Literature DB >> 23814563

Analysis of linear measurement accuracy obtained by cone beam computed tomography (CBCT-NewTom VG).

Mahkameh Moshfeghi1, Mohammad Amin Tavakoli, Ehsan Tavakoli Hosseini, Ali Tavakoli Hosseini, Iman Tavakoli Hosseini.   

Abstract

BACKGROUND: One of the major uses of cone beam computed tomography (CBCT) is presurgical implant planning. Linear measurement is used for the determination of the quantity of alveolar bone (height and width). Linear measurements are used in orthodontic analysis and definition of jaw tumor size. The objective of this study is to evaluate the accuracy of the linear measurement of CBCT (Newtom VG) in the axial and coronal planes, with two different voxel sizes.
MATERIALS AND METHODS: In this accuracy diagnostic study, 22 anatomic landmarks in four dry human skulls were marked by gutta-percha. Fifteen linear measurements were obtained using a digital caliper. These were considered to be the gold standard (real measurement). The skulls were scanned by CBCT (Newtom VG) at two settings: (a) Voxel size 0.3 mm (b) voxel size 0.15 mm High Resolution (HR). The radiographic distance measurements were made in the axial and coronal sections by three observers. The radiographic measurements were repeated two weeks later for evaluation of intraobserver reliability. SPSS software version 17 was used for data analysis. The level of significance was considered to be 5% (P ≤ 0.05).
RESULTS: The mean differences of real and radiographic measurements were -0.10±0.99 mm in the axial sections, -0.27±1.07 mm in the coronal sections, +0.14±1.44 mm in the axial (HR) sections, and 0.02±1.4 mm in the coronal (HR) sections. The intraclass correlation (ICC) for CBCT measurements in the axial sections was 0.9944, coronal sections 0.9941, axial (HR) sections 0.9935, and coronal (HR) sections 0.9937. The statistical analysis showed high interobserver and intraobserver reliability (P ≤ 0.05).
CONCLUSION: CBCT (Newtom VG) is highly accurate and reproducible in linear measurements in the axial and coronal image planes and in different areas of the maxillofacial region. According to the findings of the present study, a CBCT scan with a larger voxel size (0.3 mm in comparison to 0.15 mm) is recommended when the purpose of the CBCT scan is to measure linear distances. This will result in lower patient radiation dose and faster scan time.

Entities:  

Keywords:  Cone-beam computed tomography; implant; measurement; skull

Year:  2012        PMID: 23814563      PMCID: PMC3692201     

Source DB:  PubMed          Journal:  Dent Res J (Isfahan)        ISSN: 1735-3327


INTRODUCTION

The introduction of cone beam computed tomography (CBCT) dedicated to maxillofacial imaging introduces an innovation in maxillofacial imaging. Due to rapid volumetric image acquisition (as low as 18 seconds) from a single low radiation dose scan of the patient and the low mA, the effective dose with the CBCT technique is significantly lower than that achieved with other CT imaging methods. CBCT provides multiple planar images of both jaws by a single rotational scan.[1234] Applications of CBCT in maxillofacial imaging are increasing. Currently CBCT is commonly used for a variety of purposes in dentomaxillofacial imaging, implantology, periodontics, temporomandibular joint (TMJ) pathology, airway analysis, bone pathology, and endodontics.[56789] Cone Beam Computed Tomography can provide submillimeter spatial resolution for images of the craniofacial complex, with scanning time comparable to panoramic radiography. The cone-beam technique uses rotational scanning of an X-ray source, reciprocating an X-ray detector around the patient head. CT/CBCT images are displayed as a matrix of individual blocks called voxels (volume element). CBCT can perform imaging of maxillofacial structures with different voxel sizes. The voxel size in CBCT may be as low as 0.125 mm, smaller than that achieved with conventional CT units. Smaller voxel size provides better image resolution and requires higher radiation dose. The voxel in CBCT is isotropic (uniform in all directions). Isotropic voxels in CBCT enable maintenance of the image quality in all three orthogonal planes (axial, sagittal, and coronal). Several CBCT systems are commercially available.[101112] CBCT software provide tools to measure distances, angles, zoom, invert the gray scale, adjust contrast, and gamma changes.[13] One of the major uses of CBCT is presurgical implant planning. Linear measurement is used often in presurgical implant planning for determination of the exact amount of alveolar bone (height and width) and consequently size of the dental implants. Also linear measurements are used in orthodontic analysis and definition of jaw tumor size. Studies show that 94% of the CBCT measurements have been accurate, within 1 mm.[141516] The objective of this study is to evaluate the accuracy of the linear measurement of CBCT (Newtom VG) in axial and coronal planes with two voxel sizes: 0.15 mm and 0.3 mm.

