Mechanical behavior of screws in normal and osteoporotic bone.

J Seebeck1, J Goldhahn, M M Morlock, E Schneider.   

Abstract

Fracture fixation in severe osteoporotic bone by means of implants that rely on screw anchorage is still a clinical problem. So far, a sufficiently accurate prediction of the holding capacity of screws as a function of local bone morphology has not been obtained. In this study the ultimate pullout loads of screws in the epi-, meta-, and diaphyseal regions of human tibiae were correlated to the cortical thicknesses and cancellous bone mineral densities at the screw axes determined from QCT densitometric data. Stepwise multiple linear regression showed that in regions with cortical thicknesses below 1.5 mm, cancellous density determined the ultimate pullout load (R2 = 0.85, p < 0.001), while in regions with cortices above 1.5 mm, cortical thickness alone significantly influenced the holding capacity of a screw (R2 = 0.90, p < 0.001). The findings of this study provide a basis for a bone morphology-related pre-operative estimation of the holding capacity of screws, which could help to improve their proper application in osteoporotic bone.

Entities:  

Mesh:

Year:  2004        PMID: 15517184     DOI: 10.1007/s00198-004-1777-0

Source DB:  PubMed          Journal:  Osteoporos Int        ISSN: 0937-941X            Impact factor:   4.507


  15 in total

1.  Determination of the role of the cancellous bone in generation of screw holding power at metaphysis.

Authors:  T Harnroongroj; A Techataweewan
Journal:  Clin Biomech (Bristol, Avon)       Date:  1999-06       Impact factor: 2.063

2.  Effect of bone quality on the forces generated by compression screws.

Authors:  K J Faran; N Ichioka; M A Trzeciak; S Han; J Medige; O J Moy
Journal:  J Biomech       Date:  1999-08       Impact factor: 2.712

3.  Bone strength at clinically relevant sites displays substantial heterogeneity and is best predicted from site-specific bone densitometry.

Authors:  Felix Eckstein; Eva-Maria Lochmüller; Christoph A Lill; Volker Kuhn; Erich Schneider; Günter Delling; Ralph Müller
Journal:  J Bone Miner Res       Date:  2002-01       Impact factor: 6.741

4.  Locking Compression Plate (LCP). An osteosynthesis plate based on the Dynamic Compression Plate and the Point Contact Fixator (PC-Fix).

Authors:  R Frigg
Journal:  Injury       Date:  2001-09       Impact factor: 2.586

5.  Estimation of mechanical properties of cortical bone by computed tomography.

Authors:  S M Snyder; E Schneider
Journal:  J Orthop Res       Date:  1991-05       Impact factor: 3.494

Review 6.  [Conventional radiology of osteoporosis and radiographic absorptiometry].

Authors:  M Jergas; G Schmid
Journal:  Radiologe       Date:  1999-03       Impact factor: 0.635

7.  European semi-anthropomorphic phantom for the cross-calibration of peripheral bone densitometers: assessment of precision accuracy and stability.

Authors:  J Pearson; P Ruegsegger; J Dequeker; M Henley; J Bright; J Reeve; W Kalender; D Felsenberg; A M Laval-Jeantet; J E Adams
Journal:  Bone Miner       Date:  1994-11

8.  Holding power of the 4.5 mm AO/ASIF cortex screw in cortical bone in relation to bone mineral.

Authors:  K Strømsøe; W L Kok; A Høiseth; A Alho
Journal:  Injury       Date:  1993-12       Impact factor: 2.586

9.  A comparative estimation of the combined cortical thickness of various bone sites.

Authors:  R A Bloom
Journal:  Skeletal Radiol       Date:  1980       Impact factor: 2.199

10.  Volumetric model determination of the tibia based on 2D radiographs using a 2D/3D database.

Authors:  P Messmer; G Long; N Suhm; P Regazzoni; A L Jacob
Journal:  Comput Aided Surg       Date:  2001
View more
  22 in total

1.  Short-term and long-term orthopaedic issues in patients with fragility fractures.

Authors:  Susan V Bukata; Stephen L Kates; Regis J O'Keefe
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

2.  Inadvertent screw stripping during ankle fracture fixation in elderly bone.

Authors:  A Feroz Dinah; Simon C Mears; Trevor A Knight; Sandeep P Soin; John T Campbell; Stephen M Belkoff
Journal:  Geriatr Orthop Surg Rehabil       Date:  2011-05

Review 3.  Currents of plate osteosynthesis in osteoporotic bone.

Authors:  Kalliopi Lampropoulou-Adamidou; Panagiotis K Karampinas; Efstathios Chronopoulos; John Vlamis; Demetrios S Korres
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-03-31

4.  [The biomechanics of screws, cerclage wire and cerclage cable].

Authors:  C Schröder; M Woiczinski; S Utzschneider; M Kraxenberger; P Weber; V Jansson
Journal:  Orthopade       Date:  2013-05       Impact factor: 1.087

5.  Preoperative assessment of the cancellous bone mineral density of the proximal humerus using CT data.

Authors:  Dietmar Krappinger; Tobias Roth; Martin Gschwentner; Armin Suckert; Michael Blauth; Clemens Hengg; Franz Kralinger
Journal:  Skeletal Radiol       Date:  2011-04-21       Impact factor: 2.199

6.  Early versus delayed weight bearing after surgical fixation of distal femur fractures: a non-randomized comparative study.

Authors:  Paolo Consigliere; Efthymios Iliopoulos; Tamer Ads; Alex Trompeter
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-07-02

7.  A surrogate long-bone model with osteoporotic material properties for biomechanical testing of fracture implants.

Authors:  Mark B Sommers; Daniel C Fitzpatrick; Steven M Madey; Corey Vande Zanderschulp; Michael Bottlang
Journal:  J Biomech       Date:  2007-06-18       Impact factor: 2.712

8.  Three-dimensional analyses of proximal humeral fractures using computed tomography with multiplanar reconstruction: early stability of fixation after osteosynthesis in relation to preoperative bone quality.

Authors:  Koki Ueda; Satoshi Ikemura; Akihisa Yamashita; Takashi Harada; Tetsuya Watanabe; Kenzo Shirasawa
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-09-17

9.  A demineralized calf vertebra model as an alternative to classic osteoporotic vertebra models for pedicle screw pullout studies.

Authors:  Atilla Akbay; Gokhan Bozkurt; Ozgur Ilgaz; Selcuk Palaoglu; Nejat Akalan; Edward C Benzel
Journal:  Eur Spine J       Date:  2007-11-17       Impact factor: 3.134

10.  Kirschner wire prepared pilot holes improve screw pullout strength in synthetic osteoporotic-type bone.

Authors:  Hrayr G Basmajian; Joseph N Liu; Travis Scudday; Seth T Campbell; Nirav H Amin
Journal:  J Clin Orthop Trauma       Date:  2019-08-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.