Sequential Blood Filtration for Extracorporeal Circulation: Initial Results from a Proof-of-Concept Prototype.

Daniel P Herbst.   

Abstract

Micropore filters are used during extracorporeal circulation to prevent gaseous and solid particles from entering the patient's systemic circulation. Although these devices improve patient safety, limitations in current designs have prompted the development of a new concept in micropore filtration. A prototype of the new design was made using 40-μm filter screens and compared against four commercially available filters for performance in pressure loss and gross air handling. Pre- and postfilter bubble counts for 5- and 10-mL bolus injections in an ex vivo test circuit were recorded using a Doppler ultrasound bubble counter. Statistical analysis of results for bubble volume reduction between test filters was performed with one-way repeated-measures analysis of variance using Bonferroni post hoc tests. Changes in filter performance with changes in microbubble load were also assessed with dependent t tests using the 5- and 10-mL bolus injections as the paired sample for each filter. Significance was set at p < .05. All filters in the test group were comparable in pressure loss performance, showing a range of 26-33 mmHg at a flow rate of 6 L/min. In gross air-handling studies, the prototype showed improved bubble volume reduction, reaching statistical significance with three of the four commercial filters. All test filters showed decreased performance in bubble volume reduction when the microbubble load was increased. Findings from this research support the underpinning theories of a sequential arterial-line filter design and suggest that improvements in microbubble filtration may be possible using this technique.

Entities:  

Mesh:

Year:  2014        PMID: 26357790      PMCID: PMC4566833     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  16 in total

1.  In-vitro quantification of gaseous microemboli in two extracorporeal life support circuits.

Authors:  Jeffrey Burnside; Daniel Gomez; Thomas J Preston; Vincent F Olshove; Alistair Phillips
Journal:  J Extra Corpor Technol       Date:  2011-09

2.  In vitro evaluation of gaseous microemboli handling of cardiopulmonary bypass circuits with and without integrated arterial line filters.

Authors:  Saifei Liu; Richard F Newland; Phillip J Tully; Sigrid C Tuble; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2011-09

3.  Pre-clinical laboratory evaluation of the new 'AF' arterial line filter range.

Authors:  Gemma Yarham; John Mulholland
Journal:  Perfusion       Date:  2010-07       Impact factor: 1.972

4.  In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.

Authors:  Timothy A Dickinson; Jeffrey B Riley; Jeffrey C Crowley; Paul M Zabetakis
Journal:  J Extra Corpor Technol       Date:  2006-09

5.  The effects of pressure on gases in solution: possible insights to improve microbubble filtration for extracorporeal circulation.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2013-06

6.  Development of a new arterial-line filter design using computational fluid dynamics analysis.

Authors:  Daniel P Herbst; Hani K Najm
Journal:  J Extra Corpor Technol       Date:  2012-09

7.  Extracorporeal bubbles: a word of caution.

Authors:  Filip M J J De Somer; Maria R Vetrano; Jeroen P A J Van Beeck; Guido J Van Nooten
Journal:  Interact Cardiovasc Thorac Surg       Date:  2010-03-02

8.  Arterial line filters ranked for gaseous micro-emboli separation performance: an in vitro study.

Authors:  Jeffrey B Riley
Journal:  J Extra Corpor Technol       Date:  2008-03

9.  Microemboli in our bypass circuits: a contemporary audit.

Authors:  Timothy W Willcox; Simon J Mitchell
Journal:  J Extra Corpor Technol       Date:  2009-12

10.  Clinical gaseous microemboli assessment of an oxygenator with integral arterial filter in the pediatric population.

Authors:  Thomas J Preston; Daniel Gomez; Vincent F Olshove; Alistair Phillips; Mark Galantowicz
Journal:  J Extra Corpor Technol       Date:  2009-12
View more
  2 in total

1.  Working toward Best Practice: Microbubble Filtration and Patient Safety During Extracorporeal Circulation.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2015-06

2.  Application of Micropore Filter Technology: Exploring the Blood Flow Path in Arterial-Line Filters and Its Effect on Bubble Trapping Functions.

Authors:  Daniel P Herbst
Journal:  J Extra Corpor Technol       Date:  2017-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.