Susceptibility of interstitial continuous glucose monitor performance to sleeping position.

Brett D Mensh1, Natalie A Wisniewski, Brian M Neil, Daniel R Burnett.   

Abstract

BACKGROUND: Developing a round-the-clock artificial pancreas requires accurate and stable continuous glucose monitoring. The most widely used continuous glucose monitors (CGMs) are percutaneous, with the sensor residing in the interstitial space. Inaccuracies in percutaneous CGM readings during periods of lying on the devices (e.g., in various sleeping positions) have been anecdotally reported but not systematically studied.
METHODS: In order to assess the impact of sleep and sleep position on CGM performance, we conducted a study in human subjects in which we measured the variability of interstitial CGM data at night as a function of sleeping position. Commercially available sensors were placed for 4 days in the abdominal subcutaneous tissue in healthy, nondiabetic volunteers (four sensors per person, two per side). Nocturnal sleeping position was determined from video recordings and correlated to sensor data.
RESULTS: We observed that, although the median of the four sensor readings was typically 70-110 mg/dl during sleep, individual sensors intermittently exhibited aberrant glucose readings (>25 mg/dl away from median) and that these aberrant readings were strongly correlated with subjects lying on the sensors. We expected and observed that most of these aberrant sleep-position-related CGM readings were sudden decreases in reported glucose values, presumably due to local blood-flow decreases caused by tissue compression. Curiously, in rare cases, the aberrant CGM readings were elevated values.
CONCLUSIONS: These findings highlight limitations in our understanding of interstitial fluid physiology in the subcutaneous space and have significant implications for the utilization of sensors in the construction of an artificial pancreas.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23911167      PMCID: PMC3879750          DOI: 10.1177/193229681300700408

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  32 in total

Review 1.  Real-time continuous glucose monitoring.

Authors:  Bruce Buckingham; Kimberly Caswell; Darrell M Wilson
Journal:  Curr Opin Endocrinol Diabetes Obes       Date:  2007-08       Impact factor: 3.243

2.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

3.  Local heat increases blood flow and oxygen tension in wounds.

Authors:  J M Rabkin; T K Hunt
Journal:  Arch Surg       Date:  1987-02

4.  Multinational study of subcutaneous model-predictive closed-loop control in type 1 diabetes mellitus: summary of the results.

Authors:  Boris Kovatchev; Claudio Cobelli; Eric Renard; Stacey Anderson; Marc Breton; Stephen Patek; William Clarke; Daniela Bruttomesso; Alberto Maran; Silvana Costa; Angelo Avogaro; Chiara Dalla Man; Andrea Facchinetti; Lalo Magni; Giuseppe De Nicolao; Jerome Place; Anne Farret
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

5.  Nocturnal hypoglycaemia in Type 1 diabetic patients, assessed with continuous glucose monitoring: frequency, duration and associations.

Authors:  I M E Wentholt; A Maran; N Masurel; R J Heine; J B L Hoekstra; J H DeVries
Journal:  Diabet Med       Date:  2007-03-22       Impact factor: 4.359

6.  Lack of accuracy of continuous glucose sensors in healthy, nondiabetic children: results of the Diabetes Research in Children Network (DirecNet) accuracy study.

Authors:  Nelly Mauras; Roy W Beck; Katrina J Ruedy; Craig Kollman; William V Tamborlane; H Peter Chase; Bruce A Buckingham; Eva Tsalikian; Stuart Weinzimer; Andrea D Booth; Dongyuan Xing
Journal:  J Pediatr       Date:  2004-06       Impact factor: 4.406

7.  Decreased epinephrine responses to hypoglycemia during sleep.

Authors:  T W Jones; P Porter; R S Sherwin; E A Davis; P O'Leary; F Frazer; G Byrne; S Stick; W V Tamborlane
Journal:  N Engl J Med       Date:  1998-06-04       Impact factor: 91.245

8.  Nocturnal hypoglycaemias in type 1 diabetic patients: what can we learn with continuous glucose monitoring?

Authors:  L Guillod; S Comte-Perret; D Monbaron; R C Gaillard; J Ruiz
Journal:  Diabetes Metab       Date:  2007-07-24       Impact factor: 6.041

