The effects of pressure on gases in solution: possible insights to improve microbubble filtration for extracorporeal circulation.

Daniel P Herbst1.   

Abstract

Improvements in micropore arterial line filter designs used for extracorporeal circulation are still needed because microbubbles larger than the rated pore sizes are being detected beyond the filter outlet. Linked to principles governing the function of micropore filters, fluid pressures contained in extracorporeal circuits also influence the behavior of gas bubbles and the extent to which they are carried in a fluid flow. To better understand the relationship between pressure and microbubble behavior, two ex vivo test circuits with and without inline resistance were designed to assess changes in microbubble load with changes in pressure. Ultrasound Doppler probes were used to measure and compare the quality and quantity of microbubbles generated in each test circuit. Analysis of microbubble load was separated into two distinct phases, the time periods during and immediately after bubble generation. Although microbubble number decreased similarly in both test circuits, changes in microbubble volume were significant only in the test circuit with inline resistance. The test circuit with inline resistance also showed a decrease in the rate of volume transferred across each ultrasound Doppler probe and the microbubble number and size range measured in the postbubble generation period. The present research proposes that fluid pressures contained in extracorporeal circuits may be used to affect gases in solution as a possible method to improve microbubble filtration during extracorporeal circulation.

Mesh:

Substances:

Year:  2013        PMID: 23930378      PMCID: PMC4557586     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  18 in total

1.  How effective are cardiopulmonary bypass circuits at removing gaseous microemboli?

Authors:  Timothy J Jones; Dwight D Deal; Jason C Vernon; Noel Blackburn; David A Stump
Journal:  J Extra Corpor Technol       Date:  2002-03

2.  Evaluation of Quadrox-i adult hollow fiber oxygenator with integrated arterial filter.

Authors:  Yulong Guan; Xiaowei Su; Robert McCoach; Robert Wise; Allen Kunselman; Akif Undar
Journal:  J Extra Corpor Technol       Date:  2010-06

3.  In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.

Authors:  Timothy A Dickinson; Jeffrey B Riley; Jeffrey C Crowley; Paul M Zabetakis
Journal:  J Extra Corpor Technol       Date:  2006-09

4.  Comparison of conventional extracorporeal circulation and minimal extracorporeal circulation with respect to microbubbles and microembolic signals.

Authors:  M Perthel; S Kseibi; F Sagebiel; A Alken; J Laas
Journal:  Perfusion       Date:  2005-10       Impact factor: 1.972

5.  Bubble tracking through computational fluid dynamics in arterial line filters for cardiopulmonary bypass.

Authors:  Gianfranco Beniamino Fiore; Umberto Morbiducci; Raffaele Ponzini; Alberto Redaelli
Journal:  ASAIO J       Date:  2009 Sep-Oct       Impact factor: 2.872

6.  Cerebral microemboli during cardiopulmonary bypass: increased emboli during perfusionist interventions.

Authors:  R L Taylor; M A Borger; R D Weisel; L Fedorko; C M Feindel
Journal:  Ann Thorac Surg       Date:  1999-07       Impact factor: 4.330

7.  Arterial line filters ranked for gaseous micro-emboli separation performance: an in vitro study.

Authors:  Jeffrey B Riley
Journal:  J Extra Corpor Technol       Date:  2008-03

Review 8.  Microemboli detection on extracorporeal bypass circuits.

Authors:  J E Lynch; J B Riley
Journal:  Perfusion       Date:  2008-01       Impact factor: 1.972

9.  Evaluation of air handling in a new generation neonatal oxygenator with integral arterial filter.

Authors:  D Gomez; T J Preston; V F Olshove; A B Phillips; M E Galantowicz
Journal:  Perfusion       Date:  2009-03       Impact factor: 1.972

10.  Post-arterial filter gaseous microemboli activity of five integral cardiotomy reservoirs during venting: an in vitro study.

Authors:  Gerard J Myers; Cheri Voorhees; Rob Haynes; Bob Eke
Journal:  J Extra Corpor Technol       Date:  2009-03
View more
  7 in total

1.  Working toward Best Practice: Microbubble Filtration and Patient Safety During Extracorporeal Circulation.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2015-06

2.  Air Transmission Comparison of the Affinity Fusion Oxygenator with an Integrated Arterial Filter to the Affinity NT Oxygenator with a Separate Arterial Filter.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2014-09

3.  Sequential Blood Filtration for Extracorporeal Circulation: Initial Results from a Proof-of-Concept Prototype.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2014-09

4.  Application of Micropore Filter Technology: Exploring the Blood Flow Path in Arterial-Line Filters and Its Effect on Bubble Trapping Functions.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2017-03

5.  Technology Use and Patient Safety: At the Intersection of Change.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2016-12

6.  Can the oxygenator screen filter reduce gaseous microemboli?

Authors:  Daniel Johagen; Micael Appelblad; Staffan Svenmarker
Journal:  J Extra Corpor Technol       Date:  2014-03

7.  To Purge or Not to Purge.

Authors:  Juan D V Hugo; Alexander Yeung; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2020-03
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.