MATERIALS AND METHODS

Four dry skulls, which were not identified by gender, age, or ethnicity, were used in this study. Fifteen linear distances were used for linear measurement accuracy between 22 anatomic landmarks [Table 1]. The landmarks were marked by 2 mm diameter round gutta-percha fixed to the skull, using heat. The linear distances were selected horizontally to account for linear measurements in the axial and coronal sections. The gold standard for each linear measurement was obtained by physical measurement, using a digital caliper (Mitutoyo Corp. Kawasaki, Japan) with the accuracy of 0.01 mm. The physical measurements were repeated twice by an observer. The mean was considered as the gold standard or real measurement.
Table 1

Anatomical landmarks used as references for measurements

Anatomical landmarks used as references for measurements The radiographic scans were obtained using CBCT Newtom VG (Quantitative Radiology, Verona, Italy). The skull was then centered and fixed in the CBCT system. The skulls were positioned according to the recommendations of the CBCT manufacturer. The midline laser beam of the CBCT system was adjusted to the midsagittal plane of the skull. The horizontal laser beam was parallel to the Frankfort line of the skull. Each skull was scanned twice: Full scan: Voxel size 0.3 mm (5.56 mAs, 110 kVp) and an 18-second scan time. High resolution scan: Voxel size 0.15 mm (8.3 mAs, 110 kVp) and a 36-second scan time. The raw data were reconstructed using the CBCT software (QRNNT V 2.21 Quantitative Radiology). Image reconstructions were made in the axial and coronal image planes. Radiographic measurements were made by three trained observers, with no knowledge of the real measurements, in the axial and coronal image planes using a measuring tool of the NNT software [Figure 1]. The images were viewed in a dimly lit room, on a 19-inch LG flatron monitor (LG, Seoul, Korea), with a screen resolution of 144 0× 900 pixels and a 32-bit depth color. Observers were free to choose the tools of the CBCT software, including brightness, contrast, and gamma changes, with no time limitation. In total, 120 radiographic measurements were performed by each observer. The radiographic measurements were repeated two weeks later for evaluation of intraobserver reliability. The radiographic measurements were recorded in comparison with the real measurements, with a negative number indicating underestimation and a positive number indicating overestimation.
Figure 1

Linear measurement between right and left mental foramen obtained by CBCT in axial plane

Linear measurement between right and left mental foramen obtained by CBCT in axial plane

Statistical analysis

The SPSS software version 17 was used for data analysis. The mean of all radiographic measurements for each image sequence was calculated. The intraclass correlation coefficient was used to determine the accuracy of the radiographic measurements and also interobserver and intraobserver reliability. The level of significance was considered to be 5% (P ≤ 0.05).

RESULTS

The mean difference and standard deviation of the radiographic measurements of each image plane from the gold standard are summarized in Table 2.
Table 2

Mean difference and standard deviation of radiographic and real measurements

Mean difference and standard deviation of radiographic and real measurements The mean of all the radiographic measurements was smaller than that of the real measurements, except for the axial high resolution. The mean difference of the radiographic measurements from that of the real measurements ranged from -0.2781 to 0.1418, and the standard deviation ranged from 0.991 to 1.441. No statistically significant difference was found between the radiographic measurements and real measurements (P value > 0.05). The ICC of each image plane comparing the radiographic measurements and real measurements in the first reading is given in Figure 2. The ICC ranged from 0.9935 to 0.9944, showing that the radiographic measurements were accurate (P value > 0.05). Regarding the statistical analysis, no statistically significant difference was seen for both interobserver and intraobserver reliability. The ICC for interobserver reliability varied from 0.9991 to 0.9996 and for intraobserver reliability it varied from 0.9849 to 0.9998 [Tables 3 and 4]. The ICC for each individual landmark is given in Table 5. The ICC for each individual landmark ranged from 0.7407 to 0.9996, showing no statistically significant difference between the radiographic measurements and real measurements in each landmark (P value > 0.05). The plotting of radiographic measurements against the gold standard is given in Figure 3.
Figure 2

