Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes.

Kilty McGowan1, William Thomas, Antoinette Moran.   

Abstract

OBJECTIVE: The Medtronic MiniMed Continuous Glucose Monitoring System (CGMS) is designed to continuously monitor interstitial fluid glucose levels within a range of 40-400 mg/dl. It is considered an important tool for overnight glucose monitoring. The goal of this study was to determine the accuracy of this system in individuals with tightly controlled diabetes. RESEARCH DESIGN AND METHODS: Seven adolescents and young adults with HbA(1c) levels 6.6 +/- 0.6% (range 5.7-7.1) were admitted to the Clinical Research Center. Simultaneous glucose measurements obtained by glucose analyzer, Accu-Check Advantage meter, and CGMS were compared. The analyzer levels were considered the standard.
RESULTS: The CGMS results were lower than analyzer readings in 74% of simultaneous pairs of tests performed during the 24-h period; the average correlation was 0.76. There was a trend for the poorest correlation to occur in patients with the narrowest range in daily glucose levels. When the lowest CGMS reading of the night was compared with the simultaneous analyzer reading, the CGMS level was lower in all cases by an average of 38 +/- 15%. In six of seven patients, the discrepancy was believed to be clinically significant; in at least four patients, overnight glucose levels reported by CGMS were falsely low, in a range that might have resulted in inappropriate reduction of overnight insulin dose.
CONCLUSIONS: CGMS reports of asymptomatic nighttime hypoglycemia may be spurious and should be interpreted with caution in patients with tightly controlled diabetes.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12196417     DOI: 10.2337/diacare.25.9.1499

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  23 in total

Review 1.  Evolving approaches to intensive insulin therapy in type 1 diabetes: multiple daily injections, insulin pumps and new methods of monitoring.

Authors:  Elizabeth Stephens; Matthew Riddle
Journal:  Rev Endocr Metab Disord       Date:  2003-12       Impact factor: 6.514

2.  Do different glucose levels at calibration influence accuracy of continuous glucose monitoring readings in vitro?

Authors:  Katherine E Iscoe; Raymond J Davey; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

4.  Evaluation of factors affecting CGMS calibration.

Authors:  Bruce A Buckingham; Craig Kollman; Roy Beck; Andrea Kalajian; Rosanna Fiallo-Scharer; Michael J Tansey; Larry A Fox; Darrell M Wilson; Stuart A Weinzimer; Katrina J Ruedy; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

5.  Clinical Use of Professional Continuous Glucose Monitoring.

Authors:  Eugene E Wright; James R Gavin
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

6.  The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes.

Authors:  Darrell M Wilson; Roy W Beck; William V Tamborlane; Mariya J Dontchev; Craig Kollman; Peter Chase; Larry A Fox; Katrina J Ruedy; Eva Tsalikian; Stuart A Weinzimer
Journal:  Diabetes Care       Date:  2007-01       Impact factor: 19.112

Review 7.  Exercise and glucose metabolism in persons with diabetes mellitus: perspectives on the role for continuous glucose monitoring.

Authors:  Michael Riddell; Bruce A Perkins
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

8.  Lack of accuracy of continuous glucose sensors in healthy, nondiabetic children: results of the Diabetes Research in Children Network (DirecNet) accuracy study.

Authors:  Nelly Mauras; Roy W Beck; Katrina J Ruedy; Craig Kollman; William V Tamborlane; H Peter Chase; Bruce A Buckingham; Eva Tsalikian; Stuart Weinzimer; Andrea D Booth; Dongyuan Xing
Journal:  J Pediatr       Date:  2004-06       Impact factor: 4.406

9.  The accuracy of the CGMS in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

10.  Susceptibility of interstitial continuous glucose monitor performance to sleeping position.

Authors:  Brett D Mensh; Natalie A Wisniewski; Brian M Neil; Daniel R Burnett
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.