A method for energy window optimization for quantitative tasks that includes the effects of model-mismatch on bias: application to Y-90 bremsstrahlung SPECT imaging.

Xing Rong1, Yong Du, Eric C Frey.   

Abstract

Quantitative Yttrium-90 ((90)Y) bremsstrahlung single photon emission computed tomography (SPECT) imaging has shown great potential to provide reliable estimates of (90)Y activity distribution for targeted radionuclide therapy dosimetry applications. One factor that potentially affects the reliability of the activity estimates is the choice of the acquisition energy window. In contrast to imaging conventional gamma photon emitters where the acquisition energy windows are usually placed around photopeaks, there has been great variation in the choice of the acquisition energy window for (90)Y imaging due to the continuous and broad energy distribution of the bremsstrahlung photons. In quantitative imaging of conventional gamma photon emitters, previous methods for optimizing the acquisition energy window assumed unbiased estimators and used the variance in the estimates as a figure of merit (FOM). However, for situations, such as (90)Y imaging, where there are errors in the modeling of the image formation process used in the reconstruction there will be bias in the activity estimates. In (90)Y bremsstrahlung imaging this will be especially important due to the high levels of scatter, multiple scatter, and collimator septal penetration and scatter. Thus variance will not be a complete measure of reliability of the estimates and thus is not a complete FOM. To address this, we first aimed to develop a new method to optimize the energy window that accounts for both the bias due to model-mismatch and the variance of the activity estimates. We applied this method to optimize the acquisition energy window for quantitative (90)Y bremsstrahlung SPECT imaging in microsphere brachytherapy. Since absorbed dose is defined as the absorbed energy from the radiation per unit mass of tissues in this new method we proposed a mass-weighted root mean squared error of the volume of interest (VOI) activity estimates as the FOM. To calculate this FOM, two analytical expressions were derived for calculating the bias due to model-mismatch and the variance of the VOI activity estimates, respectively. To obtain the optimal acquisition energy window for general situations of interest in clinical (90)Y microsphere imaging, we generated phantoms with multiple tumors of various sizes and various tumor-to-normal activity concentration ratios using a digital phantom that realistically simulates human anatomy, simulated (90)Y microsphere imaging with a clinical SPECT system and typical imaging parameters using a previously validated Monte Carlo simulation code, and used a previously proposed method for modeling the image degrading effects in quantitative SPECT reconstruction. The obtained optimal acquisition energy window was 100-160 keV. The values of the proposed FOM were much larger than the FOM taking into account only the variance of the activity estimates, thus demonstrating in our experiment that the bias of the activity estimates due to model-mismatch was a more important factor than the variance in terms of limiting the reliability of activity estimates.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22617760      PMCID: PMC3739432          DOI: 10.1088/0031-9155/57/12/3711

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  13 in total

1.  Optimization of Ga-67 imaging for detection and estimation tasks: dependence of imaging performance on spectral acquisition parameters.

Authors:  Georges El Fakhri; Stephen C Moore; Marie Foley Kijewski
Journal:  Med Phys       Date:  2002-08       Impact factor: 4.071

2.  Effect of errors in the system matrix on maximum a posteriori image reconstruction.

Authors:  Jinyi Qi; Ronald H Huesman
Journal:  Phys Med Biol       Date:  2005-07-06       Impact factor: 3.609

3.  A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT.

Authors:  Bin He; Yong Du; Xiyun Song; W Paul Segars; Eric C Frey
Journal:  Phys Med Biol       Date:  2005-08-24       Impact factor: 3.609

4.  Development and evaluation of an improved quantitative (90)Y bremsstrahlung SPECT method.

Authors:  Xing Rong; Yong Du; Michael Ljungberg; Erwann Rault; Stefaan Vandenberghe; Eric C Frey
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

5.  Recommendations of the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for 90Y microsphere brachytherapy in the treatment of hepatic malignancies.

