Comparison of cone beam computed tomography imaging with physical measures.

S A Stratemann1, J C Huang, K Maki, A J Miller, D C Hatcher.   

Abstract

OBJECTIVES: The goal of this study was to determine the accuracy of measuring linear distances between landmarks commonly used in orthodontic analysis on a human skull using two cone beam CT (CBCT) systems.
METHODS: Measurements of length were taken using volumetric data from two CBCT systems and were compared with physical measures using a calliper applied to one human adult skull. Landmarks were identified with chromium steel balls embedded at 32 cranial and 33 mandibular landmarks and the linear measures were taken with a digital calliper. The skull was then scanned with two different CBCT systems: the NewTom QR DVT 9000 (Aperio Inc, Sarasota, FL) and the Hitachi MercuRay (Hitachi Medico Technology, Tokyo, Japan). CT data including the landmark point data were threshold segmented using CyberMed's CB Works software (CB Works 1.0, CyberMed Inc., Seoul, Korea). The resulting segmentations were exported from CB Works as VRML (WRL) files to Amira software (Amira 3.1, Mercury Computer Systems GmbH, Berlin, Germany).
RESULTS: The error was small compared with the gold standard of the physical calliper measures for both the NewTom (0.07+/-0.41 mm) and CB MercuRay (0.00+/-0.22 mm) generated data. Absolute error to the gold standard was slightly positive, indicating minor compression relative to the calliper measurement. The error was slightly smaller in the CB MercuRay than in the NewTom, probably related to a broader greyscale range for describing beam attenuation in 12-bit vs 8-bit data.
CONCLUSIONS: The volumetric data rendered with both CBCT systems provided highly accurate data compared with the gold standard of physical measures directly from the skulls, with less than 1% relative error.

Entities:  

Mesh:

Year:  2008        PMID: 18239035     DOI: 10.1259/dmfr/31349994

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  65 in total

1.  Digital method for quantification of circumferential periodontal bone level using cone beam CT.

Authors:  Jonathan Fleiner; Christian Hannig; Dirk Schulze; Andres Stricker; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2012-03-21       Impact factor: 3.573

2.  Accuracy of CBCT measurements of a human skull.

Authors:  Kivanç Kamburoğlu; Eray Kolsuz; Hakan Kurt; Cenk Kiliç; Tuncer Özen; Candan Semra Paksoy
Journal:  J Digit Imaging       Date:  2011-10       Impact factor: 4.056

Review 3.  The current status of cone beam computed tomography imaging in orthodontics.

Authors:  S Kapila; R S Conley; W E Harrell
Journal:  Dentomaxillofac Radiol       Date:  2011-01       Impact factor: 2.419

4.  Geometric analysis of alveolar bone around the incisors after anterior retraction following premolar extraction.

Authors:  Fan Zhang; Suk-Cheol Lee; Jun-Beom Lee; Kyung-Min Lee
Journal:  Angle Orthod       Date:  2019-11-26       Impact factor: 2.079

5.  High-resolution CT of transplanted teeth: imaging technique and measurement accuracy.

Authors:  André Gahleitner; Ulrike Kuchler; Peter Homolka; Janina Heschl; Georg Watzek; Herwig Imhof
Journal:  Eur Radiol       Date:  2008-08-09       Impact factor: 5.315

6.  Accuracy of linear measurement in the Galileos cone beam computed tomography under simulated clinical conditions.

Authors:  R Ganguly; A Ruprecht; S Vincent; J Hellstein; S Timmons; F Qian
Journal:  Dentomaxillofac Radiol       Date:  2011-07       Impact factor: 2.419

7.  Assessing the need for computed tomography for lower-third-molar extraction: a survey among 322 dentists.

Authors:  S Sivolella; G Boccuzzo; E Gasparini; G De Conti; M Berengo
Journal:  Radiol Med       Date:  2011-04-19       Impact factor: 3.469

8.  Accuracy assessment of the axial images obtained from cone beam computed tomography.

Authors:  Francine Kühl Panzarella; J L C Junqueira; L B Oliveira; N S de Araújo; C Costa
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

9.  Accuracy of anatomical landmark identification using different CBCT- and MSCT-based 3D images: an in vitro study.

Authors:  Jürgen Medelnik; Klaus Hertrich; Stefanie Steinhäuser-Andresen; Ursula Hirschfelder; Elisabeth Hofmann
Journal:  J Orofac Orthop       Date:  2011-09-08       Impact factor: 1.938

10.  Calculating nasoseptal flap dimensions: a cadaveric study using cone beam computed tomography.

Authors:  Ellen Ten Dam; Astrid G W Korsten-Meijer; Rutger H Schepers; Wicher J van der Meer; Peter O Gerrits; Bernard F A M van der Laan; Robert A Feijen
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-10-31       Impact factor: 2.503

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.