A collimator optimization method for quantitative imaging: application to Y-90 bremsstrahlung SPECT.

Xing Rong1, Eric C Frey.   

Abstract

PURPOSE: Post-therapy quantitative 90Y bremsstrahlung single photon emission computed tomography (SPECT) has shown great potential to provide reliable activity estimates, which are essential for dose verification. Typically 90Y imaging is performed with high- or medium-energy collimators. However, the energy spectrum of 90Y bremsstrahlung photons is substantially different than typical for these collimators. In addition, dosimetry requires quantitative images, and collimators are not typically optimized for such tasks. Optimizing a collimator for 90Y imaging is both novel and potentially important. Conventional optimization methods are not appropriate for 90Y bremsstrahlung photons, which have a continuous and broad energy distribution. In this work, the authors developed a parallel-hole collimator optimization method for quantitative tasks that is particularly applicable to radionuclides with complex emission energy spectra. The authors applied the proposed method to develop an optimal collimator for quantitative 90Y bremsstrahlung SPECT in the context of microsphere radioembolization.
METHODS: To account for the effects of the collimator on both the bias and the variance of the activity estimates, the authors used the root mean squared error (RMSE) of the volume of interest activity estimates as the figure of merit (FOM). In the FOM, the bias due to the null space of the image formation process was taken in account. The RMSE was weighted by the inverse mass to reflect the application to dosimetry; for a different application, more relevant weighting could easily be adopted. The authors proposed a parameterization for the collimator that facilitates the incorporation of the important factors (geometric sensitivity, geometric resolution, and septal penetration fraction) determining collimator performance, while keeping the number of free parameters describing the collimator small (i.e., two parameters). To make the optimization results for quantitative 90Y bremsstrahlung SPECT more general, the authors simulated multiple tumors of various sizes in the liver. The authors realistically simulated human anatomy using a digital phantom and the image formation process using a previously validated and computationally efficient method for modeling the image-degrading effects including object scatter, attenuation, and the full collimator-detector response (CDR). The scatter kernels and CDR function tables used in the modeling method were generated using a previously validated Monte Carlo simulation code.
RESULTS: The hole length, hole diameter, and septal thickness of the obtained optimal collimator were 84, 3.5, and 1.4 mm, respectively. Compared to a commercial high-energy general-purpose collimator, the optimal collimator improved the resolution and FOM by 27% and 18%, respectively.
CONCLUSIONS: The proposed collimator optimization method may be useful for improving quantitative SPECT imaging for radionuclides with complex energy spectra. The obtained optimal collimator provided a substantial improvement in quantitative performance for the microsphere radioembolization task considered.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23927349      PMCID: PMC3732303          DOI: 10.1118/1.4813297

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS.

Authors:  H O ANGER
Journal:  J Nucl Med       Date:  1964-07       Impact factor: 10.057

Review 2.  Dosimetry of internal emitters.

Authors:  George Sgouros
Journal:  J Nucl Med       Date:  2005-01       Impact factor: 10.057

3.  Accelerated image reconstruction using ordered subsets of projection data.

Authors:  H M Hudson; R S Larkin
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

4.  Development and validation of a Monte Carlo simulation of photon transport in an Anger camera.

Authors:  D J De Vries; S C Moore; R E Zimmerman; S P Mueller; B Friedland; R C Lanza
Journal:  IEEE Trans Med Imaging       Date:  1990       Impact factor: 10.048

5.  A method for energy window optimization for quantitative tasks that includes the effects of model-mismatch on bias: application to Y-90 bremsstrahlung SPECT imaging.

Authors:  Xing Rong; Yong Du; Eric C Frey
Journal:  Phys Med Biol       Date:  2012-05-23       Impact factor: 3.609

6.  Development and evaluation of an improved quantitative (90)Y bremsstrahlung SPECT method.

Authors:  Xing Rong; Yong Du; Michael Ljungberg; Erwann Rault; Stefaan Vandenberghe; Eric C Frey
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

7.  Recommendations of the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for 90Y microsphere brachytherapy in the treatment of hepatic malignancies.

