Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

Stephen J Vincent1, Roslyn A Vincent, David Shields, Graham A Lee.   

Abstract

BACKGROUND: To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients.
DESIGN: A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES: One hundred nine (54 males : 55 females) patients including only eyes under medical treatment for glaucoma.
METHODS: Measurement by Goldmann applanation tonometry, iCare rebound tonometry and 7CR non-contact tonometry. MAIN OUTCOME MEASURES: Intraocular pressure.
RESULTS: There were strong correlations between the intraocular pressure measurements obtained with Goldmann and both the rebound and non-contact tonometers (Spearman r-values ≥ 0.79, P < 0.001). However, there were small, statistically significant differences between the average readings for each tonometer. For the rebound tonometer, the mean intraocular pressure was slightly higher compared with the Goldmann applanation tonometer in the right eyes (P = 0.02), and similar in the left eyes (P = 0.93); however, these differences did not reach statistical significance. The Goldmann correlated measurements from the non-contact tonometer were lower than the average Goldmann reading for both right (P < 0.001) and left (P > 0.01) eyes. The corneal compensated measurements from the non-contact tonometer were significantly higher compared with the other tonometers (P ≤ 0.001).
CONCLUSIONS: The iCare rebound tonometer and the 7CR non-contact tonometer measure intraocular pressure in fundamentally different ways to the Goldmann applanation tonometer. The resulting intraocular pressure values vary between the instruments and will need to be considered when comparing clinical versus home acquired measurements.
© 2011 The Authors. Clinical and Experimental Ophthalmology © 2011 Royal Australian and New Zealand College of Ophthalmologists.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21883774     DOI: 10.1111/j.1442-9071.2011.02670.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  14 in total

1.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  [Differences in the measurement results of Goldmann applanation tonometry with and without fluorescein].

Authors:  N Arend; C Hirneiss; M Kernt
Journal:  Ophthalmologe       Date:  2014-03       Impact factor: 1.059

3.  COMPARISON OF IOP READINGS USING REBOUND I CARE TONOMETER AND PERKINS APPLANATION TONOMETER IN AN AFRICAN POPULATION.

Authors:  D S Ademola-Popoola; A F Odi; T M Akande
Journal:  J West Afr Coll Surg       Date:  2014 Jan-Mar

4.  Efficacy and safety of 1% forskolin eye drops in open angle glaucoma - An open label study.

Authors:  Muhammed Majeed; Kalyanam Nagabhushanam; Sankaran Natarajan; Priti Vaidyanathan; Suresh Kumar Karri; Jyolsna Agnes Jose
Journal:  Saudi J Ophthalmol       Date:  2015-03-05

5.  The utility of rebound tonometer in late elderly subjects.

Authors:  Rei Sakata; Jiro Numaga
Journal:  Clin Interv Aging       Date:  2014-01-22       Impact factor: 4.458

6.  Intraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea.

Authors:  Mohammad Reza Razeghinejad; Ramin Salouti; Mohammad Reza Khalili
Journal:  Iran J Med Sci       Date:  2014-01

7.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

8.  A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients.

Authors:  Yoshitaka Nakao; Yoshiaki Kiuchi; Satoshi Okimoto
Journal:  PLoS One       Date:  2017-01-17       Impact factor: 3.240

9.  Distribution of IOP measured with an air puff tonometer in a young population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Payam Nabovati; Negareh Yazdani; Hadi Ostadimoghaddam; Ehsan Shiralivand; Akbar Derakhshan; AbbasAli Yekta
Journal:  J Curr Ophthalmol       Date:  2017-04-02

10.  The effectiveness of schemes that refine referrals between primary and secondary care--the UK experience with glaucoma referrals: the Health Innovation & Education Cluster (HIEC) Glaucoma Pathways Project.

Authors:  Gokulan Ratnarajan; Wendy Newsom; Stephen A Vernon; Cecilia Fenerty; David Henson; Fiona Spencer; Yanfang Wang; Robert Harper; Andrew McNaught; Lisa Collins; Mike Parker; John Lawrenson; Robyn Hudson; Peng Tee Khaw; Richard Wormald; David Garway-Heath; Rupert Bourne
Journal:  BMJ Open       Date:  2013-07-21       Impact factor: 2.692

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.