Spatial and rotational quality assurance of 6DOF patient tracking systems.

Andrew H Belcher1, Xinmin Liu1, Zachary Grelewicz1, Rodney D Wiersma1.   

Abstract

PURPOSE: External tracking systems used for patient positioning and motion monitoring during radiotherapy are now capable of detecting both translations and rotations. In this work, the authors develop a novel technique to evaluate the 6 degree of freedom 6(DOF) (translations and rotations) performance of external motion tracking systems. The authors apply this methodology to an infrared marker tracking system and two 3D optical surface mapping systems in a common tumor 6DOF workspace.
METHODS: An in-house designed and built 6DOF parallel kinematics robotic motion phantom was used to perform motions with sub-millimeter and subdegree accuracy in a 6DOF workspace. An infrared marker tracking system was first used to validate a calibration algorithm which associates the motion phantom coordinate frame to the camera frame. The 6DOF positions of the mobile robotic system in this space were then tracked and recorded independently by an optical surface tracking system after a cranial phantom was rigidly fixed to the moveable platform of the robotic stage. The calibration methodology was first employed, followed by a comprehensive 6DOF trajectory evaluation, which spanned a full range of positions and orientations in a 20 × 20 × 16 mm and 5° × 5° × 5° workspace. The intended input motions were compared to the calibrated 6DOF measured points.
RESULTS: The technique found the accuracy of the infrared (IR) marker tracking system to have maximal root-mean square error (RMSE) values of 0.18, 0.25, 0.07 mm, 0.05°, 0.05°, and 0.09° in left-right (LR), superior-inferior (SI), anterior-posterior (AP), pitch, roll, and yaw, respectively, comparing the intended 6DOF position and the measured position by the IR camera. Similarly, the 6DOF RSME discrepancy for the HD optical surface tracker yielded maximal values of 0.46, 0.60, 0.54 mm, 0.06°, 0.11°, and 0.08° in LR, SI, AP, pitch, roll, and yaw, respectively, over the same 6DOF evaluative workspace. An earlier generation 3D optical surface tracking unit was observed to have worse tracking capabilities than both the IR camera unit and the newer 3D surface tracking system with maximal RMSE of 0.69, 0.74, 0.47 mm, 0.28°, 0.19°, and 0.18°, in LR, SI, AP, pitch, roll, and yaw, respectively, in the same 6DOF evaluation space.
CONCLUSIONS: The proposed technique was found to be effective at evaluating the performance of 6DOF patient tracking systems. All observed optical tracking systems were found to exhibit tracking capabilities at the sub-millimeter and subdegree level within a 6DOF workspace.

Entities:  

Mesh:

Year:  2016        PMID: 27277026      PMCID: PMC4866954          DOI: 10.1118/1.4948506

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  Initial clinical experience with frameless radiosurgery for patients with intracranial metastases.

Authors:  Reena Kamath; Timothy C Ryken; Sanford L Meeks; Edward C Pennington; Justine Ritchie; John M Buatti
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-04-01       Impact factor: 7.038

2.  Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).

Authors:  Nobutaka Mukumoto; Mitsuhiro Nakamura; Akira Sawada; Yasunobu Suzuki; Kunio Takahashi; Yuki Miyabe; Shuji Kaneko; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

3.  Magnetic resonance imaging-guided versus surrogate-based motion tracking in liver radiation therapy: a prospective comparative study.

Authors:  Chiara Paganelli; Matteo Seregni; Giovanni Fattori; Paul Summers; Massimo Bellomi; Guido Baroni; Marco Riboldi
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-03-15       Impact factor: 7.038

4.  The dosimetric impact of prostate rotations during electromagnetically guided external-beam radiation therapy.

Authors:  Hanan Amro; Daniel A Hamstra; Daniel L Mcshan; Howard Sandler; Karen Vineberg; Scott Hadley; Dale Litzenberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-05-02       Impact factor: 7.038

5.  Development of a 6DOF robotic motion phantom for radiation therapy.

Authors:  Andrew H Belcher; Xinmin Liu; Zachary Grelewicz; Erik Pearson; Rodney D Wiersma
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

6.  Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax.

Authors:  Philipp J Schöffel; Wolfgang Harms; Gabriele Sroka-Perez; Wolfgang Schlegel; Christian P Karger
Journal:  Phys Med Biol       Date:  2007-06-06       Impact factor: 3.609

7.  Intra-fraction setup variability: IR optical localization vs. X-ray imaging in a hypofractionated patient population.

Authors:  Maria Francesca Spadea; Barbara Tagaste; Marco Riboldi; Eleonora Preve; Daniela Alterio; Gaia Piperno; Cristina Garibaldi; Roberto Orecchia; Antonio Pedotti; Guido Baroni
Journal:  Radiat Oncol       Date:  2011-04-15       Impact factor: 3.481

8.  A comparative study between the imaging system and the optical tracking system in proton therapy at CNAO.

Authors:  Maxime Desplanques; Barbara Tagaste; Giulia Fontana; Andrea Pella; Marco Riboldi; Giovanni Fattori; Andrea Donno; Guido Baroni; Roberto Orecchia
Journal:  J Radiat Res       Date:  2013-07       Impact factor: 2.724

Review 9.  Image guidance in radiation therapy: techniques and applications.

Authors:  Shikha Goyal; Tejinder Kataria
Journal:  Radiol Res Pract       Date:  2014-12-17

10.  Dosimetric consequences of rotational setup errors with direct simulation in a treatment planning system for fractionated stereotactic radiotherapy.

Authors:  Jean L Peng; Chihray Liu; Yu Chen; Robert J Amdur; Kenneth Vanek; Jonathan G Li
Journal:  J Appl Clin Med Phys       Date:  2011-04-04       Impact factor: 2.102

View more
  5 in total

1.  Towards frameless maskless SRS through real-time 6DoF robotic motion compensation.

Authors:  Andrew H Belcher; Xinmin Liu; Steven Chmura; Kamil Yenice; Rodney D Wiersma
Journal:  Phys Med Biol       Date:  2017-11-13       Impact factor: 3.609

2.  Optimization based trajectory planning for real-time 6DoF robotic patient motion compensation systems.

Authors:  Xinmin Liu; Rodney D Wiersma
Journal:  PLoS One       Date:  2019-01-11       Impact factor: 3.240

3.  A phantom study to evaluate three different registration platform of 3D/3D, 2D/3D, and 3D surface match with 6D alignment for precise image-guided radiotherapy.

Authors:  Hsiang-Chi Kuo; Michael M Lovelock; Guang Li; Åse Ballangrud; Brian Wolthuis; Cesar Della Biancia; Margie A Hunt; Sean L Berry
Journal:  J Appl Clin Med Phys       Date:  2020-11-12       Impact factor: 2.102

Review 4.  Recent advanced in Surface Guided Radiation Therapy.

Authors:  P Freislederer; M Kügele; M Öllers; A Swinnen; T-O Sauer; C Bert; D Giantsoudi; S Corradini; V Batista
Journal:  Radiat Oncol       Date:  2020-07-31       Impact factor: 3.481

5.  A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy.

Authors:  Takehiro Shiinoki; Fumitake Fujii; Koya Fujimoto; Yuki Yuasa; Tatsuhiro Sera
Journal:  J Appl Clin Med Phys       Date:  2020-04-13       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.