INTRODUCTION: The objective of this study was to determine if a synthetic bone substitute would provide results similar to bone from osteoporotic femoral heads during in vitro testing with orthopaedic implants. If the synthetic material could produce results similar to those of the osteoporotic bone, it could reduce or eliminate the need for testing of implants on bone. METHODS: Pushout studies were performed with the dynamic hip screw (DHS) and the DHS Blade in both cadaveric femoral heads and artificial bone substitutes in the form of polyurethane foam blocks of different density. The pushout studies were performed as a means of comparing the force displacement curves produced by each implant within each material. RESULTS: The results demonstrated that test material with a density of 0.16 g/cm(3) (block A) produced qualitatively similar force displacement curves for the DHS and qualitatively and quantitatively similar force displacement curves for the DHS Blade, whereas the test material with a density of 0.08 g/cm(3) (block B) did not produce results that were predictive of those recorded within the osteoporotic cadaveric femoral heads. CONCLUSION: This study demonstrates that synthetic material with a density of 0.16 g/cm(3) can provide a good substitute for cadaveric osteoporotic femoral heads in the testing of implants. However we do recognise that no synthetic material can be considered as a definitive substitute for bone, therefore studies performed with artificial bone substrates may need to be validated by further testing with a small bone sample in order to produce conclusive results.
INTRODUCTION: The objective of this study was to determine if a synthetic bone substitute would provide results similar to bone from osteoporotic femoral heads during in vitro testing with orthopaedic implants. If the synthetic material could produce results similar to those of the osteoporotic bone, it could reduce or eliminate the need for testing of implants on bone. METHODS: Pushout studies were performed with the dynamic hip screw (DHS) and the DHS Blade in both cadaveric femoral heads and artificial bone substitutes in the form of polyurethane foam blocks of different density. The pushout studies were performed as a means of comparing the force displacement curves produced by each implant within each material. RESULTS: The results demonstrated that test material with a density of 0.16 g/cm(3) (block A) produced qualitatively similar force displacement curves for the DHS and qualitatively and quantitatively similar force displacement curves for the DHS Blade, whereas the test material with a density of 0.08 g/cm(3) (block B) did not produce results that were predictive of those recorded within the osteoporotic cadaveric femoral heads. CONCLUSION: This study demonstrates that synthetic material with a density of 0.16 g/cm(3) can provide a good substitute for cadaveric osteoporotic femoral heads in the testing of implants. However we do recognise that no synthetic material can be considered as a definitive substitute for bone, therefore studies performed with artificial bone substrates may need to be validated by further testing with a small bone sample in order to produce conclusive results.
Entities:
Keywords:
Artificial bone substitute; Biomechanical study; DHS; DHS Blade; Intertrochanteric fracture; Osteoporotic Bone
This study attempted to determine if readily available commercial
artificial bone substrates produce results similar to those from
osteoporotic cadaveric femoral heads in the testing of orthopaedic
implants, by testing an artificial bone substrate suggested in the
literature to demonstrate mechanical properties in the range of osteoporotic
cancellous boneArtificial bone substrate with a density of 0.16
g/cm3 produces similar results to those that were observed with
the osteoporotic cadaveric femoral heads, with similar force displacement
curves and peak forces. Therefore this material can be considered
as a good substitute for osteoporotic bone when testing orthopaedic implantsAlthough the artificial bone substrate with a density of 0.16
g/cm3 may be a good substitute for bone, we do recognise
that studies conducted with it may need to be validated by further
tests with cadaveric bone to produce indisputable results. However,
by conducting the initial aspect of a study with the 0.16 g/cm3 material,
the samples of cadaveric bone required may be greatly reducedOne of the strengths of this study is that it provides evidence
that the artificial bone substrate tested produces results that
are consistent with those achieved with a cadaveric bone sampleLiterature on the simulation of osteoporotic bone for mechanical
testing is sparse and this study may offer guidance to researchers
