Evaluation of a new transcutaneous bilirubinometer in Chinese newborns.

H T Ho1, T K Ng, K C Tsui, Y C Lo.   

Abstract

OBJECTIVE: To evaluate the use of a new transcutaneous bilirubinometer (JM-103 Minolta Airshields) for detection of hyperbilirubinaemia in term or near-term healthy Chinese newborns.
METHODS: Transcutaneous bilirubin (TcB) was used to screen for severe hyperbilirubinaemia in newborn infants. Blood was taken for total serum bilirubin (TSB) measurement if the initial TcB level was higher than the 40th centile in Bhutani's nomogram. Paired TcB and TSB results were then reviewed over 6 months. The correlation as well as the mean difference between the two methods were calculated. The clinical application of TcB with Bhutani's nomogram in the prediction of severe hyperbilirubinaemia in low-risk, medium-risk and high-risk thresholds for phototherapy was also analysed.
RESULTS: 997 paired TcB and TSB measurements were evaluated in term or near-term newborns. TcB was significantly correlated with TSB, with a correlation coefficient of 0.83 (p<0.001). Their mean difference was 21.7 micromol/l (SD 21.2, p<0.001), with the 95% limits of agreement between -19.9 and 63.3 micromol/l. In both low-risk and medium-risk thresholds for phototherapy, using the 75th centile of Bhutani's nomogram as threshold, TcB could identify all cases and had a sensitivity and negative predictive value of 100% each, a specificity of 56% and positive predictive value of 23%. For high-risk cases, using the 75th centile as cut-off, the sensitivity and negative predictive value were reduced to 86.7% and 97.0%, respectively.
CONCLUSION: An accurate point-of-care bilirubin analyser facilitates bilirubin screening and avoids unnecessary blood tests. Although using the transcutaneous bilirubinometer JM-103 might result in a significant difference between TcB and TSB measured in Chinese newborns, combining the use of TcB and the 75th centile in Bhutani's nomogram as the cut-off level can identify all cases of significant hyperbilirubinaemia.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16849367      PMCID: PMC2672758          DOI: 10.1136/adc.2005.090217

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  20 in total

1.  Measuring bilirubin through the skin?

Authors:  Vinod K Bhutani; Lois H Johnson; Glenn Gourley
Journal:  Pediatrics       Date:  2003-04       Impact factor: 7.124

Review 2.  An evidence-based review of important issues concerning neonatal hyperbilirubinemia.

Authors:  Stanley Ip; Mei Chung; John Kulig; Rebecca O'Brien; Robert Sege; Stephan Glicken; M Jeffrey Maisels; Joseph Lau
Journal:  Pediatrics       Date:  2004-07       Impact factor: 7.124

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  New transcutaneous jaundice device with two optical paths.

Authors:  Saneyuki Yasuda; Susumu Itoh; Kenichi Isobe; Masahiko Yonetani; Hajime Nakamura; Makoto Nakamura; Yoshitada Yamauchi; Akio Yamanishi
Journal:  J Perinat Med       Date:  2003       Impact factor: 1.901

5.  Transcutaneous bilirubinometry: bilirubin kinetics of the skin and serum during and after phototherapy.

Authors:  Y Yamauchi; I Yamanouchi
Journal:  Biol Neonate       Date:  1989

6.  Candidate reference method for determination of total bilirubin in serum: development and validation.

Authors:  B T Doumas; P P Kwok-Cheung; B W Perry; B Jendrzejczak; R B McComb; R Schaffer; L L Hause
Journal:  Clin Chem       Date:  1985-11       Impact factor: 8.327

7.  Neonatal jaundice: its prevalence in Chinese babies and associating factors.

Authors:  T F Fok; S P Lau; C W Hui
Journal:  Aust Paediatr J       Date:  1986-08

8.  Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation.

Authors: 
Journal:  Pediatrics       Date:  2004-07       Impact factor: 7.124

9.  Neonatal jaundice in full-term infants. Role of breast-feeding and other causes.

Authors:  M J Maisels; K Gifford
Journal:  Am J Dis Child       Date:  1983-06

10.  Transcutaneous bilirubinometry: preliminary studies of noninvasive transcutaneous bilirubin meter in the Okayama National Hospital.

Authors:  I Yamanouchi; Y Yamauchi; I Igarashi
Journal:  Pediatrics       Date:  1980-02       Impact factor: 7.124

View more
  5 in total

1.  Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants.

Authors:  Zhang-Bin Yu; Xiao-Yue Dong; Shu-Ping Han; Yu-Lin Chen; Yu-Fang Qiu; Li Sha; Qing Sun; Xi-Rong Guo
Journal:  Eur J Pediatr       Date:  2010-09-03       Impact factor: 3.183

2.  Validation of a transcutaneous bilirubin meter in Mongolian neonates: comparison with total serum bilirubin.

Authors:  Moe Akahira-Azuma; Naohiro Yonemoto; Battsengel Ganzorig; Rintaro Mori; Shinichi Hosokawa; Takeji Matsushita; Bayasgalantai Bavuusuren; Enkhtur Shonkhuuz
Journal:  BMC Pediatr       Date:  2013-09-27       Impact factor: 2.125

3.  Comparison between the Transcutaneous and Total Serum Bilirubin Measurement in Malay Neonates with Neonatal Jaundice.

Authors:  Mazrah Mohamed; Nor Rosidah Ibrahim; Noraida Ramli; Noorizan Abdul Majid; Najib Majdi Yacob; Ariffin Nasir
Journal:  Malays J Med Sci       Date:  2022-02-23

4.  Evaluation of transcutaneous bilirubinometer (DRAEGER JM 103) use in Zimbabwean newborn babies.

Authors:  Gwendoline Lilly Tanyaradzwa Chimhini; Simbarashe Chimhuya; Vasco Chikwasha
Journal:  Matern Health Neonatol Perinatol       Date:  2018-01-18

5.  Evaluation of the Relationship Between Transcutaneous Bilirubin Measurement and Total Serum Bilirubin in Neonatal Patients Followed for Jaundice.

Authors:  Fatma Cetinkaya Cat; Abdulkadir Cat; Tuba Cicek; Seda Geylani Gulec
Journal:  Sisli Etfal Hastan Tip Bul       Date:  2021-07-02
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.