Authors:  William A Dezarn; Jeffery T Cessna; Larry A DeWerd; Wenzheng Feng; Vanessa L Gates; James Halama; Andrew S Kennedy; Subir Nag; Mehrdad Sarfaraz; Varun Sehgal; Reed Selwyn; Michael G Stabin; Bruce R Thomadsen; Lawrence E Williams; Riad Salem
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

8.  Letter: Parallel-hole collimator design.

Authors:  M S Gerber; D W Miller
Journal:  J Nucl Med       Date:  1974-08       Impact factor: 10.057

9.  EM reconstruction algorithms for emission and transmission tomography.

Authors:  K Lange; R Carson
Journal:  J Comput Assist Tomogr       Date:  1984-04       Impact factor: 1.826

10.  4D XCAT phantom for multimodality imaging research.

Authors:  W P Segars; G Sturgeon; S Mendonca; Jason Grimes; B M W Tsui
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

View more
  9 in total

Review 1.  Quantitative Imaging of Alpha-Emitting Therapeutic Radiopharmaceuticals.

Authors:  Youngho Seo
Journal:  Nucl Med Mol Imaging       Date:  2019-02-18

2.  Repeatability of Radiotracer Uptake in Normal Abdominal Organs with ¹¹¹In-Pentetreotide Quantitative SPECT/CT.

Authors:  Steven P Rowe; Esther Vicente; Nadège Anizan; Hao Wang; Jeffrey P Leal; Martin A Lodge; Eric C Frey; Richard L Wahl
Journal:  J Nucl Med       Date:  2015-05-14       Impact factor: 10.057

3.  Practical reconstruction protocol for quantitative (90)Y bremsstrahlung SPECT/CT.

Authors:  W Siman; J K Mikell; S C Kappadath
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

4.  Effect of ME Collimator Characteristic, Energy Window Width, and Reconstruction Algorithm Selection on Imaging Performance of Yttrium-90: Simulation Study.

Authors:  Payvand Taherparvar; Nazila Shahmari
Journal:  Nucl Med Mol Imaging       Date:  2019-11-13

Review 5.  Advances in SPECT for Optimizing the Liver Tumors Radioembolization Using Yttrium-90 Microspheres.

Authors:  Hoda Rezaei Roshan; Ahmadreza Azarm; Babak Mahmoudian; Jalil Pirayesh Islamian
Journal:  World J Nucl Med       Date:  2015 May-Aug

Review 6.  Theranostic Imaging of Yttrium-90.

Authors:  Chadwick L Wright; Jun Zhang; Michael F Tweedle; Michael V Knopp; Nathan C Hall
Journal:  Biomed Res Int       Date:  2015-05-28       Impact factor: 3.411

7.  Comparison of Image Quality of Different Radionuclides Technetium-99m, Samarium-153, and Iodine-123.

Authors:  Youssef Bouzekraoui; Farida Bentayeb; Hicham Asmi; Faustino Bonutti
Journal:  Indian J Nucl Med       Date:  2019 Jul-Sep

8.  Energy Window and Contrast Optimization for Single-photon Emission Computed Tomography Bremsstrahlung Imaging with Yttrium-90.

Authors:  Youssef Bouzekraoui; Farida Bentayeb; Hicham Asmi; Faustino Bonutti
Journal:  Indian J Nucl Med       Date:  2019 Apr-Jun

Review 9.  Quantitative Imaging for Targeted Radionuclide Therapy Dosimetry - Technical Review.

Authors:  Tiantian Li; Edwin C I Ao; Bieke Lambert; Boudewijn Brans; Stefaan Vandenberghe; Greta S P Mok
Journal:  Theranostics       Date:  2017-10-13       Impact factor: 11.556

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.