Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial.

Tilman Pfitzner1, Matthew P Abdel, Philipp von Roth, Carsten Perka, Hagen Hommel.   

Abstract

BACKGROUND: Patient-specific instrumentation in TKA has the proposed benefits of improving coronal and sagittal alignment and rotation of the components. In contrast, the literature is inconsistent if the use of patient-specific instrumentation improves alignment in comparison to conventional instrumentation. Depending on the manufacturer, patient-specific instrumentation is based on either MRI or CT scans. However, it is unknown whether one patient-specific instrumentation approach is more accurate than the other and if there is a potential benefit in terms of reduction of duration of surgery. QUESTIONS/PURPOSES: We compared the accuracy of MRI- and CT-based patient-specific instrumentation with conventional instrumentation and with each other in TKAs. The three approaches also were compared with respect to validated outcomes scores and duration of surgery.
METHODS: A randomized clinical trial was conducted in which 90 patients were enrolled and divided into three groups: CT-based, MRI-based patient-specific instrumentation, and conventional instrumentation. The groups were not different regarding age, male/female sex distribution, and BMI. In all groups, coronal and sagittal alignments were measured on postoperative standing long-leg and lateral radiographs. Component rotation was measured on CT scans. Clinical outcomes (Knee Society and WOMAC scores) were evaluated preoperatively and at a mean of 3 months postoperatively and the duration of surgery was analyzed for each patient. MRI- and CT-based patient-specific instrumentation groups were first compared with conventional instrumentation, the patient-specific instrumentation groups were compared with each other, and all three approaches were compared for clinical outcome measures and duration of surgery.
RESULTS: Compared with conventional instrumentation MRI- and CT-based patient-specific instrumentation showed higher accuracy regarding the coronal limb axis (MRI versus conventional, 1.0° [range, 0°-4°] versus 4.5° [range, 0°-8°], p < 0.001; CT versus conventional, 3.0° [range, 0°-5°] versus 4.5° [range, 0°-8°], p = 0.02), femoral rotation (MRI versus conventional, 1.0° [range, 0°-2°] versus 4.0° [range, 1°-7°], p < 0.001; CT versus conventional, 1.0° [range, 0°-2°] versus 4.0° [range, 1°-7°], p < 0.001), and tibial slope (MRI versus conventional, 1.0° [range, 0°-2°] versus 3.5° [range, 1°-7°], p < 0.001; CT versus conventional, 1.0° [range, 0°-2°] versus 3.5° [range, 1°-7°], p < 0.001), but the differences were small. Furthermore, MRI-based patient-specific instrumentation showed a smaller deviation in the postoperative coronal mechanical limb axis compared with CT-based patient-specific instrumentation (MRI versus CT, 1.0° [range, 0°-4°] versus 3.0° [range, 0°-5°], p = 0.03), while there was no difference in femoral rotation or tibial slope. Although there was a significant reduction of the duration of surgery in both patient-specific instrumentation groups in comparison to conventional instrumentation (MRI versus conventional, 58 minutes [range, 53-67 minutes] versus 76 minutes [range, 57-83 minutes], p < 0.001; CT versus conventional, 63 minutes [range, 59-69 minutes] versus 76 minutes [range, 57-83 minutes], p < .001), there were no differences in the postoperative Knee Society pain and function and WOMAC scores among the groups.
CONCLUSIONS: Although this study supports that patient-specific instrumentation increased accuracy compared with conventional instrumentation and that MRI-based patient-specific instrumentation is more accurate compared with CT-based patient-specific instrumentation regarding coronal mechanical limb axis, differences are only subtle and of questionable clinical relevance. Because there are no differences in the long-term clinical outcome or survivorship yet available, the widespread use of this technique cannot be recommended.

Entities:  

Mesh:

Year:  2014        PMID: 25024031      PMCID: PMC4160476          DOI: 10.1007/s11999-014-3784-6

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  52 in total

1.  Analysis of procedure-related costs and proposed benefits of using patient-specific approach in total knee arthroplasty.

Authors:  Tyler Steven Watters; Richard C Mather; James A Browne; Keith R Berend; Adolph V Lombardi; Michael P Bolognesi
Journal:  J Surg Orthop Adv       Date:  2011

2.  Rotational alignment of femoral components in total knee arthroplasty: nonimage-based navigation system versus conventional technique.

