Evaluation of a new transcutaneous bilirubinometer.

M Jeffrey Maisels1, Enrique M Ostrea, Suzanne Touch, Sarah E Clune, Eugene Cepeda, Elizabeth Kring, Karin Gracey, Cheryl Jackson, Deborah Talbot, Raywin Huang.   

Abstract

OBJECTIVE: The objective of this study was to evaluate the Minolta/Hill-Rom Air-Shields Transcutaneous Jaundice Meter model JM-103.
METHODS: We studied a convenience sample of 849 newborns > or =35 weeks of gestation in 3 hospitals. These infants had total serum bilirubin (TSB) levels measured on clinical indication, and transcutaneous bilirubin (TcB) levels were obtained within 1 hour of the TSB levels. The population was 59.2% white, 29.8% black, 4.5% East Asian, 3.8% Middle Eastern, 1.6% Indian/Pakistani, and 1.1% Hispanic.
RESULTS: There was a close correlation between TSB and TcB values in all of the population groups: white (n = 503, r =.949); black (n = 253, r =.822); and East Asian, Indian/Pakistani, and Hispanic (n = 93, r =.926). In the black population, the correlation was less close than in the other groups, and differences between the TcB and TSB measurements tended to increase with rising TSB values. JM-103 values differed from TSB values by 3 mg/dL or more in 2% of white, 3.2% of other, and 17.4% of black infants. In these black infants, the JM-103 value was always greater than the TSB value.
CONCLUSIONS: We conclude that TcB measurements using the JM-103 jaundice meter correlate very closely with TSB levels over the range of TSB encountered in this study. Because only 3.3% of our infants had TSB values >15 mg/dL (257 micro mol/L), more data are needed in this range of TSB concentration. The correlation in black infants is not as close as in other groups, but because the tendency in blacks is for the JM-103 to overestimate serum bilirubin levels, dangerous clinical errors are unlikely to occur. The measurement technique is rapid and simple, and it is easy to perform repeated measurements over time, thus reducing the likelihood of error. TcB measurements with the JM-103 jaundice meter should obviate the need for most serum bilirubin levels in newborn infants > or =35 weeks of gestation, although serum bilirubin measurements are still required when treatment with phototherapy or exchange transfusion is being considered.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15173483     DOI: 10.1542/peds.113.6.1628

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  32 in total

1.  Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants.

Authors:  Zhang-Bin Yu; Xiao-Yue Dong; Shu-Ping Han; Yu-Lin Chen; Yu-Fang Qiu; Li Sha; Qing Sun; Xi-Rong Guo
Journal:  Eur J Pediatr       Date:  2010-09-03       Impact factor: 3.183

2.  Surveillance of severe neonatal hyperbilirubinemia: a view from south of the border.

Authors:  M Jeffrey Maisels; Thomas B Newman
Journal:  CMAJ       Date:  2006-09-12       Impact factor: 8.262

3.  Transcutaneous bilirubin in predicting hyperbilirubinemia in term neonates.

Authors:  Y Ramesh Bhat; Amitha Rao
Journal:  Indian J Pediatr       Date:  2008-02       Impact factor: 1.967

4.  "Transcutaneous bilirubin in predicting hyperbilirubinemia in term neonates." Correspondence.

Authors:  Deepak Chawla; Y Ramesh Bhat; Amitha Rao
Journal:  Indian J Pediatr       Date:  2009-10       Impact factor: 1.967

5.  Natural history of early neonatal bilirubinemia: a global perspective.

Authors:  Michael Kaplan; M Jeffrey Maisels
Journal:  J Perinatol       Date:  2021-01-04       Impact factor: 2.521

6.  Reliability of transcutaneous bilirubin determination based on skin color determined by a neonatal skin color scale of our own.

Authors:  Silvia Maya-Enero; Júlia Candel-Pau; Jordi Garcia-Garcia; Xavier Duran-Jordà; María Ángeles López-Vílchez
Journal:  Eur J Pediatr       Date:  2021-01-06       Impact factor: 3.183

7.  Photoacoustic microscopy of bilirubin in tissue phantoms.

Authors:  Yong Zhou; Chi Zhang; Da-Kang Yao; Lihong V Wang
Journal:  J Biomed Opt       Date:  2012-12       Impact factor: 3.170

8.  Transcutaneous bilirubin measurement at the time of hospital discharge in a multiethnic newborn population.

Authors:  Douglas M Campbell; Karoon C Danayan; Valleverdina McGovern; Sohail Cheema; Brenda Stade; Michael Sgro
Journal:  Paediatr Child Health       Date:  2011-03       Impact factor: 2.253

9.  Utility of Decision Rules for Transcutaneous Bilirubin Measurements.

Authors:  James A Taylor; Anthony E Burgos; Valerie Flaherman; Esther K Chung; Elizabeth A Simpson; Neera K Goyal; Isabelle Von Kohorn; Niramol Dhepyasuwan
Journal:  Pediatrics       Date:  2016-04-06       Impact factor: 7.124

10.  Comparison of a new transcutaneous bilirubinometer (Bilimed) with serum bilirubin measurements in preterm and full-term infants.

Authors:  Tanja Karen; Hans Ulrich Bucher; Jean-Claude Fauchère
Journal:  BMC Pediatr       Date:  2009-11-12       Impact factor: 2.125

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.