Reliability and accuracy of four dental shade-matching devices.

Seungyee Kim-Pusateri1, Jane D Brewer, Elaine L Davis, Alvin G Wee.   

Abstract

STATEMENT OF PROBLEM: There are several electronic shade-matching instruments available for clinical use, but the reliability and accuracy of these instruments have not been thoroughly investigated.
PURPOSE: The purpose of this in vitro study was to evaluate the reliability and accuracy of 4 dental shade-matching instruments in a standardized environment.
MATERIAL AND METHODS: Four shade-matching devices were tested: SpectroShade, ShadeVision, VITA Easyshade, and ShadeScan. Color measurements were made of 3 commercial shade guides (Vitapan Classical, Vitapan 3D-Master, and Chromascop). Shade tabs were placed in the middle of a gingival matrix (Shofu GUMY) with shade tabs of the same nominal shade from additional shade guides placed on both sides. Measurements were made of the central region of the shade tab positioned inside a black box. For the reliability assessment, each shade tab from each of the 3 shade guide types was measured 10 times. For the accuracy assessment, each shade tab from 10 guides of each of the 3 types evaluated was measured once. Differences in reliability and accuracy were evaluated using the Standard Normal z test (2 sided) (alpha=.05) with Bonferroni correction.
RESULTS: Reliability of devices was as follows: ShadeVision, 99.0%; SpectroShade, 96.9%; VITA Easyshade, 96.4%; and ShadeScan, 87.4%. A significant difference in reliability was found between ShadeVision and ShadeScan (P=.008). All other comparisons showed similar reliability. Accuracy of devices was as follows: VITA Easyshade, 92.6%; ShadeVision, 84.8%; SpectroShade, 80.2%; and ShadeScan, 66.8%. Significant differences in accuracy were found between all device pairs (P<.001) for all comparisons except for SpectroShade versus ShadeVision (P=.033).
CONCLUSIONS: Most devices had similar high reliability (over 96%), indicating predictable shade values from repeated measurements. However, there was more variability in accuracy among devices (67-93%), and differences in accuracy were seen with most device comparisons.

Mesh:

Year:  2009        PMID: 19231572     DOI: 10.1016/S0022-3913(09)60028-7

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  33 in total

1.  Color change of some aesthetic dental materials: Effect of immersion solutions and finishing of their surfaces.

Authors:  Elizabeth Sarkis
Journal:  Saudi Dent J       Date:  2012-03-10

2.  Comparison of the Easyshade Compact and Advance in vitro and in vivo.

Authors:  Andreas Zenthöfer; Tomas Cabrera; Nicoleta Corcodel; Peter Rammelsberg; Alexander J Hassel
Journal:  Clin Oral Investig       Date:  2013-10-08       Impact factor: 3.573

3.  Reliability of shade selection using an intraoral spectrophotometer.

Authors:  Siegbert Witkowski; Nao-Daniel Yajima; Martin Wolkewitz; Jorge R Strub
Journal:  Clin Oral Investig       Date:  2011-08-10       Impact factor: 3.573

4.  In vivo study for tooth colour determination-visual versus digital.

Authors:  Jan Brandt; Shirley Nelson; Hans-Christoph Lauer; Ulrike von Hehn; Silvia Brandt
Journal:  Clin Oral Investig       Date:  2017-03-06       Impact factor: 3.573

5.  Is personality relevant in the choice of bleaching?

Authors:  A Herrera; J Martín; F Pérez; E Bonafé; A Reis; A Loguercio Dourado; E Fernández
Journal:  Clin Oral Investig       Date:  2016-01-11       Impact factor: 3.573

6.  A novel regression model from RGB image data to spectroradiometric correlates optimized for tooth colored shades.

Authors:  Melody N Carney; William M Johnston
Journal:  J Dent       Date:  2016-05-31       Impact factor: 4.379

7.  In Vivo and in Vitro Evaluations of Repeatability and Accuracy of VITA Easyshade® Advance 4.0 Dental Shade-Matching Device.

Authors:  Dubravka Knezović; Davor Zlatarić; Iva Ž Illeš; Maja Alajbeg
Journal:  Acta Stomatol Croat       Date:  2015-06

8.  How reliable and accurate is the shade determination of premolars by spectrophotometry?

Authors:  Anna-Luisa Klotz; Yasamin Habibi; Alexander Jochen Hassel; Peter Rammelsberg; Andreas Zenthöfer
Journal:  Clin Oral Investig       Date:  2019-12-14       Impact factor: 3.573

9.  The Effect of two Shading Techniques on Value of Zirconia-Based Crowns.

Authors:  Ahmad Hassan Ahangari; Kianoosh Torabi Ardakani; Farideh Mahdavi; Mahshid Torabi Ardakani
Journal:  J Dent (Shiraz)       Date:  2015-06

10.  Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up.

Authors:  Rafael Francisco Lia Mondelli; Juliana Felipi David E Góes de Azevedo; Ana Carolina Francisconi; Cristiane Machado de Almeida; Sérgio Kiyoshi Ishikiriama
Journal:  J Appl Oral Sci       Date:  2012 Jul-Aug       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.