Authors:  Hyuk Soo Han; Sang Cheol Seong; Sahnghoon Lee; Myung Chul Lee
Journal:  Orthopedics       Date:  2006-10       Impact factor: 1.390

3.  A quantitative approach to radiography of the lower limb. Principles and applications.

Authors:  T D Cooke; R A Scudamore; J T Bryant; C Sorbie; D Siu; B Fisher
Journal:  J Bone Joint Surg Br       Date:  1991-09

4.  Comparison of two minimally invasive implantation instrument-sets for total knee arthroplasty.

Authors:  Arno Martin; Mitchell B Sheinkop; Mary M Langhenry; Mark Widemschek; Thomas Benesch; Archibald von Strempel
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-10-22       Impact factor: 4.342

5.  The gap technique does not rotate the femur parallel to the epicondylar axis.

Authors:  Georg Matziolis; Hinrich Boenicke; Sascha Pfiel; Georgi Wassilew; Carsten Perka
Journal:  Arch Orthop Trauma Surg       Date:  2010-05-18       Impact factor: 3.067

6.  Do patient-specific guides improve coronal alignment in total knee arthroplasty?

Authors:  Ryan M Nunley; Bradley S Ellison; Jinjun Zhu; Erin L Ruh; Stephen M Howell; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2011-12-20       Impact factor: 4.176

7.  Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA.

Authors:  Vincent Y Ng; Jeffrey H DeClaire; Keith R Berend; Bethany C Gulick; Adolph V Lombardi
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

8.  Neutral mechanical alignment: a requirement for successful TKA: affirms.

Authors:  Adolph V Lombardi; Keith R Berend; Vincent Y Ng
Journal:  Orthopedics       Date:  2011-09-09       Impact factor: 1.390

Review 9.  Advances in imaging of osteoarthritis and cartilage.

Authors:  Frank W Roemer; Michel D Crema; Siegfried Trattnig; Ali Guermazi
Journal:  Radiology       Date:  2011-08       Impact factor: 11.105

10.  Postoperative alignment and ROM affect patient satisfaction after TKA.

Authors:  Shuichi Matsuda; Shinya Kawahara; Ken Okazaki; Yasutaka Tashiro; Yukihide Iwamoto
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

View more
  26 in total

1.  Clinical faceoff: Neutrally versus kinematically aligned TKA.

Authors:  Matthew P Abdel; Sam Oussedik; Michael B Cross
Journal:  Clin Orthop Relat Res       Date:  2014-10-24       Impact factor: 4.176

2.  Structure-enhanced local phase filtering using L0 gradient minimization for efficient semiautomated knee magnetic resonance imaging segmentation.

Authors:  Mikhiel Lim; Ilker Hacihaliloglu
Journal:  J Med Imaging (Bellingham)       Date:  2016-12-02

3.  Editorial: Our love affair with technology and the choices we make.

Authors:  Seth S Leopold
Journal:  Clin Orthop Relat Res       Date:  2014-08-07       Impact factor: 4.176

Review 4.  Favourable alignment outcomes with MRI-based patient-specific instruments in total knee arthroplasty.

Authors:  Martijn G M Schotanus; Elke Thijs; Marion Heijmans; Rein Vos; Nanne P Kort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-07-11       Impact factor: 4.342

5.  Handheld Navigation Device and Patient-Specific Cutting Guides Result in Similar Coronal Alignment for Primary Total Knee Arthroplasty: a Retrospective Matched Cohort Study.

Authors:  Michael E Steinhaus; Alexander S McLawhorn; Shawn S Richardson; Patrick Maher; David J Mayman
Journal:  HSS J       Date:  2016-02-29

6.  Medical 3D Printing Cost-Savings in Orthopedic and Maxillofacial Surgery: Cost Analysis of Operating Room Time Saved with 3D Printed Anatomic Models and Surgical Guides.

Authors:  David H Ballard; Patrick Mills; Richard Duszak; Jeffery A Weisman; Frank J Rybicki; Pamela K Woodard
Journal:  Acad Radiol       Date:  2019-09-18       Impact factor: 3.173

7.  Verification of in vivo accuracy of Trumatch™ patient-specific instrumentation in total knee replacement using pin-less computer navigation.

Authors:  Nikolai Briffa; Mohamed A Imam; Ravi Mallina; Ashraf Abdelkafy; Ajeya Adhikari
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-09-07

8.  No correlation between rotation of femoral components in the transverse plane and clinical outcome after total knee arthroplasty.

Authors:  Roland Becker; Katharina Bäker; Hagen Hommel; Manfred Bernard; Sebastian Kopf
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-05-16       Impact factor: 4.342

9.  Outcomes following total knee arthroplasty with CT-based patient-specific instrumentation.

Authors:  Meng Zhu; Jerry Yongqiang Chen; Hwei Chi Chong; Andy Khye Soon Yew; Leon Siang Shen Foo; Shi-Lu Chia; Ngai Nung Lo; Seng Jin Yeo
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-09-26       Impact factor: 4.342

10.  [Total and unicompartmental knee replacement. Patient-specific Instrumentation].

Authors:  G Köster; C Biró
Journal:  Orthopade       Date:  2016-04       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.