Evaluation of the BioPlex 2200 syphilis system as a first-line method of reverse-sequence screening for syphilis diagnosis.

Antonella Marangoni1, Paola Nardini, Claudio Foschi, Alessandra Moroni, Antonietta D'Antuono, Letizia Bacchi Reggiani, Roberto Cevenini.   

Abstract

Despite recent technological advances, the diagnosis of syphilis remains a challenging enterprise. Actually, most high-volume laboratories have adopted the "reverse algorithm" due several factors, including the potential to automate testing. Recently, immunoassays processed on random-access systems have been proposed as screening tests. The purpose of this study was to evaluate diagnostic performances of BioPlex 2200 Syphilis IgG and BioPlex 2200 Syphilis IgM, tests based on Multiplex Flow technology, in comparison with the performance of Architect Syphilis TP, a chemiluminescent immunoassay for the detection of IgG and/or IgM anti-Treponema pallidum antibodies. A retrospective study was performed with a panel of 100 blood donor sera, a panel of 350 clinical and laboratory-characterized syphilitic sera, and 170 samples obtained from subjects with potentially interfering conditions. Moreover, 200 unselected samples submitted to the Microbiology Laboratory of St. Orsola Hospital in Bologna for routine screening for syphilis were evaluated. As confirmatory tests, T. pallidum hemagglutination and Western blot assays were used. Considering the IgG Western blot (WB) assay to be the gold standard method, BioPlex 2200 Syphilis IgG specificity was far higher than Architect Syphilis TP specificity (89.7% versus 78.4%, respectively), whereas the sensitivity was 100% for both automated methods. Compared to the IgM WB assay, BioPlex 2200 Syphilis IgM performed with a specificity of 94.9%, whereas the sensitivity was 84.8%. Considering the excellent ease of use and automation, the high sample throughput and its valuable analytical performances, BioPlex Syphilis 2200 IgG could represent a suitable choice for high-volume laboratories. BioPlex Syphilis 2200 IgM could be considered a good addition to IgG testing for uncovering active infections.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23697575      PMCID: PMC3697445          DOI: 10.1128/CVI.00316-13

Source DB:  PubMed          Journal:  Clin Vaccine Immunol        ISSN: 1556-679X


  36 in total

1.  Evaluation of an enzyme immunoassay technique for detection of antibodies against Treponema pallidum.

Authors:  Rita Castro; Emília S Prieto; Irene Santo; Jacinta Azevedo; Filomena da L Exposto
Journal:  J Clin Microbiol       Date:  2003-01       Impact factor: 5.948

2.  Western immunoblotting with five Treponema pallidum recombinant antigens for serologic diagnosis of syphilis.

Authors:  V Sambri; A Marangoni; C Eyer; C Reichhuber; E Soutschek; M Negosanti; A D'Antuono; R Cevenini
Journal:  Clin Diagn Lab Immunol       Date:  2001-05

3.  Electrophoretic transfer of proteins from polyacrylamide gels to nitrocellulose sheets: procedure and some applications.

Authors:  H Towbin; T Staehelin; J Gordon
Journal:  Proc Natl Acad Sci U S A       Date:  1979-09       Impact factor: 11.205

4.  Evaluation of recomWell Treponema, a novel recombinant antigen-based enzyme-linked immunosorbent assay for the diagnosis of syphilis.

Authors:  V Sambri; A Marangoni; M A Simone; A D'Antuono; M Negosanti; R Cevenini
Journal:  Clin Microbiol Infect       Date:  2001-04       Impact factor: 8.067

5.  Evaluation of an enzyme immunoassay for IgM antibodies to Treponema pallidum in syphilis in man.

Authors:  F Müller; M Moskophidis
Journal:  Br J Vener Dis       Date:  1984-10

6.  Screening for treponemal infection by a new enzyme immunoassay.

Authors:  H Young; A Moyes; A McMillan; D H Robertson
Journal:  Genitourin Med       Date:  1989-04

7.  Four-step enzyme-linked immunosorbent assay for detection of Treponema pallidum antibody.

Authors:  C E Farshy; E F Hunter; L O Helsel; S A Larsen
Journal:  J Clin Microbiol       Date:  1985-03       Impact factor: 5.948

8.  Enzyme-linked immunosorbent assay for detection of immunoglobulin M antibody against the Reiter treponeme flagellum in syphilis.

