Comparison of erythrocyte sedimentation rate measurement by the automated SEDIsystem and conventional Westergren method using the Bland and Altman statistical method.

Suad M AlFadhli1, Anwar M Al-Awadhi.   

Abstract

OBJECTIVES: To compare the performance of SEDIsystem(TM), a fully automated analyzer for the measurement of the erythrocyte sedimentation rate (ESR), with the manual Westergren method.
MATERIALS AND METHODS: Both methods were applied to 150 randomly selected subjects. The linear regression and Bland and Altman data analysis methods were used to measure the agreement between the automated and manual methods.
RESULTS: The regression analysis showed a good correlation between the two methods (r=0.91). The Bland and Altman data analysis showed no systematic bias (95% confidence interval for mean difference); however, limits of agreement were between 11.52 and -37.88. This indicates that ESR values measured by the SEDIsystem may be 11.52 mm/h above or 37.88 mm/h below the reference method. A greater scatter of data was also observed with abnormally high (>25 mm/h) ESR results (mean of difference=-21.4 and limits of agreement=-45.2 and 2.26) compared with normal (<25 mm/h) readings (mean of difference=-3.9 and limits of agreement=-13.5 and 5.7).
CONCLUSION: The Bland and Altman statistical analysis showed a wide degree of scatter between results obtained by the two ESR techniques that was not clearly demonstrated using the linear regression analysis. The automated system was found to underestimate ESR with the Bland and Altman statistical analysis, and therefore a correction factor is recommended. Copyright (c) 2005 S. Karger AG, Basel.

Mesh:

Year:  2005        PMID: 15961933     DOI: 10.1159/000085742

Source DB:  PubMed          Journal:  Med Princ Pract        ISSN: 1011-7571            Impact factor:   1.927


  6 in total

1.  Evaluation of Ves-Matic Cube 200 for erythrocyte sedimentation rate determination.

Authors:  Sevilay Sezer; Fatma Meriç Yilmaz; Oğuzhan Kaya; Sema Uysal
Journal:  J Clin Lab Anal       Date:  2013-09       Impact factor: 2.352

2.  Erythrocyte sedimentation rate measurement by VES Matic Cube 80 in relation to inflammation plasma proteins.

Authors:  Helena Cerutti; Chiara Muzzi; Roberto Leoncini; Carlo Scapellato; Alessio Cortelazzo; Emiliano Furlani; Roberto Guerranti
Journal:  J Clin Lab Anal       Date:  2011       Impact factor: 2.352

3.  A Comparison Between the StaRRsed Auto-Compact Erythrocyte Sedimentation Rate Instrument and the Westergren Method.

Authors:  Juha Horsti; Riikka Rontu; Auni Collings
Journal:  J Clin Med Res       Date:  2010-12-11

4.  Diagnosis of periprosthetic joint infection: the threshold for serological markers.

Authors:  Pouya Alijanipour; Hooman Bakhshi; Javad Parvizi
Journal:  Clin Orthop Relat Res       Date:  2013-10       Impact factor: 4.176

5.  Incidence of giant cell arteritis in Western Norway 1972-2012: a retrospective cohort study.

Authors:  L K Brekke; A P Diamantopoulos; B-T Fevang; J Aβmus; E Esperø; C G Gjesdal
Journal:  Arthritis Res Ther       Date:  2017-12-15       Impact factor: 5.156

6.  Early evidence of efficacy for orally administered SPM-enriched marine lipid fraction on quality of life and pain in a sample of adults with chronic pain.

Authors:  Nini Callan; Doug Hanes; Ryan Bradley
Journal:  J Transl Med       Date:  2020-10-21       Impact factor: 5.531

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.