Literature DB >> 12661149

New transcutaneous jaundice device with two optical paths.

Saneyuki Yasuda1, Susumu Itoh, Kenichi Isobe, Masahiko Yonetani, Hajime Nakamura, Makoto Nakamura, Yoshitada Yamauchi, Akio Yamanishi.   

Abstract

OBJECTIVES: To confirm the accuracy and precision of transcutaneous bilirubin (TcB) values measured by a new device with two optical paths (JM-103) and the value of total serum bilirubin (TSB) level in clinical units of measurement.
METHODS: For comparison of the levels of accuracy and precision of JM-103 and the old device (JM-102), serum samples were collected from 77 Japanese infants in three different hospitals including 24 preterm infants and 53 term infants. Measurement of TcB by JM-103 and JM-102 were performed on the forehead of each infant within 30 min before or after blood sampling.
RESULTS: The range of TSB was limited to 19.6 mg/dL and to 17 mg/dL for preterm infants. The correlation coefficients for all subjects (r = 0.94) and for term and preterm subjects between TcB measured by JM-103 and TSB was higher than that between TcB measured by JM-102 and TSB. The regression line in term infants between TcB measured by JM-103 and TcB was similar to that in preterm infants. The error distribution of TcB measured by JM-102 and TSB for all subjects (0.00 +/- 2.21) and for term and preterm subjects was larger than that of TcB measured by JM-103 and TSB (all subjects, 0.30 +/- 1.55).

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12661149     DOI: 10.1515/JPM.2003.012

Source DB:  PubMed          Journal:  J Perinat Med        ISSN: 0300-5577            Impact factor:   1.901


  10 in total

1.  Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants.

Authors:  Zhang-Bin Yu; Xiao-Yue Dong; Shu-Ping Han; Yu-Lin Chen; Yu-Fang Qiu; Li Sha; Qing Sun; Xi-Rong Guo
Journal:  Eur J Pediatr       Date:  2010-09-03       Impact factor: 3.183

2.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

3.  Transcutaneous bilirubin in predicting hyperbilirubinemia in term neonates.

Authors:  Y Ramesh Bhat; Amitha Rao
Journal:  Indian J Pediatr       Date:  2008-02       Impact factor: 1.967

4.  Evaluation of a new transcutaneous bilirubinometer in Chinese newborns.

Authors:  H T Ho; T K Ng; K C Tsui; Y C Lo
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2006-07-18       Impact factor: 5.747

5.  Hyperbilirubinemia in Term Newborns Needing Phototherapy within 48 Hours after Birth in a Japanese Birth Center.

Authors:  Saeko Tsujimae; Katsuhiko Yoshii; Keiji Yamana; Kazumichi Fujioka; Kazumoto Iijima; Ichiro Morioka
Journal:  Kobe J Med Sci       Date:  2018-09-11

6.  Accuracy and precision of transcutaneous bilirubinometry in postdischarge Asian neonates.

Authors:  Ratchada Kitsommart; Pornpat Pornladnun; Chulathida Chomchai; Pornvilai Urujchutchairut; Bosco Paes
Journal:  Eur J Pediatr       Date:  2013-02-09       Impact factor: 3.183

7.  Validation of a transcutaneous bilirubin meter in Mongolian neonates: comparison with total serum bilirubin.

Authors:  Moe Akahira-Azuma; Naohiro Yonemoto; Battsengel Ganzorig; Rintaro Mori; Shinichi Hosokawa; Takeji Matsushita; Bayasgalantai Bavuusuren; Enkhtur Shonkhuuz
Journal:  BMC Pediatr       Date:  2013-09-27       Impact factor: 2.125

8.  Should transcutaneous bilirubin be measured in preterm infants receiving phototherapy? The relationship between transcutaneous and total serum bilirubin in preterm infants with and without phototherapy.

Authors:  Christian V Hulzebos; Deirdre E Vader-van Imhoff; Arend F Bos; Peter H Dijk
Journal:  PLoS One       Date:  2019-06-14       Impact factor: 3.240

9.  How skin anatomy influences transcutaneous bilirubin determinations: an in vitro evaluation.

Authors:  Marlijn D van Erk; Alida J Dam-Vervloet; Foky-Anna de Boer; Martijn F Boomsma; Henrica van Straaten; Nienke Bosschaart
Journal:  Pediatr Res       Date:  2019-06-24       Impact factor: 3.756

10.  Evaluation of transcutaneous bilirubinometer (DRAEGER JM 103) use in Zimbabwean newborn babies.

Authors:  Gwendoline Lilly Tanyaradzwa Chimhini; Simbarashe Chimhuya; Vasco Chikwasha
Journal:  Matern Health Neonatol Perinatol       Date:  2018-01-18