Authors:  N S Pedersen; C S Petersen; N H Axelsen
Journal:  J Clin Microbiol       Date:  1982-10       Impact factor: 5.948

9.  An outbreak no longer: factors contributing to the return of syphilis in Greater Manchester.

Authors:  M Ashton; W Sopwith; P Clark; D McKelvey; L Lighton; D Mandal
Journal:  Sex Transm Infect       Date:  2003-08       Impact factor: 3.519

10.  Syphilis.

Authors:  R W Peeling; D C W Mabey
Journal:  Nat Rev Microbiol       Date:  2004-06       Impact factor: 60.633

View more
  13 in total

1.  Evaluation of the HISCL Anti-Treponema pallidum Assay as a Screening Test for Syphilis.

Authors:  Jingna An; Qixia Chen; Qianqian Liu; Chenli Rao; Dongdong Li; Tingting Wang; Chuanmin Tao; Lanlan Wang
Journal:  Clin Vaccine Immunol       Date:  2015-05-13

2.  Performance evaluation of the Elecsys syphilis assay for the detection of total antibodies to Treponema pallidum.

Authors:  Martin Enders; Andrea Hunjet; Michael Gleich; Roland Imdahl; Annelies Mühlbacher; Harald Schennach; Kriangsak Chaiwong; Tasanee Sakuldamrongpanich; Ajda Turhan; Rüchan Sertöz; Eva Wolf; Wolfgang Mayer; Chuanmin Tao; Lan Lan Wang; Simona Semprini; Vittorio Sambri
Journal:  Clin Vaccine Immunol       Date:  2014-10-29

3.  Evaluation of the Vitros Syphilis TPA chemiluminescence immunoassay as a first-line method for reverse syphilis screening.

Authors:  Victoria González; Gema Fernández; Eva Dopico; Nuria Margall; Juliana Esperalba; Carme Muñoz; Elisabeth Castro; Elena Sulleiro; Lurdes Matas
Journal:  J Clin Microbiol       Date:  2015-01-21       Impact factor: 5.948

4.  Current trends in donor testing to detect syphilis infection.

Authors:  Karoll J Cortez; Melissa A Greenwald
Journal:  Curr Infect Dis Rep       Date:  2014-09       Impact factor: 3.725

5.  Analytical Comparison of the Architect Syphilis TP and Liaison Treponema Automated Chemiluminescent Immunoassays and Their Performance in a Reverse Syphilis Screening Algorithm.

Authors:  Alan M Sanfilippo; Kristie Freeman; John L Schmitz
Journal:  J Clin Microbiol       Date:  2018-07-26       Impact factor: 5.948

6.  Evaluation of the Lumipulse G TP-N Chemiluminescent Immunoassay as a Syphilis Screening Test.

Authors:  Daniel A Ortiz; Michael J Loeffelholz
Journal:  J Clin Microbiol       Date:  2017-09-06       Impact factor: 5.948

7.  Clinical Evaluation of Fully Automated Elecsys® Syphilis Assay for the Detection of Antibodies of Treponema pallidum.

Authors:  Dongdong Li; Jingna An; Tingting Wang; Chuanmin Tao; Lanlan Wang
Journal:  J Clin Lab Anal       Date:  2016-05-26       Impact factor: 2.352

8.  Contribution of a Comparative Western Blot Method to Early Postnatal Diagnosis of Congenital Syphilis.

Authors:  Antonella Marangoni; Claudio Foschi; Maria Grazia Capretti; Paola Nardini; Monica Compri; Luigi Tommaso Corvaglia; Giacomo Faldella; Roberto Cevenini
Journal:  Clin Vaccine Immunol       Date:  2016-05-06

9.  Automation in Blood Centre: Its impact on Blood Safety.

Authors:  Snehalata C Gupte
Journal:  Asian J Transfus Sci       Date:  2015-04

10.  Multiplex Antibody Detection for Noninvasive Genus-Level Diagnosis of Prosthetic Joint Infection.

Authors:  Simon Marmor; Thomas Bauer; Nicole Desplaces; Beate Heym; Anne-Laure Roux; Olivier Sol; Julie Rogé; Florence Mahé; Laurent Désiré; Philippe Aegerter; Idir Ghout; Jacques Ropers; Jean-Louis Gaillard; Martin Rottman
Journal:  J Clin Microbiol       Date:  2016-02-10       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.