planning biomechanical studies using osteoporotic boneOne of the limitations of this study was in the sourcing of the
femoral heads, as it proved difficult to get sufficient numbers
of suitable osteoporotic femoral heads, due to a variety of reasons
Introduction
Biomechanical testing of implants plays a vital role in the evaluation
of any new implant technology.[1] Obtaining fresh
disease-free cadaveric bone to be used in mechanical testing of
orthopaedic implants is difficult and can be extremely expensive.[2] Another problem
is that cadaveric specimens are not uniform, resulting in the inclusion
of specimens with vastly heterogeneous bone quality and strength.[3-5] Large variations have been noted in
the shape and material properties of cadaveric bone, with sds in excess
of 100%.[2] Ranges
in the age and degree of osteoporosis of cadaver specimens can also
partially account for the variability in mechanical properties.[4,6] Therefore, this variability in the
geometric and material properties of cadaveric specimens often requires
prohibitively large sample sizes in order to detect statistically
significant differences in implant performance.With constraints regarding availability, handling and reproducibility
of cadaveric specimens, bone surrogate models have been introduced
for mechanical testing of fracture fixation implants.[3,4,7,8] These specimens,
such as Sawbones composite bones (Malmö, Sweden) exhibit known mechanical
characteristics with minimal variation, allowing statistically valid
comparisons to be made with much smaller sample sizes. It is widely
accepted that the performance of fracture fixation, including mechanisms of
failure, differ between strong and weak bone.[9-11] Several studies have confirmed that
current bone surrogates possess mechanical properties adequate to
evaluate the performance of implants in normal bone.[3,4,7,8] The mechanical properties
of composite femora have been shown to fall well within the range
for cadaveric specimens, with no significant differences being detected between
the synthetic femurs and cadaveric femurs.[3] The inter-femur variability for the
composite femurs was between 20 and 200 times lower than that for
the cadaveric specimens, thus allowing smaller differences to be characterised
as significant using the same sample size, with the use of composite
femora.[3]With the aim of providing a uniform test base, artificial bone
substrates in the form of polyurethane foam blocks have also been
developed.[2] These
materials can be sourced in different densities with varying associated
predetermined mechanical properties. These materials are manufactured
to provide consistent and uniform -materials with properties in
the range of human cancellous bone.The objective of this study was to determine if a synthetic bone
substitute could provide results similar to those from bone from
cadaveric osteoporotic femoral heads when testing orthopaedic implants.
The implants tested for the purpose of this study were the dynamic
hip screw (DHS) and the DHS Blade (both Synthes GmbH, Zuchwil, Switzerland),
which are used in the treatment of intertrochanteric fractures.
Materials and Methods
Ethical approval for this study was obtained from the -ethics
committee within the hospital and the university.The lag screw elements of the DHS screw and the DHS Blade were
chosen as the test implants, as these are inserted into osteoporotic
femoral heads during the fixation of intertrochanteric fractures.
A factor in choosing these devices was that a supply of osteoporotic
femoral heads should be available as these can be obtained after
hemiarthroplasty for intracapsular fracture of the femoral neck.Polyurethane foam blocks (Sawbones) were chosen to be the artificial
bone substrates. These blocks were 180 mm in length, 130 mm in width
and 40 mm in depth. Blocks of two different densities were used
(0.16 g/cm3 (block A) and 0.08 g/cm3 (block
B)), as these were two densities described as having properties
in the range of osteoporotic bone.[12] Patel et al[12] concluded that the material with
a density of 0.08 g/cm3 was more brittle than osteoporotic
bone and that the 0.16 g/cm3 material was a good alternative
for in vitro testing of osteoporotic bone. The
mechanical properties of both these materials are documented in
Table I.Mechanical properties of the polyurethane
test blocksThe femoral heads used in our study were harvested from patients
who had undergone a hemiarthroplasty at Midwestern Regional Hospital,
Limerick, for a fracture of the femoral neck and were classified
as osteoporotic on that basis. The 12 femoral heads were taken from 12 patients
(nine female and three male) with a mean age at operation of 75
years (sd 9.66). Protocol for tissue -storage was followed
after harvesting the femoral heads and they were stored in a bone