Intraclass correlation coefficient for first reading

Table 3

Interobserver correlation for first reading

Table 4

Intraobserver reliability

Table 5

The intraclass correlation for each individual landmark

Figure 3

Scatter plot of radiographic measurements versus real measurements (gold standard): (a) Axial, (b) Coronal, (c) Axial H, (d) Coronal H

Intraclass correlation coefficient for first reading Interobserver correlation for first reading Intraobserver reliability The intraclass correlation for each individual landmark Scatter plot of radiographic measurements versus real measurements (gold standard): (a) Axial, (b) Coronal, (c) Axial H, (d) Coronal H

DISCUSSION

CBCT provides a valuable tool for evaluating craniofacial region. Effective radiation dose from a scan of maxillofacial volume is significantly lower than medical CT and is in the range of conventional dental radiographies.[14] The CBCT software provides some useful tools for clinical practice, such as, tools to measure distances and angles, to zoom, invert the gray scale, adjust the contrast, and gamma changes.[131415] One of the major uses of CBCT is presurgical implant planning. The linear measurement of distances is often used in presurgical implant planning for the determination of the exact amount of alveolar bone (height and width) and consequently the size of the dental implants. Also, linear measurements are used in orthodontic analysis and in the definition of jaw tumor size. The image data is acquired from a single 360 rotation scan around the patient.[17] Image reconstruction provides multiplanar images. The purpose of this study was to analyze the accuracy of linear measurements made in images obtained by CBCT in different image planes and different areas of the field of view (FOV), in different voxel sizes. In this study, anatomic landmarks were chosen in different areas of the skull, to check the accuracy of CBCT linear measurements in different areas of FOV. An analysis of the measurements showed that the CBCT measurements were highly accurate and reproducible. The mean differences between the radiographic measurements and real measurements varied from - 0.2781 mm for the coronal images (voxel size 0.3 mm) to 0.1418 mm for the axial high resolution images (voxel size 0.15 mm). The mean of differences was less than 0.5 mm in all measured image planes. This showed that CBCT provided a valuable measuring tool in different areas of FOV. The linear measurement accuracy of CBCT was tested in various units. Fatemitabar et al. evaluated the accuracy of CBCT (planmeca), and found the mean differences varying from 0.37 mm to 0.58 mm for CBCT, and from 0.37 mm to 0.72 mm for a 64-channel CT (Siemens).[12] Pinsky et al. found that CBCT (i-CAT) could be an accurate diagnostic tool for small osseous defects. They found mean differences varying from - 0.01 mm to 0.27 mm for width and height accuracy.[13] Stratemann et al. also found high accuracy in the CBCT images for linear distances compared to the real measurements. The error was small for two evaluated CBCT systems varying from to 0.07 mm ± 0.41 mm for NewTom 9000 and 0.00 ± 0.22 mm for CB Hitachi MercuRay.[14] Lascala et al., in a similarly designed study in which distances between 13 sites on human dry skull were measured using NewTom 9000 (Quantitative Radiology, Verona, Italy), found that the CBCT images underestimated the real distances between the skull sites, however, the differences between the CBCT and real measurements were only statistically significant for measurements of the skull base. In our study, the CBCT measurements were slightly underestimated (but not statistically significant) and the mean differences were negative, except for measurements in axial high resolution.[16] Interobserver reliability varied from 0.9991 to 0.9996. Intraobserver reliability varied from 0.9849 to 0.9998. This showed that CBCT measurements in the craniofacial area were highly reproducible. This was in agreement with the results of Kamburoglu et al., who found the interobserver reliability to be 0.995 to 1 and intraobserver reliability to be 0.992 to 1. Oz et al. also found high interobserver reliability of the CBCT measurements in the craniofacial area.[18] Measurements were also performed in two different voxel sizes: (a) Voxel size 0.3 mm. (b) voxel size 0.15 mm High Resolution (HR). The effect of the voxel size did not seem to be an important factor in linear measurement accuracy, as no significant difference was found between the measurements made in images with a voxel size of 0.3 mm and 0.15 mm (P value > 0.05). Stratemann et al. had the same results.[14] Brown et al. found that reducing the number of image projections did not result in a higher accuracy of CBCT measurements.[15] This was of clinical significance, as the patient radiation dose would be directly related to the voxel size and the number of projection images acquired. By reducing the number of image projections or increasing the voxel size, images with the same dimensional accuracy could be obtained. According to the findings of the present study, a CBCT scan with larger voxel size (0.3 mm in comparison to 0.15 mm) was recommended when the purpose of the CBCT scan was to measure linear distances, as in presurgical implant site evaluation and orthodontic analysis. This would result in a lower patient radiation dose and faster scan time. Faster scan time would decrease motion artifact due to patient movement. Patcas et al. evaluated the accuracy of linear intraoral measurements using CBCT. The radiological measurements were accurate, with a mean difference from the anatomical measurements of 0.14 mm.[19] Dalessandri et al., in an ex-vivo study, evaluated measurement reliability using two different CBCT scanners for orthodontic purposes. They found both scanners reliable for linear measurements.[20] In the present study, we found no significant difference between the accuracy of CBCT measurements in the axial and coronal sections. This was expected because the voxels in CBCT were isotropic (equal in all three dimensions) and this would result in the same image quality in the three orthogonal image planes. The accuracy of measurement distances on patients may be affected by a reduction in image quality due to soft tissue attenuation, restoration metallic artifacts, and patient movement. The skill of the operator in CBCT measurements and calibration of the hardware/software of the CBCT system are also effective in the accuracy of the CBCT measurements of the craniofacial area.

