickness, and Resin Cement Shade.

Authors:  Seyed Reza Khosravani; Mehdi Abed Kahnamoui; Soodabeh Kimyai; Elmira Jafari Navimipour; Farzaneh Sadeghi Mahounak; Fatemeh Pournaghi Azar
Journal:  Biomed Res Int       Date:  2022-06-01       Impact factor: 3.246

4.  Effect of enamel sealants on tooth bleaching and on the color stability of the result.

Authors:  N Corcodel; A J Hassel; S Sen; D Saure; P Rammelsberg; C J Lux; S Zingler
Journal:  Odontology       Date:  2016-05-20       Impact factor: 2.634

5.  Efficacy of whitening oral rinses and dentifrices on color stability of bleached teeth.

Authors:  Muhammet Karadas
Journal:  Acta Biomater Odontol Scand       Date:  2015-04-27

6.  A Digital Shade-Matching Device for Dental Color Determination Using the Support Vector Machine Algorithm.

Authors:  Minah Kim; Byungyeon Kim; Byungjun Park; Minsuk Lee; Youngjae Won; Choul-Young Kim; Seungrag Lee
Journal:  Sensors (Basel)       Date:  2018-09-12       Impact factor: 3.576

7.  Correlation of natural tooth colour with aging in the Spanish population.

Authors:  Cristina Gómez Polo; Miguel Gómez Polo; Javier Montero; Juan Antonio Martínez Vazquez De Parga; Alicia Celemin Viñuela
Journal:  Int Dent J       Date:  2015-07-27       Impact factor: 2.607

8.  Discoloration of surface sealants by plaque disclosing solution.

Authors:  Sinan Şen; Ralf Erber; Gözde Şen; Nadine Deurer; Sebastian Zingler; Christopher J Lux
Journal:  J Orofac Orthop       Date:  2020-05-06       Impact factor: 1.938

9.  Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study.

Authors:  Hee-Kyung Kim
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.