Literature DB >> 21963617

The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study.

Mei Bai1, Bin Liu, Hongqiu Mu, Xianghua Liu, Yuan Jiang.   

Abstract

PURPOSE: To assess and compare the radiation dose of DynaCT in phantoms to that of MSCT. METHODS AND MATERIALS: A male Anderson Radiation Therapy 200 phantom with embedded thermoluminescence dosimeters (TLDs) was scanned with DynaCT (Axiom Artis dTA, Siemens Healthcare, Forchheim, Germany) and MSCTs (MSCT1: Lightspeed VCT, GE, Milwaukee, USA; MSCT2: Sensation Cardiac 64, Siemens Healthcare, Erlangen, Germany), respectively. For DynaCT, radiation exposure data with automatic exposure control were obtained from scanning of head, chest and abdomen. For MSCT, scanning protocols for head, chest, and abdomen were examined under conventional exposure conditions. Effective doses were calculated from LiF-TLD measurements according to ICRP103. A Catphan phantom (Phantom Lab, USA) was used to evaluate the spatial resolution and low contrast detectability of DynaCT and MSCT.
RESULTS: The effective doses for DynaCT from head, chest and abdomen scanning were 1.18, 7.32, 7.48 mSv (20 s scan) and 0.85, 6.01, 7.04 mSv (8 s scan); the effective doses for MSCT from head, chest and abdomen scanning were 3.33, 7.62 and 8.42 mSv (MSCT1), and 1.89, 7.52, 8.23 mSv (MSCT2). Significant difference between the organ doses from DynaCT and from MSCT (p<.05) was shown. The spatial resolution of 12 lp/cm was achieved and it was able to recognize a 3mm low contrast object at 0.5% contrast level in DynaCT, which was on the same level as in the MSCT images.
CONCLUSION: The phantom study indicates that DynaCT applies significantly less dose to patient and achieves similar spatial resolution and low contrast detectability to standard diagnostic MSCT.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21963617     DOI: 10.1016/j.ejrad.2011.09.006

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  27 in total

1.  New developed urological protocols for the Uro Dyna-CT reduce radiation exposure of endourological patients below the levels of the low dose standard CT scans.

Authors:  M-C Rassweiler; R Banckwitz; C Koehler; B Mueller-Allissat; M-S Michel; A Häcker; M Ritter
Journal:  World J Urol       Date:  2013-10-30       Impact factor: 4.226

2.  Influence of endourological devices on 3D reconstruction image quality using the Uro Dyna-CT.

Authors:  M-C Rassweiler; M Ritter; M-S Michel; A Häcker
Journal:  World J Urol       Date:  2012-08-03       Impact factor: 4.226

3.  Real-time image guidance in laparoscopic liver surgery: first clinical experience with a guidance system based on intraoperative CT imaging.

Authors:  Hannes G Kenngott; Martin Wagner; Matthias Gondan; Felix Nickel; Marco Nolden; Andreas Fetzer; Jürgen Weitz; Lars Fischer; Stefanie Speidel; Hans-Peter Meinzer; Dittmar Böckler; Markus W Büchler; Beat P Müller-Stich
Journal:  Surg Endosc       Date:  2013-11-01       Impact factor: 4.584

Review 4.  Imaging Techniques to Aid IR Treatment of Musculoskeletal Malignancy.

Authors:  A Kyle Jones; Steven Yevich
Journal:  Semin Intervent Radiol       Date:  2019-03-29       Impact factor: 1.513

5.  CT reconstruction and MRI fusion of 3D rotational angiography in the evaluation of pediatric cerebrovascular lesions.

Authors:  Prakash Muthusami; Nicholas Shkumat; Vanessa Rea; Albert H Chiu; Manohar Shroff
Journal:  Neuroradiology       Date:  2017-03-27       Impact factor: 2.804

6.  Feasibility of intravenous flat panel detector CT angiography for intracranial arterial stenosis.

Authors:  J S Jeon; S H Sheen; G J Hwang; H C Kim; B J Kwon
Journal:  AJNR Am J Neuroradiol       Date:  2012-08-30       Impact factor: 3.825

7.  Accuracy and speed of robotic assisted needle interventions using a modern cone beam computed tomography intervention suite: a phantom study.

Authors:  Boris Schulz; Katrin Eichler; Petra Siebenhandl; Tatjana Gruber-Rouh; Christoph Czerny; Thomas Josef Vogl; Stephan Zangos
Journal:  Eur Radiol       Date:  2012-07-21       Impact factor: 5.315

8.  Effective dose span of ten different cone beam CT devices.

Authors:  D Rottke; S Patzelt; P Poxleitner; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-04-12       Impact factor: 2.419

9.  Efficacy of DynaCT for surgical navigation during complex laparoscopic surgery: an initial experience.

Authors:  Tetsuo Nozaki; Yasuyoshi Fujiuchi; Akira Komiya; Hideki Fuse
Journal:  Surg Endosc       Date:  2012-10-06       Impact factor: 4.584

10.  Three-dimensional rotational angiography in the assessment of vascular and airway compression in children after a cavopulmonary anastomosis.

Authors:  Sharon Borik; Sabina Volodina; Rajiv Chaturvedi; Kyong Jin Lee; Lee N Benson
Journal:  Pediatr Cardiol       Date:  2015-03-12       Impact factor: 1.655

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.