CONCLUSION

Cone beam computed tomography (Newtom VG) is highly accurate and reproducible in linear measurements in the axial and coronal image planes, and in different areas of the maxillofacial region. According to the findings of the present study, a CBCT scan with a larger voxel size (0.3 mm in comparison with 0.15 mm) is recommended when the purpose of the scan is to measure linear distances. This will result in lower patient radiation dose and faster scan time. We recommend methods with a lower patient radiation dose and faster scan time.
  20 in total

1.  Development of dento-maxillofacial cone beam X-ray computed tomography system.

Authors:  K Yamamoto; K Ueno; K Seo; D Shinohara
Journal:  Orthod Craniofac Res       Date:  2003       Impact factor: 1.826

2.  Cone beam computed tomography in craniofacial imaging.

Authors:  P Sukovic
Journal:  Orthod Craniofac Res       Date:  2003       Impact factor: 1.826

3.  Comparison of linear and angular measurements using two-dimensional conventional methods and three-dimensional cone beam CT images reconstructed from a volumetric rendering program in vivo.

Authors:  U Oz; K Orhan; N Abe
Journal:  Dentomaxillofac Radiol       Date:  2011-12       Impact factor: 2.419

4.  Dose reduction in maxillofacial imaging using low dose Cone Beam CT.

Authors:  Kostas Tsiklakis; Catherine Donta; Sophia Gavala; Kety Karayianni; Vasiliki Kamenopoulou; Costas J Hourdakis
Journal:  Eur J Radiol       Date:  2005-06-22       Impact factor: 3.528

5.  Accuracy of three-dimensional measurements using cone-beam CT.

Authors:  H M Pinsky; S Dyda; R W Pinsky; K A Misch; D P Sarment
Journal:  Dentomaxillofac Radiol       Date:  2006-11       Impact factor: 2.419

Review 6.  State-of-the-art on cone beam CT imaging for preoperative planning of implant placement.

Authors:  Maria Eugenia Guerrero; Reinhilde Jacobs; Miet Loubele; Filip Schutyser; Paul Suetens; Daniel van Steenberghe
Journal:  Clin Oral Investig       Date:  2006-02-16       Impact factor: 3.573

7.  Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography.