bank at -80° until required for testing. The femoral heads also
underwent dual--emission X-ray absorptiometry (DEXA) scanning prior
to testing, in order to determine their bone mineral density (BMD,
g/cm2). An equal number of femoral heads were initially
obtained for testing with each implant group and were assigned to
each implant after DEXA scanning, so that an equal number of femoral
heads of similar density would be tested with each implant. However,
difficulties encountered during testing and the limited supply of specimens
of bone resulted in unequal numbers of femoral heads in each group
at the end of testing. Some femo-ral heads being damaged at time
of extraction and were therefore unsuitable for testing, and on
other occasions problems were encountered when securing the femoral heads
in the customised testing rig. If a problem was encountered that
was considered to affect the results of testing, the femoral head
was discarded.Pushout studies were chosen as the experimental protocol for this study as
they are a good representation of the most common mode of failure
with these implants in clinical practice, which is ‘cut-out’ or migration
of the implants through the femoral heads until failure.[13] This has been
documented in the literature with previous studies using pushout
tests to investigate implant performance in relation to the treatment of
intertrochanteric fractures.[14-16] Pushout studies involve
pushing each implant through the test material and recording the
corresponding force-displacement curves produced.Pushout studies were performed in the osteoporotic femoral heads
and in the polyurethane blocks A and B. With the cadaveric femoral
heads, both implants were inserted as per clinical practice, with
the femoral heads being reamed and tapped to the correct depth for
the DHS and the DHS Blades being tapped into the correct position.
They were both placed within the bone to a target tip apex distance
of 25 mm.All the polyurethane test blocks were 40 mm in depth and each
implant was inserted to a depth of 30 mm. For the DHS each test
block was reamed and tapped to 30 mm with the DHS inserted to the
same depth. The DHS blade was tapped into each test block to a depth
of 30 mm as per manufacturers’ instructions.In order to perform the pushout tests, the implants with either
the synthetic bone blocks or the femoral heads were mounted on a
Tinius Olsen testing machine (Tinius Olsen Ltd, Redhill, United
Kingdom), as shown in Figure 1. Testing was standardised and conducted
at a rate of 2 mm/min. Tests were stopped after either dramatic
failure or when displacement > 7 mm was measured. Individual force
displacement curves were recorded for each test. For comparative
purposes, trendlines were created for each group tested by fitting
a sixth-order polynomial curve to the force displacement curves
for each group tested.Photograph showing the experimental
set-up for pushout testing of the implant in a synthetic bone construct.
Results
The results from testing using the DHS implant are shown in Table
II. The mean peak forces reached in block A and in the femoral heads
were similar (1035 N (sd 46) and 1377 N (sd 406.5),
respectively), while the mean peak force reached in block B was
much lower (305 N (sd 28)). The force displacement curves
for all three materials tested with the DHS reached their peak forces
at similar mean displacements of 1.48 mm (sd 0.12), 1.2
mm (sd 0.14) and 1.5 mm (sd 0.41) for block A,
block B and the femoral heads, respectively. The force displacement curves
for the three materials is shown in Figure 2a. The osteoporotic
cadaveric femoral heads and block A produce similar curves, whereas
block B produces a curve depicting substantially lower values than
those achieved with the cadaveric femoral heads.Force–displacement curves for the
pushout tests involving a) the dynamic hip screw (DHS) and b) the
DHS Blade in cadaveric femoral heads and -synthetic bone constructs
with densities of 0.16 g/cm3 (block A) and 0.08 g/cm3 (block
B).Peak forces and distance to peak
force for the dynamic hip screw (DHS) and DHS Blade in test blocks
A and B and in the femoral heads* BMD, bone mineral density (as measured on dual-emission
X-ray absorptiometry)The results from testing using the DHS Blade are shown in Table
II. The mean peak forces reached in block A and in the cadaveric
femoral heads were again similar (1020 N (sd 14) and 1302
N (sd 512.96), respectively), and the mean peak force achieved
by block B was again much lower, at 297 N (sd 16). However,
as shown by the force displacement curves in Figure 2b, the block
A and block B test materials reached their peak forces at mean displacements of 3.17 mm (sd 0.63)
and 3.19 mm (sd 0.52), respectively, while the cadaveric
femoral heads reached their peak forces at a mean of 5.89 mm (sd 2.14).
