[Differences in the measurement results of Goldmann applanation tonometry with and without fluorescein].

N Arend1, C Hirneiss, M Kernt.   

Abstract

BACKGROUND: Goldmann applanation tonometry (GAT) is still the gold standard for measuring the intraocular pressure (IOP). Usually fluorescein eye drops are used additional to topical anesthesia to gain the best visualization results. The present study evaluated the differences in the results of GAT with and without fluorescein.
METHODS: A total of 400 eyes of 200 patients without known glaucoma were enrolled in this study and randomized to two groups: group A (first measurement without, second measurement with fluorescein) and group B (first measurement with fluorescein, second without). All measurements were performed by the same examiner with the same slit lamp. Results were analyzed by Bland-Altman plots and Spearman's correlation test.
RESULTS: The examined groups showed no significant differences regarding patient age, astigmatism or reason for consulting. In both groups performing GAT without the application of fluorescein led to significantly lower measurement results. The differences were 1.5 ± 1.7 mmHg SD in group A, 1.2 ± 1.6 mmHg SD in group B and 1.4 ± 1.65 mmHg SD for all eyes. The intraocular pressure (IOP) and the differences between the groups were independent of patient age, astigmatism, reason of consulting or IOP level. Both groups showed outliers up to 10 mmHg difference because of corneal edema.
CONCLUSIONS: Measurements from GAT without fluorescein differed significantly from measurements with fluorescein, independently of patient age, astigmatism, reason for consulting or IOP level. The mean difference of 1.4 mmHg may seem negligible in daily routine but can lead to serious consequences in borderline cases.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 23604252     DOI: 10.1007/s00347-013-2843-9

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  16 in total

1.  Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings.

Authors:  N Feltgen; D Leifert; J Funk
Journal:  Br J Ophthalmol       Date:  2001-01       Impact factor: 4.638

2.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

3.  Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis.

Authors:  Jun Liu; Cynthia J Roberts
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

Review 4.  [Glaucoma diagnostics and corneal thickness].

Authors:  J von Eicken; M Kohlhaas; H Höh
Journal:  Ophthalmologe       Date:  2005-09       Impact factor: 1.059

5.  Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.

Authors:  Markus Kohlhaas; Andreas G Boehm; Eberhard Spoerl; Antje Pürsten; Hans J Grein; Lutz E Pillunat
Journal:  Arch Ophthalmol       Date:  2006-04

6.  Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

Authors:  Stephen J Vincent; Roslyn A Vincent; David Shields; Graham A Lee
Journal:  Clin Exp Ophthalmol       Date:  2011-10-20       Impact factor: 4.207

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Goldmann applanation tonometry without fluorescein.

Authors:  D C Bright; J W Potter; D C Allen; R D Spruance
Journal:  Am J Optom Physiol Opt       Date:  1981-12

9.  Applanation tonometry and correction according to corneal thickness.

Authors:  R Stodtmeister
Journal:  Acta Ophthalmol Scand       Date:  1998-06

Review 10.  Systematic review of the agreement of tonometers with Goldmann applanation tonometry.

Authors:  Jonathan Alistair Cook; Adriana Paola Botello; Andrew Elders; Alia Fathi Ali; Augusto Azuara-Blanco; Cynthia Fraser; Kirsty McCormack; Jennifer Margaret Burr
Journal:  Ophthalmology       Date:  2012-05-10       Impact factor: 12.079

View more
  3 in total

1.  A comparison of NCT, Goldman application tonometry values with and without fluorescein.

Authors:  Hakika Erdogan; Ziya Akingol; Ozlem Cam; Sadik Sencan
Journal:  Clin Ophthalmol       Date:  2018-10-29

2.  [Pachymetry and intraocular pressure measurement by corneal visualization Scheimpflug technology (Corvis ST): A clinical comparison to the gold standard].

Authors:  J Steinberg; J Mehlan; A Frings; V Druchkiv; G Richard; T Katz; S J Linke
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

3.  Association of Diabetes With Central Corneal Thickness Among a Multiethnic Asian Population.

Authors:  Xiao-Yang Luo; Wei Dai; Miao-Li Chee; Yijin Tao; Jacqueline Chua; Nicholas Y Q Tan; Yih-Chung Tham; Tin Aung; Tien Yin Wong; Ching-Yu Cheng
Journal:  JAMA Netw Open       Date:  2019-01-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.