Abstract

BACKGROUND: Recently, patient-specific guides (PSGs) have been introduced, claiming a significant improvement in accuracy and reproducibility of component positioning in TKA. Despite intensive marketing by the manufacturers, this claim has not yet been confirmed in a controlled prospective trial. QUESTIONS/PURPOSES: We (1) compared three-planar component alignment and overall coronal mechanical alignment between PSG and conventional instrumentation and (2) logged the need for applying changes in the suggested position of the PSG.
METHODS: In this randomized controlled trial, we enrolled 128 patients. In the PSG cohort, surgical navigation was used as an intraoperative control. When the suggested cut deviated more than 3° from target, the use of PSG was abandoned and marked as an outlier. When cranial-caudal position or size was adapted, the PSG was marked as modified. All patients underwent long-leg standing radiography and CT scan. Deviation of more than 3° from the target in any plane was defined as an outlier.
RESULTS: The PSG and conventional cohorts showed similar numbers of outliers in overall coronal alignment (25% versus 28%; p = 0.69), femoral coronal alignment (7% versus 14%) (p = 0.24), and femoral axial alignment (23% versus 17%; p = 0.50). There were more outliers in tibial coronal (15% versus 3%; p = 0.03) and sagittal 21% versus 3%; p = 0.002) alignment in the PSG group than in the conventional group. PSGs were abandoned in 14 patients (22%) and modified in 18 (28%).
CONCLUSIONS: PSGs do not improve accuracy in TKA and, in our experience, were somewhat impractical in that the procedure needed to be either modified or abandoned with some frequency.

Entities:  

Mesh:

Year:  2014        PMID: 23616267      PMCID: PMC3889461          DOI: 10.1007/s11999-013-2997-4

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  44 in total

1.  The Ranawat Award. Femoral component rotation during total knee arthroplasty.

Authors:  C W Olcott; R D Scott
Journal:  Clin Orthop Relat Res       Date:  1999-10       Impact factor: 4.176

2.  The effect of femoral component position on patellar tracking after total knee arthroplasty.

Authors:  D D Rhoads; P C Noble; J D Reuben; O M Mahoney; H S Tullos
Journal:  Clin Orthop Relat Res       Date:  1990-11       Impact factor: 4.176

3.  Custom-fit total knee arthroplasty (OtisKnee) results in malalignment.

Authors:  Brian A Klatt; Nitin Goyal; Matthew S Austin; William J Hozack
Journal:  J Arthroplasty       Date:  2008-01       Impact factor: 4.757

4.  How precise can bony landmarks be determined on a CT scan of the knee?

Authors:  J Victor; D Van Doninck; L Labey; B Innocenti; P M Parizel; J Bellemans
Journal:  Knee       Date:  2009-02-05       Impact factor: 2.199

Review 5.  Rotational alignment of the distal femur: a literature review.

Authors:  J Victor
Journal:  Orthop Traumatol Surg Res       Date:  2009-07-09       Impact factor: 2.256

6.  Evaluation of the accuracy of a patient-specific instrumentation by navigation.

Authors:  Fabio Conteduca; Raffaele Iorio; Daniele Mazza; Ludovico Caperna; Gabriele Bolle; Giuseppe Argento; Andrea Ferretti
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-06-27       Impact factor: 4.342

7.  Do patient-specific guides improve coronal alignment in total knee arthroplasty?

Authors:  Ryan M Nunley; Bradley S Ellison; Jinjun Zhu; Erin L Ruh; Stephen M Howell; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2011-12-20       Impact factor: 4.176

8.  The effects of axial rotational alignment of the femoral component on knee stability and patellar tracking in total knee arthroplasty demonstrated on autopsy specimens.

Authors:  Y S Anouchi; L A Whiteside; A D Kaiser; M T Milliano
Journal:  Clin Orthop Relat Res       Date:  1993-02       Impact factor: 4.176

9.  The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus.

Authors:  Johan Bellemans; William Colyn; Hilde Vandenneucker; Jan Victor
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

10.  Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA.

Authors:  Vincent Y Ng; Jeffrey H DeClaire; Keith R Berend; Bethany C Gulick; Adolph V Lombardi
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

View more
  60 in total

1.  Assessing the accuracy of patient-specific guides for total knee arthroplasty.

Authors:  Jong-Keun Seon; Hyeong-Won Park; Seung-Hyun Yoo; Eun-Kyoo Song
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-11-16       Impact factor: 4.342

2.  Assessment of patient-specific instrumentation precision through bone resection measurements.

Authors:  F Zambianchi; A Colombelli; V Digennaro; A Marcovigi; R Mugnai; F Fiacchi; D Sandoni; A Belluati; F Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-12-24       Impact factor: 4.342

3.  Comparison of custom cutting guides based on three-dimensional computerized CT-scan planning and a conventional ancillary system based on two-dimensional planning in total knee arthroplasty: a randomized controlled trial.

Authors:  Elhadi Sariali; Charles Kajetanek; Yves Catonné
Journal:  Int Orthop       Date:  2019-06-21       Impact factor: 3.075

4.  Alignment in total knee arthroplasty, still more questions than answers….

Authors:  Emmanuel Thienpont; Johan Bellemans; Jan Victor; Roland Becker
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-10       Impact factor: 4.342

5.  Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial.

Authors:  Nienke M Kosse; Petra J C Heesterbeek; Janneke J P Schimmel; Gijs G van Hellemondt; Ate B Wymenga; Koen C Defoort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-11-28       Impact factor: 4.342

Review 6.  Patient specific instrumentation in total knee arthroplasty: a state of the art.

Authors:  Lorenzo Mattei; Pietro Pellegrino; Michel Calò; Alessandro Bistolfi; Filippo Castoldi
Journal:  Ann Transl Med       Date:  2016-04

Review 7.  No difference in mechanical alignment and femoral component placement between patient-specific instrumentation and conventional instrumentation in TKA.

Authors:  Huichao Fu; Jiaxing Wang; Shenyuan Zhou; Tao Cheng; Wen Zhang; Qi Wang; Xianlong Zhang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-06-11       Impact factor: 4.342

8.  Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial.

Authors:  Tilman Pfitzner; Matthew P Abdel; Philipp von Roth; Carsten Perka; Hagen Hommel
Journal:  Clin Orthop Relat Res       Date:  2014-07-15       Impact factor: 4.176

9.  Inter-observer precision and physiologic variability of mri landmarks used to determine rotational alignment in conventional and patient-specific TKA.

Authors:  Andrew Park; Denis Nam; Michael V Friedman; Stephen T Duncan; Travis J Hillen; Robert L Barrack
Journal:  J Arthroplasty       Date:  2014-09-06       Impact factor: 4.757

10.  Patient-specific positioning guides do not consistently achieve the planned implant position in UKA.

Authors:  Justin A M J van Leeuwen; Stephan M Röhrl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-08-12       Impact factor: 4.342

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.