9.  Continuous glucose monitoring and intensive treatment of type 1 diabetes.

Authors:  William V Tamborlane; Roy W Beck; Bruce W Bode; Bruce Buckingham; H Peter Chase; Robert Clemons; Rosanna Fiallo-Scharer; Larry A Fox; Lisa K Gilliam; Irl B Hirsch; Elbert S Huang; Craig Kollman; Aaron J Kowalski; Lori Laffel; Jean M Lawrence; Joyce Lee; Nelly Mauras; Michael O'Grady; Katrina J Ruedy; Michael Tansey; Eva Tsalikian; Stuart Weinzimer; Darrell M Wilson; Howard Wolpert; Tim Wysocki; Dongyuan Xing
Journal:  N Engl J Med       Date:  2008-09-08       Impact factor: 91.245

10.  Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes.

Authors:  Kilty McGowan; William Thomas; Antoinette Moran
Journal:  Diabetes Care       Date:  2002-09       Impact factor: 19.112

View more
  23 in total

1.  Comparison of Insulins Glargine and Degludec in Diabetic Rhesus Macaques (Macaca mulatta) with CGM Devices.

Authors:  Samantha C Puglisi; Alexis L Mackiewicz; Amir Ardeshir; Laura M Garzel; Kari L Christe
Journal:  Comp Med       Date:  2021-05-25       Impact factor: 0.982

2.  A novel method to detect pressure-induced sensor attenuations (PISA) in an artificial pancreas.

Authors:  Nihat Baysal; Fraser Cameron; Bruce A Buckingham; Darrell M Wilson; H Peter Chase; David M Maahs; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2014-10-14

3.  Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology.

Authors:  Udo Hoss; Erwin Satrya Budiman
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

4.  Continuous glucose monitoring in 2013.

Authors:  Tadej Battelino; Bruce W Bode
Journal:  Diabetes Technol Ther       Date:  2014-02       Impact factor: 6.118

Review 5.  Update on Clinical Utility of Continuous Glucose Monitoring in Type 1 Diabetes.

Authors:  Nalani Haviland; John Walsh; Ruth Roberts; Timothy S Bailey
Journal:  Curr Diab Rep       Date:  2016-11       Impact factor: 4.810

Review 6.  Fault detection and safety in closed-loop artificial pancreas systems.

Authors:  B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2014-07-21

Review 7.  Practical Considerations on the Use of Continuous Glucose Monitoring in Pediatrics and Older Adults and Nonadjunctive Use.

Authors:  Gregory P Forlenza; Nicholas B Argento; Lori M Laffel
Journal:  Diabetes Technol Ther       Date:  2017-06       Impact factor: 6.118

8.  Continuous Glucose Monitors and Automated Insulin Dosing Systems in the Hospital Consensus Guideline.

Authors:  Rodolfo J Galindo; Guillermo E Umpierrez; Robert J Rushakoff; Ananda Basu; Suzanne Lohnes; James H Nichols; Elias K Spanakis; Juan Espinoza; Nadine E Palermo; Dessa Garnett Awadjie; Leigh Bak; Bruce Buckingham; Curtiss B Cook; Guido Freckmann; Lutz Heinemann; Roman Hovorka; Nestoras Mathioudakis; Tonya Newman; David N O'Neal; Michaela Rickert; David B Sacks; Jane Jeffrie Seley; Amisha Wallia; Trisha Shang; Jennifer Y Zhang; Julia Han; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2020-09-28

9.  Glycemic Patterns Are Distinct in Post-Bariatric Hypoglycemia After Gastric Bypass (PBH-RYGB).

Authors:  Daniel Lee; Jonathan M Dreyfuss; Amanda Sheehan; Alexa Puleio; Christopher M Mulla; Mary Elizabeth Patti
Journal:  J Clin Endocrinol Metab       Date:  2021-07-13       Impact factor: 5.958

10.  Evaluation of Reference Metrics for Continuous Glucose Monitoring in Persons Without Diabetes and Prediabetes.

Authors:  Sheyda Sofizadeh; Anders Pehrsson; Arndís F Ólafsdóttir; Marcus Lind
Journal:  J Diabetes Sci Technol       Date:  2020-10-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.