Youssef Bouzekraoui1, Farida Bentayeb1, Hicham Asmi1, Faustino Bonutti2.
Abstract
PURPOSE: In yttrium-90 (Y-90) single-photon emission computed tomography (SPECT) imaging, the choice of the acquisition energy window is not trivial, due to the continuous and broad energy distribution of the bremsstrahlung photons. In this work, we investigate the effects of the energy windows on the image contrast to noise ratio (CNR), in order to select the optimal energy window for Y-90 imaging.Entities:
Keywords: Bremsstrahlung; CNR; Jaszczak phantom; SIMIND; yttrium-90 single-photon emission computed tomography imaging
Year: 2019 PMID: 31040523 PMCID: PMC6481209 DOI: 10.4103/ijnm.IJNM_150_18
Source DB: PubMed Journal: Indian J Nucl Med ISSN: 0974-0244
Collimator specifications
| Collimator | Diameter (cm) | Septa (cm) | Length (cm) | Hole shape | Col type |
|---|---|---|---|---|---|
| HE | 0.400 | 0.200 | 5.970 | Hexagonal | Parallel hole |
HE: High energy
Size and central position for the sub-windows of the large window-set
| 1 | 2 | 3 | 4 | 5 | |
|---|---|---|---|---|---|
| Subwindow (keV) | 30–100 | 100–170 | 170–240 | 240–310 | 210–380 |
| Center (keV) | 65 | 135 | 205 | 275 | 345 |
Size and central position for the sub-windows of the narrow window-set
| 1 | 2 | 3 | 4 | 5 | |
|---|---|---|---|---|---|
| Subwindow (keV) | 60–90 | 90–120 | 120–150 | 150–180 | 180–210 |
| Center (keV) | 75 | 105 | 135 | 165 | 195 |
Figure 1Monte Carlo simulated spectrum for yttrium-90
Figure 2Contributions of geometric, penetration, scatter, and X-rays components with energy windows
Figure 3Contrast of the six spheres with energy window
Figure 4Images of simulated Jaszczak phantom with a high-energy collimator using large energy windows (above) and narrow energy windows (below)
Figure 5CNR of the six spheres
Figure 6CNR of the six spheres with energy windows