Authors:  William A Dezarn; Jeffery T Cessna; Larry A DeWerd; Wenzheng Feng; Vanessa L Gates; James Halama; Andrew S Kennedy; Subir Nag; Mehrdad Sarfaraz; Varun Sehgal; Reed Selwyn; Michael G Stabin; Bruce R Thomadsen; Lawrence E Williams; Riad Salem
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

6.  Analysis of the reconstructibility and noise properties of scattered photons in 99mTc SPECT.

Authors:  D J Kadrmas; E C Frey; B M Tsui
Journal:  Phys Med Biol       Date:  1997-12       Impact factor: 3.609

7.  A Monte Carlo program for the simulation of scintillation camera characteristics.

Authors:  M Ljungberg; S E Strand
Journal:  Comput Methods Programs Biomed       Date:  1989-08       Impact factor: 5.428

8.  4D XCAT phantom for multimodality imaging research.

Authors:  W P Segars; G Sturgeon; S Mendonca; Jason Grimes; B M W Tsui
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

Review 9.  Three-dimensional imaging-based radiobiological dosimetry.

Authors:  George Sgouros; Eric Frey; Richard Wahl; Bin He; Andrew Prideaux; Robert Hobbs
Journal:  Semin Nucl Med       Date:  2008-09       Impact factor: 4.446

10.  Planar gamma camera imaging and quantitation of yttrium-90 bremsstrahlung.

Authors:  S Shen; G L DeNardo; A Yuan; D A DeNardo; S J DeNardo
Journal:  J Nucl Med       Date:  1994-08       Impact factor: 10.057

View more
  11 in total

1.  A collimator optimization method for quantitative imaging: application to Y-90 bremsstrahlung SPECT.

Authors:  Xing Rong; Eric C Frey
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

2.  Cerenkov Luminescence Imaging for Radiation Dose Calculation of a ⁹⁰Y-Labeled Gastrin-Releasing Peptide Receptor Antagonist.

Authors:  Christian Lohrmann; Hanwen Zhang; Daniel L J Thorek; Pooja Desai; Pat B Zanzonico; Joseph O'Donoghue; Christopher P Irwin; Thomas Reiner; Jan Grimm; Wolfgang A Weber
Journal:  J Nucl Med       Date:  2015-04-03       Impact factor: 10.057

Review 3.  Quantitative Imaging of Alpha-Emitting Therapeutic Radiopharmaceuticals.

Authors:  Youngho Seo
Journal:  Nucl Med Mol Imaging       Date:  2019-02-18

Review 4.  The role of SPECT/CT in radioembolization of liver tumours.

Authors:  Hojjat Ahmadzadehfar; Heying Duan; Alexander R Haug; Stephan Walrand; Martha Hoffmann
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-01-18       Impact factor: 9.236

Review 5.  Post Yttrium-90 Imaging.

Authors:  Mitchell Rice; Matthew Krosin; Paul Haste
Journal:  Semin Intervent Radiol       Date:  2021-10-07       Impact factor: 1.780

6.  Optimization of energy window for 90Y bremsstrahlung SPECT imaging for detection tasks using the ideal observer with model-mismatch.

Authors:  Xing Rong; Michael Ghaly; Eric C Frey
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

7.  Effect of ME Collimator Characteristic, Energy Window Width, and Reconstruction Algorithm Selection on Imaging Performance of Yttrium-90: Simulation Study.

Authors:  Payvand Taherparvar; Nazila Shahmari
Journal:  Nucl Med Mol Imaging       Date:  2019-11-13

Review 8.  Advances in SPECT for Optimizing the Liver Tumors Radioembolization Using Yttrium-90 Microspheres.

Authors:  Hoda Rezaei Roshan; Ahmadreza Azarm; Babak Mahmoudian; Jalil Pirayesh Islamian
Journal:  World J Nucl Med       Date:  2015 May-Aug

Review 9.  Theranostic Imaging of Yttrium-90.

Authors:  Chadwick L Wright; Jun Zhang; Michael F Tweedle; Michael V Knopp; Nathan C Hall
Journal:  Biomed Res Int       Date:  2015-05-28       Impact factor: 3.411

Review 10.  Combination Radioimmunotherapy Approaches and Quantification of Immuno-PET.

Authors:  Jin Su Kim
Journal:  Nucl Med Mol Imaging       Date:  2016-01-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.