The force displacement curves for the osteoporotic femoral heads and block A can be seen
to be very similar, with block B providing a much shallower curve
(Fig. 2b).
Discussion
The objective of this study was to determine if a synthetic bone
substitute could provide results similar to those from cadaveric
osteoporotic femoral heads when testing orthopaedic implants. The
results suggest that block A, with a density of 0.16 g/cm3,
provides a good alternative test material to osteoporotic cadaveric
femoral heads, as it produced qualitatively similar force displacement
curves for the DHS (Fig. 2a) and qualitatively and quantitatively
similar force displacement curves for the DHS Blade (Fig. 2b). This
same is true for the DHS Blade, although it reaches its peak force
at a different point with the polyurethane test materials (means
of 3.17 mm (sd 0.63) for block A and 3.19 mm (sd 0.52)
for block B) than with the cadaveric femoral heads (mean of 5.89
mm (sd 2.14)). This may be due to the fact that peak force
is not well defined with the DHS Blade, as the force displacement
curve reaches a plateau early on in the curve and then continues
along this plateau until failure. Where the peak force occurs along
this curve can be therefore influenced by a small variation in the
resistance of the test material.Block B produced lower results than those recorded with the cadaveric
femoral heads (Fig. 2). It is recognised that osteoporotic bone
does not have one absolute value, but instead exists within a spectrum
of values. However, the results achieved with block B were consistently
below those achieved with the cadaveric femoral heads. This material
is therefore a less suitable substitute for osteoporotic bone than
the higher density block A (0.16 g/cm3). The most important
finding in this study is that in using these artificial test materials,
both implants produced force displacement curves that had a similar
pattern of failure as those produced with the femoral heads. As the
values of the forces reached before failure within block A were
similar to those achieved within the cadaveric femoral heads, it
proved to be a good substitute.Another point of discussion is that it is recognised that patients
who suffer an intertrochanteric fracture often are more osteoporotic
than those who suffer an intracapsular fracture.[17] As all the femoral
heads that were harvested for this study were collected after hemiarthroplasty
for intracapsular fractures, it is possible that the femoral heads
that these implants (the DHS and DHS Blade) will be used in will
be more osteoporotic than those used in this study. So the forces
achieved with these theoretically more osteoporotic femoral heads may
even more closely resemble those that were achieved with block A.The number of femoral heads used in this study was relatively
small and it is worth mentioning that it was not attempted to statistically
prove this association between the block A test material and the
cadaveric femoral heads. It is widely recognised that large interbone
material property variations exist and as the mechanical properties
of the cadaveric femoral heads also vary according to their bone
mineral density, a large number of femoral heads may be required
for any results to approach statistical significance. Therefore
this would be outside the capabilities of this study.Patel et al[12] had
previously investigated whether low--density polyurethane foams
are suitable for mimicking humanosteoporotic cancellous bone. However,
their study used quasi-static compression tests on polyurethane
foam cylinders of different densities (0.09 g/cm3, 0.16 g/cm3 and 0.32 g/cm3)
in order to determine the Young’s modulus, yield strength and energy
absorbed to yield. The results from these tests with the synthetic
polyurethane foam were then compared with the results of another
study by Li and Aspden,[18] which
investigated the material properties of -cancellous bone from patients
with osteoporosis. The comparison revealed that polyurethane foam
of 0.09 g/cm3 was much weaker than osteoporotic bone,
but that poly-urethane foam of 0.16 g/cm3 demonstrated
compressive Young’s modulus and yield strength values similar to
the osteoporotic bone that was tested in compression.[12]Further work has also been carried out on comparing synthetic
bone models with cadaveric bone, such as in studies by Cristofolini
et al[3] and
McNamara et al.[19] These studies
concentrated on comparing the mechanical properties of cadaveric
whole femurs against composite femurs, whereas our study focused
on proving that the artificial bone substrate tested was a good
substitute for osteoporotic bone in cadaveric femoral heads by comparing
the pattern of failure of the implants tested within them. We are
also aware that there may not be universal agreement about the use
of artificial materials as a substitute for cadaveric bone. Schoenfeld
et al[20] looked
at the pullout strength and load to failure properties of self--tapping
cortical screws in synthetic and cadaveric environments representative
of healthy and osteoporotic bone, and found that although trends
may be similar, screw performance in synthetic models was markedly
different from that in cadavers. However, we believe we have demonstrated
that block A (density 0.16 g/cm3) can be substituted for
osteoporotic cancellous bone and that the results produced with
it provide a good basis for further testing.
