| Literature DB >> 35283687 |
Mazrah Mohamed1, Nor Rosidah Ibrahim1,2, Noraida Ramli1,2, Noorizan Abdul Majid1,2, Najib Majdi Yacob3, Ariffin Nasir1,2.
Abstract
Background: This study aims to investigate the reliability of the Dräger Jaundice Meter JM-105 for the screening of neonatal jaundice in Malay neonates.Entities:
Keywords: JM-105 bilirubinometer; neonatal jaundice; total serum bilirubin; transcutaneous bilirubin
Year: 2022 PMID: 35283687 PMCID: PMC8887975 DOI: 10.21315/mjms2022.29.1.5
Source DB: PubMed Journal: Malays J Med Sci ISSN: 1394-195X
Figure 1Study flow chart
Demographic characteristics (N = 130)
| Gender | |
| Female | 73 (56.2) |
| Male | 57 (43.8) |
| Birth weight (g) | |
| 2,000–2,500 | 17 (13.1) |
| ≥ 2,501 | 113 (86.9) |
| Gestational age (weeks) | |
| 35–36 | 28 (21.5) |
| 37–42 | 102 (78.5) |
| G6PD status | |
| Deficient | 9 (6.9) |
| Normal | 121 (93.1) |
| Mother’s blood group | |
| A positive | 30 (23.1) |
| B positive | 32 (24.6) |
| O positive | 60 (46.1) |
| AB positive | 8 (6.2) |
| Baby’s blood group | |
| A positive | 34 (26.2) |
| B positive | 38 (29.2) |
| O positive | 52 (40) |
| AB positive | 6 (4.6) |
| Mode of delivery | |
| LSCS | 31 (23.8) |
| SVD | 99 (76.2) |
| Family history with neonatal jaundice | |
| Yes | 72 (55.4) |
| No | 58 (44.6) |
| Feeding | |
| Exclusive breast feeding | 66 (50.8) |
| Mixed feeding | 64 (49.2) |
Notes: LSCS = lower segment caesarean section; SVD = spontaneous vaginal delivery
Multiple pairwise comparisons of TSB, sternum and forehead TcB
| Comparison | Mean difference | 95% CI of difference | |
|---|---|---|---|
| TSB versus forehead TcB | 10.10 | 1.29, 18.90 | 0.019 |
| TSB versus sternum TcB | 9.27 | 0.05, 18.48 | 0.048 |
| Forehead TcB versus sternum TcB | −0.83 | −6.18, 4.53 | > 0.95 |
Note:
Adjustment for multiple comparisons using the Bonferroni method
Figure 2Linear regression plot for (A) TSB versus the forehead TcB, and (B) TSB versus the sternum TcB
Figure 3Bland-Altman plot showing the difference and average of (A) the forehead TcB and sternum TcB, (B) sternum TcB and TSB, and (C) forehead TcB and sternum TcB
Figure 4Comparison of the area under receiver operating characteristics (ROC) curve between the forehead TcB and sternum TcB at TSB level of 205 μmol/L
Diagnostic accuracy of the forehead TcB and sternum TcB compared to the TSB at the cut-off level TSB > 205 μmol/L
| Forehead TcB (%) | Sternum TcB (%) | |||
|---|---|---|---|---|
|
|
| |||
| Estimate | 95% CI | Estimate | 95% CI | |
| Sensitivity | 85.3 | 79.3, 91.4 | 84.4 | 78.2, 90.7 |
| Specificity | 76.4 | 69.1, 83.7 | 77.4 | 70.2, 84.6 |
| Positive predictive value | 83.1 | 76.7, 89.6 | 84.4 | 78.2, 90.7 |
| Negative predictive value | 79.3 | 72.3, 86.2 | 77.4 | 70.2, 84.6 |
Figure 5Characterisation of the difference between (A) the forehead TcB-TSB and (B) sternum TcB-TSB throughout the measurement range of TSB. The shaded area represents the 95% CI of the linear fit line