Pullout strength and load to failure properties of self-tapping cortical screws in synthetic and cadaveric environments representative of healthy and osteoporotic bone.

Andrew J Schoenfeld1, Suneel Battula, Vivek Sahai, Gregory A Vrabec, Steven Corman, Lyndsey Burton, Glen O Njus.   

Abstract

BACKGROUND: The parameters of self-tapping screw (STS) performance in normal and osteoporotic bone have been defined in representative environments, but the question remains as to the clinical application of such findings. The goal of this study was to analyze the biomechanical performance of STSs in cadaveric and synthetic environments representative of healthy and osteoporotic bone.
METHODS: Ninety-six Synthes STSs were inserted into cadaveric and synthetic models representative of osteoporotic and healthy bone. Screws were inserted to depths of 1 mm short of the far cortex, flush and 1 mm and 2 mm beyond the far cortex. Screws were tested with an Instron 8511 material testing system utilizing axial pullout forces. A SAS procedure was used to conduct analysis of variance for unbalanced datasets.
RESULTS: Substantial differences were appreciated with respect to screw performance between osteoporotic and healthy bone specimens. Although a similar pattern of increased pullout strength and loading energy with increasing depth of insertion was demonstrated, absolute values were lower in osteoporotic specimens. Although performance trends were similar in cadaveric and synthetic testing models for both osteoporotic and healthy bone, values obtained during testing were different. Incomplete insertion of STSs resulted in a 21.5% and 37% reduction of biomechanical properties in osteoporotic and normal bone, respectively.
CONCLUSIONS: These results indicate that previously published findings on the performance of STSs in synthetic models cannot reasonably be applied to the clinical realm. Although trends may be similar, screw performance in synthetic, as compared with cadaveric, models is markedly different.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18469654     DOI: 10.1097/TA.0b013e318169cd71

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  10 in total

1.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

2.  Kirschner wire prepared pilot holes improve screw pullout strength in synthetic osteoporotic-type bone.

Authors:  Hrayr G Basmajian; Joseph N Liu; Travis Scudday; Seth T Campbell; Nirav H Amin
Journal:  J Clin Orthop Trauma       Date:  2019-08-22

3.  Embalmed and fresh frozen human bones in orthopedic cadaveric studies: which bone is authentic and feasible?

Authors:  Tobias Topp; Thorben Müller; Sebastian Huss; Peter Herbert Kann; Eberhard Weihe; Steffen Ruchholtz; Ralph Peter Zettl
Journal:  Acta Orthop       Date:  2012-09-14       Impact factor: 3.717

4.  BIOMECHANICAL COMPARISON OF MEDIAL VERSUS LATERAL SIDED PLATING IN FEMORAL FRACTURES.

Authors:  Firat Al; Bilgehan Tosun; Tamer Sinmazcelik; Mustafa Ozmen
Journal:  Acta Ortop Bras       Date:  2018       Impact factor: 0.513

5.  Factors Associated With the Accuracy of Depth Gauge Measurements.

Authors:  Pengcheng Liu; Joanna Xi Xiao; Chen Zhao; Xiaodong Li; Guantong Sun; Fei Yang; Xiaoqing Wang
Journal:  Front Surg       Date:  2022-01-13

6.  Distribution of hounsfield unit values in the pelvic bones: a comparison between young men and women with traumatic fractures and older men and women with fragility fractures: a retrospective cohort study.

Authors:  Naoya Inagaki; Takaaki Tanaka; Jun Udaka; Shoshi Akiyama; Tatsuki Matsuoka; Mitsuru Saito
Journal:  BMC Musculoskelet Disord       Date:  2022-03-29       Impact factor: 2.362

7.  Fixation stability comparison of bone screws based on thread design: buttress thread, triangle thread, and square thread.

Authors:  Xiaoreng Feng; Zhaopei Luo; Yupeng Li; Yiyi Yao; Weichen Qi; Bin Chen; Hongfeng Liang
Journal:  BMC Musculoskelet Disord       Date:  2022-08-30       Impact factor: 2.562

8.  Lateral migration resistance of screw is essential in evaluating bone screw stability of plate fixation.

Authors:  Xiaoreng Feng; Weichen Qi; Teng Zhang; Christian Fang; Hongfeng Liang; Bin Chen; Frankie Leung
Journal:  Sci Rep       Date:  2021-06-15       Impact factor: 4.379

9.  Validity of synthetic bone as a substitute for osteoporotic cadaveric femoral heads in mechanical testing: A biomechanical study.

Authors:  F O'Neill; F Condon; T McGloughlin; B Lenehan; C Coffey; M Walsh
Journal:  Bone Joint Res       Date:  2012-04-01       Impact factor: 5.853

Review 10.  Designs and techniques that improve the pullout strength of pedicle screws in osteoporotic vertebrae: current status.

Authors:  Thomas M Shea; Jake Laun; Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2014-03-03       Impact factor: 3.411

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.