Conclusion
This study demonstrates that the 0.16 g/cm3 synthetic
material can provide a good substitute for osteoporotic bone within
cadaveric femoral heads when testing orthopaedic implants as the
force displacement curves produced are qualitatively and quantitatively -similar
to those produced within the osteoporotic cadaveric bone. We do
recognise however, that even though this 0.16 g/cm3 material
is a good substitute for this osteo-porotic bone, no synthetic material
can be considered as a definitive substitute for cadaveric osteoporotic bone
as large interbone material property variations exist with cadaveric
bone, therefore studies performed with artificial bone substrates
may need to be validated by further testing with a small bone sample
in order to produce conclusive results.This artificial bone substrate may then have a definite and valuable
role in designing, planning and initiating studies which may still
require a small number of bone samples to be tested in order to
produce definitive results.
Table I
Mechanical properties of the polyurethane
test blocks
Compressive
Tensile
Shear
Block
Density (g/cm3)
Strength (MPa)
Modulus (MPa)
Strength (MPa)
Modulus (MPa)
Strength (MPa)
Modulus (MPa)
A
0.16
2.2
58
2.1
86
1.6
19
B
0.08
0.60
16
1.0
32
0.59
7.1
Table II
Peak forces and distance to peak
force for the dynamic hip screw (DHS) and DHS Blade in test blocks
A and B and in the femoral heads
Mean (sd) peak force (N)
Mean (sd) distance to peak force
(mm)
Test material
Density
(g/cm3)
Mean (sd) BMD*
(g/cm2)
DHS
DHS Blade
DHS
DHS Blade
Test blocks
A (n = 7)
0.16
-
1035 (46)
1020 (14)
1.48 (0.12)
3.17 (0.63)
B (n = 7)
0.08
-
305.6 (28)
297 (16)
1.2 (0.14)
3.19 (0.52)
Femoral heads
DHS (n = 4)
-
0.912 (0.15)
1377 (406.6)
-
1.5 (0.41)
-
DHS Blade (n = 8)
-
0.784 (0.18)
-
1302 (513)
-
5.89 (2.14)
* BMD, bone mineral density (as measured on dual-emission
X-ray absorptiometry)
Authors: Arash Aminian; Fan Gao; Wasyl W Fedoriw; Li-Qun Zhang; David M Kalainov; Bradley R Merk Journal: J Orthop Trauma Date: 2007-09 Impact factor: 2.512
Authors: Daniel Bronsnick; Ryan E Harold; Ari Youderian; Giovanni Solitro; Farid Amirouche; Benjamin Goldberg Journal: Clin Orthop Relat Res Date: 2014-10-02 Impact factor: 4.176
Authors: Moritz F Lodde; J Christoph Katthagen; Clemens O Schopper; Ivan Zderic; Geoff Richards; Boyko Gueorguiev; Michael J Raschke; René Hartensuer Journal: J Clin Med Date: 2021-05-26 Impact factor: 4.241
Authors: Anderson Freitas; Gustavo Melo Torres; André Cezar de Andrade de Mello E Souza; Rafael Almeida Maciel; Diogo Ranier de Macedo Souto; George Neri de Barros Ferreira Journal: Rev Bras Ortop Date: 2014-10-27
Authors: Moritz F Lodde; J Christoph Katthagen; Clemens O Schopper; Ivan Zderic; R Geoff Richards; Boyko Gueorguiev; Michael J Raschke; René Hartensuer Journal: Medicina (Kaunas) Date: 2021-12-16 Impact factor: 2.430