| Literature DB >> 23857652 |
Paola Gandini1, Linda Orsi, Maria Francesca Sfondrini, Andrea Scribante.
Abstract
UNLABELLED: Self-ligating brackets engage the wire by means of a slide mechanism. Forces that have to be applied to open and close the sliding mechanism of brackets are still unknown.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23857652 PMCID: PMC3881911 DOI: 10.1590/1679-7757201300063
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
Figure 1Testing apparatus for opening force measurement (A: position before test; B: position after test)
Figure 2Testing apparatus for closure force measurement (A: position before test; B: position after test)
Descriptive statistics of the different brackets tested (N). All groups showed normal distributions
| 1 | Carriere LX | Upper central incisor | Opening | 3.34 | 0.12 | 3.2 | 3.3 | 3.7 | A |
| 2 | Carriere LX | Upper central incisor | Closure | 4.42 | 0.11 | 4.29 | 4.4 | 4.71 | B |
| 3 | F1000 | Upper central incisor | Opening | 5.46 | 0.07 | 5.3 | 5.5 | 5.6 | C |
| 4 | F1000 | Upper central incisor | Closure | 3.87 | 0.31 | 3.32 | 3.89 | 4.53 | D |
| 5 | Damon Q | Upper central incisor | Opening | 4.89 | 0.35 | 4.4 | 4.8 | 5.5 | E |
| 6 | Damon Q | Upper central incisor | Closure | 4.68 | 0.11 | 4.45 | 4.68 | 4.87 | E |
| 7 | Carriere LX | Lower cental incisor | Opening | 1.41 | 0.18 | 1.1 | 1.4 | 1.8 | F |
| 8 | Carriere LX | Lower cental incisor | Closure | 2.28 | 0.1 | 2.12 | 2.28 | 2.46 | G |
| 9 | F1000 | Lower cental incisor | Opening | 3.28 | 0.38 | 2.7 | 3.3 | 3.9 | A |
| 10 | F1000 | Lower cental incisor | Closure | 1.8 | 0.32 | 1.57 | 1.69 | 2.85 | H |
| 11 | Damon Q | Lower cental incisor | Opening | 2.81 | 0.09 | 2.7 | 2.8 | 3 | I |
| 12 | Damon Q | Lower cental incisor | Closure | 2.49 | 0.16 | 2.02 | 2.56 | 2.6 | G |
*Tukey grouping. Means with the same letters are not significantly different