Moe Akahira-Azuma1, Naohiro Yonemoto, Battsengel Ganzorig, Rintaro Mori, Shinichi Hosokawa, Takeji Matsushita, Bayasgalantai Bavuusuren, Enkhtur Shonkhuuz.
Abstract
BACKGROUND: Neonatal hyperbilirubinemia, especially kernicterus, can be prevented by screening for neonatal jaundice. The transcutaneous bilirubin (TcB) meter is a non-invasive medical device for screening neonates. The study aimed to investigate the validity of a TcB meter in a resource-limited setting such as Mongolia.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24074043 PMCID: PMC3850747 DOI: 10.1186/1471-2431-13-151
Source DB: PubMed Journal: BMC Pediatr ISSN: 1471-2431 Impact factor: 2.125
Clinical characteristics of the 53 ethnic Mongolian neonates
| Birth Weight (g): | <2500 g | 2 | 3.8 |
| | 2500-3499 g | 29 | 54.7 |
| | 3500-4499 g | 22 | 41.5 |
| Gestational Age (week): | 35-36 weeks | 6 | 11.3 |
| | 37-40 weeks | 43 | 81.2 |
| | 41-42 weeks | 4 | 7.5 |
| Delivery Mode: | Normal Vaginal | 41 | 77.4 |
| | Cesarean Section | 12 | 22.6 |
| Postnatal day of measurement of | Day 0-10 | 19 | 35.9 |
| | Day 11-20 | 15 | 28.2 |
| Day 21-31 | 19 | 35.9 |
Summary statistics of bilirubin measurements
| TSB | 53 | 12.55 | 1.70-21.01 | 4.819 | 1.7 | 9.20 | 12.5 | 15.95 | 21.6 |
| TcB Forehead | 53 | 12.87 | 1.07-21.56 | 4.945 | 1.0 | 9.75 | 12.9 | 16.25 | 21.7 |
| TcB Midsternum | 53 | 13.12 | 1.72-21.63 | 4.755 | 1.4 | 9.85 | 13.7 | 16.00 | 21.8 |
TSB: total serum bilirubin; TcB: transcutaneous serum bilirubin; SD: standard deviation; CI: confidence interval.
Figure 1Linear regression plots (solid lines) of JM-103 TcB TSB measurements at different measurement sites (A: forehead, B: midsternum) in the neonate.
Figure 2Bland-Altman plots of TSB and TcB in the neonates. The graphs show the difference in mean values between TSB and TcB (forehead) (A), TSB and TcB (midsternum) (B), and TcB (midsternum) and TcB (forehead) (C). The P-value by the Bradley-Blackwood test was 0.315 in (A), 0.073 in (B) and 0.247 in (C). FH: forehead; CH: midsternum.
Figure 3Receiver operating characteristics curves for JM-103 cutoff values (mg/dL) at TSB >10 mg/dL, >13 mg/dL, or >15 mg/dL. The areas under the curves were 0.962 (A), 0.945 (B), 0.908 (C), 0.987 (D), 0.913 (E), and 0.903 (F).