Authors:  A Suomalainen; T Vehmas; M Kortesniemi; S Robinson; J Peltola
Journal:  Dentomaxillofac Radiol       Date:  2008-01       Impact factor: 2.419

8.  Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: a tale of two CTs.

Authors:  R Patcas; G Markic; L Müller; O Ullrich; T Peltomäki; C J Kellenberger; C A Karlo
Journal:  Dentomaxillofac Radiol       Date:  2012-05-03       Impact factor: 2.419

9.  What is cone-beam CT and how does it work?

Authors:  William C Scarfe; Allan G Farman
Journal:  Dent Clin North Am       Date:  2008-10

10.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

View more
  13 in total

1.  Accuracy of linear measurement using cone-beam computed tomography at different reconstruction angles.

Authors:  Sima Nikneshan; Shadi Hamidi Aval; Neema Bakhshalian; Shahriyar Shahab; Mahdis Mohammadpour; Soodeh Sarikhani
Journal:  Imaging Sci Dent       Date:  2014-11-25

2.  The Effect of Mandibular Angulation on Preoperative Assessment of Dental Implant Insertion at Premolar Region: CBCT Study.

Authors:  Sadaf Sadat Mahmoudinezhad; Azarnoosh AryanKia; Sanaz Sharifi Shooshtari; Kooshan Moradi
Journal:  Biomed Res Int       Date:  2022-05-28       Impact factor: 3.246

3.  Comparison of cone-beam computed tomography with bitewing radiography for detection of periodontal bone loss and assessment of effects of different voxel resolutions: an in vitro study.

Authors:  Hayriye Cetmili; Melek Tassoker; Sevgi Sener
Journal:  Oral Radiol       Date:  2018-06-08       Impact factor: 1.852

4.  The effect of voxel size on the measurement of mandibular thickness in cone-beam computed tomography.

Authors:  Ehsan Hekmatian; Nasim Jafari-Pozve; Ladan Khorrami
Journal:  Dent Res J (Isfahan)       Date:  2014-09

5.  Comparison of Results of Measurement of Dimensions of the Placed Dental Implants on Cone Beam Computed Tomography with Dimensions of the Producers of the Implants.

Authors:  Merisa Repesa; Amela Sofic; Selma Jakupovic; Selma Tosum; Lejla Kazazic; Almir Dervisevic
Journal:  Acta Inform Med       Date:  2017-06

6.  Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography.

Authors:  Nathália Cristine da Silva; Maurício Barriviera; José Luiz Cintra Junqueira; Francine Kühl Panzarella; Ricardo Raitz
Journal:  Imaging Sci Dent       Date:  2017-03-21

7.  Effect of Dental Implant Metal Artifacts on Accuracy of Linear Measurements by Two Cone-Beam Computed Tomography Systems Before and After Crown Restoration.

Authors:  Hoorieh Bashizadeh Fakhar; Roxana Rashtchian; Milad Parvin
Journal:  J Dent (Tehran)       Date:  2017-11

8.  A new method to standardize CBCT for quantitative evaluation of alveolar ridge preservation in the mandible: a case report and review of the literature.

Authors:  Yang Xia; Lizhe Xie; Yi Zhou; Tianxi Song; Feimin Zhang; Ning Gu
Journal:  Regen Biomater       Date:  2015-10-14

9.  Automatic analysis algorithm for acquiring standard dental and mandibular shape data using cone-beam computed tomography.

Authors:  Jae Joon Hwang; Sang-Sun Han; Chena Lee; Yun-Hoa Jung
Journal:  Sci Rep       Date:  2018-09-10       Impact factor: 4.379

10.  Bone changes in the mandibular incisors after orthodontic correction of dental crowding without extraction: A cone-beam computed tomographic evaluation.

Authors:  Claudia Scigliano Valerio; Cláudia Assunção E Alves Cardoso; Eustáquio Afonso Araújo; Elton Gonçalves Zenóbio; Flávio Ricardo Manzi
Journal:  Imaging Sci Dent       Date:  2021-02-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.