Effects of wobbling angle on the stability measures of orthodontic mini-implants during insertion and removal procedures.

Il-Sik Cho1, Sang-Ho Baek, Young Ho Kim.   

Abstract

OBJECTIVE: To investigate the effects of wobbling angle on the stability measures of orthodontic mini-implants (OMIs) during insertion and removal procedures in artificial bone blocks.
MATERIALS AND METHODS: A total of 36 OMIs (self-drilling type, cylindrical shape, 7 mm in length, 1.45 mm in diameter) were allocated into three groups according to the amount of wobbling angle (W-0°, W-2°, and W-4° groups; N = 12 per group). The OMIs were installed and subsequently removed from artificial bone blocks (Sawbone) using a driving torque tester with a uniform speed of 28 rpm. Insertion peak time (IPT), maximum insertion torque (MIT), total insertion energy (TIE), near-peak insertion energy (NPIE), maximum removal torque (MRT), and near-peak removal energy (NPRE) were measured.
RESULTS: The W-4° group showed the longest IPT and highest TIE and NPIE, followed by the W-2° and W-0° groups (W-0° < W-2° < W-4°, all P < .001). The W-2° and W-4° groups showed significant increase in MIT compared with the W-0° group (W-0° < [W-4°,W-2°]; P < .001). Although there was no significant difference in NPRE among the three groups, the W-4° group showed a decrease in MRT compared with the W-0° and W-2° groups (W-4° < {W-2°,W-0°]; P < .05). Although the W-4° group showed a 14.5% (2.9 Ncm) increase in MIT compared with the W-0° group, there was only a 6% (1.3 Ncm) decrease in MRT from the W-0° group to the W-4° group.
CONCLUSION: Slight wobbling during the OMI insertion procedure may be acceptable in terms of the stability measures of OMIs during insertion and removal procedures.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23663168      PMCID: PMC8722824          DOI: 10.2319/021513-134.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  30 in total

1.  Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region.

Authors:  Cheol-Hyun Moon; Dong-Gun Lee; Hyun-Sun Lee; Jeong-Soo Im; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2008-01       Impact factor: 2.079

2.  Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length.

Authors:  Seon-A Lim; Jung-Yul Cha; Chung-Ju Hwang
Journal:  Angle Orthod       Date:  2008-03       Impact factor: 2.079

3.  Guidelines for success in placement of orthodontic mini-implants.

Authors:  Cesare Luzi; Carlalberta Verna; Birte Melsen
Journal:  J Clin Orthod       Date:  2009-01

4.  Mini-implant anchorage for the orthodontic practitioner.

Authors:  Sebastian Baumgaertel; Mohammad R Razavi; Mark G Hans
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-04       Impact factor: 2.650

5.  Rotational resistance of surface-treated mini-implants.

Authors:  Seong-Hun Kim; Shin-Jae Lee; Il-Sik Cho; Seong-Kyun Kim; Tae-Woo Kim
Journal:  Angle Orthod       Date:  2009-09       Impact factor: 2.079

6.  Microdamage in cortical bone due to the overtightening of orthodontic microscrews.

Authors:  Christian Wawrzinek; Thorsten Sommer; Helge Fischer-Brandies
Journal:  J Orofac Orthop       Date:  2008-03       Impact factor: 1.938

7.  Rationale for the use of low-torque endodontic motors in root canal instrumentation.

Authors:  G Gambarini
Journal:  Endod Dent Traumatol       Date:  2000-06

8.  Comparison of stability between cylindrical and conical type mini-implants. Mechanical and histological properties.

Authors:  Jong-Wan Kim; Seung-Hak Baek; Tae-Woo Kim; Young-Il Chang
Journal:  Angle Orthod       Date:  2008-07       Impact factor: 2.079

9.  In vitro and in vivo mechanical stability of orthodontic mini-implants.

Authors:  Il-Sik Cho; Sung-Kyun Kim; Young-Il Chang; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2011-10-19       Impact factor: 2.079

10.  Comparison of success rates of orthodontic mini-screws by the insertion method.

Authors:  Jung Suk Kim; Seong Hwan Choi; Sang Kwon Cha; Jang Han Kim; Hwa Jin Lee; Sang Seon Yeom; Chung Ju Hwang
Journal:  Korean J Orthod       Date:  2012-10-29       Impact factor: 1.372

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.