Literature DB >> 23566606

Impact of operator experience on the accuracy of implant placement with stereolithographic surgical templates: an in vitro study.

Sarra E Cushen1, Ilser Turkyilmaz.   

Abstract

STATEMENT OF PROBLEM: To achieve functional and esthetic results, implants must be placed accurately. However, little information relating to the effect of operator experience on implant placement accuracy is available.
PURPOSE: The objective of this investigation was to measure the accuracy of dental implant placement with a bone-supported stereolithographic surgical template created from a virtual implant plan and to determine the effect of operator experience on implant placement accuracy.
MATERIAL AND METHODS: Twenty photopolymer resin edentulous mandibles were scanned with cone beam computerized tomography (CBCT). Five implants were planned virtually for each mandible, and a stereolithographic surgical template was made. Four operators placed a total of 100 implants (25 per operator). Two of the operators were experienced in implant placement and 2 operators had limited prior implant placement experience. A CBCT scan of the postimplant placement mandibles was performed, and the images were superimposed on the preimplant placement images containing the virtual implant plans. The amount of angular, horizontal, and vertical deviation of the placed implants from the virtually planned implants at the apex and platform was calculated, and statistically significant differences were detected between the operator groups by using a multivariate analysis of variance (MANOVA) (α=.05).
RESULTS: For the experienced operators, the mean error of angular deviation was 2.60 ±1.25 degrees, of horizontal deviation at the apex 0.34 ±0.15 mm, of horizontal deviation at the platform 0.63 ±0.28 mm, of vertical deviation at the apex 0.59 ±0.12 mm, and of vertical deviation at the platform 0.16 ±0.11 mm. For the inexperienced group, the mean error of angular deviation was 3.96 ±1.64 degrees, of horizontal deviation at the apex 0.42 ±0.19 mm, of horizontal deviation at the platform 0.77 ±0.33 mm, of vertical deviation at the apex 0.62 ±0.13 mm, and of vertical deviation at the platform 0.15 ±0.11 mm. The MANOVA showed a statistically significant difference between the experienced and inexperienced groups for angular and horizontal error at the implant apex and platform (P<.05).
CONCLUSIONS: The results of this in vitro investigation revealed that the experience level of the operator placing the implants contributes to the accuracy of implant placement, with more experienced operators placing more implants accurately.
Copyright © 2013 The Editorial Council of the Journal of Prosthetic Dentistry. Published by Mosby, Inc. All rights reserved.

Mesh:

Substances:

Year:  2013        PMID: 23566606     DOI: 10.1016/S0022-3913(13)60053-0

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  12 in total

1.  Influence of 2D vs 3D imaging and professional experience on dental implant treatment planning.

Authors:  João Henrique Fortes; Christiano de Oliveira-Santos; Wilson Matsumoto; Raphael Jurca Gonçalves da Motta; Camila Tirapelli
Journal:  Clin Oral Investig       Date:  2018-06-16       Impact factor: 3.573

2.  Digital implant planning and guided implant surgery - workflow and reliability.

Authors:  O Schubert; J Schweiger; M Stimmelmayr; E Nold; J-F Güth
Journal:  Br Dent J       Date:  2019-01-25       Impact factor: 1.626

Review 3.  Peri-implantitis. Part 1: Scope of the problem.

Authors:  A Alani; M Kelleher; K Bishop
Journal:  Br Dent J       Date:  2014-09       Impact factor: 1.626

4.  Computer-aided flapless implant surgery and immediate loading. A technical note.

Authors:  Alfonso Baruffaldi; Pier Paolo Poli; Andrea Baruffaldi; Lorenzo Giberti; Maurizio Pigozzo; Carlo Maiorana
Journal:  Oral Maxillofac Surg       Date:  2016-04-05

5.  [Precise investigation of digital guide plates applied to implant surgery of anterior teeth].

Authors:  Fang-Zhi Lou; Mao-Rui Zhang; Peng-Cheng Rao; Shi-Hong Luo; Meng-Ying Tang; Jin-Gang Xiao
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-04-01

Review 6.  Advantages and limitations of implant surgery with CAD/CAM surgical guides: A literature review.

Authors:  Gokce-Soganci Unsal; Ilser Turkyilmaz; Samantha Lakhia
Journal:  J Clin Exp Dent       Date:  2020-04-01

7.  Accuracy of Linear Measurements Using Cone Beam Computed Tomography in Comparison with Clinical Measurements.

Authors:  Amir Reza Rokn; Kazem Hashemi; Solmaz Akbari; Mohammad Javad Kharazifard; Hamidreza Barikani; Mehrdad Panjnoosh
Journal:  J Dent (Tehran)       Date:  2016-09

8.  The Use of Orientation Templates and Free-Hand Implant Insertion in Artificial Mandibles-An Experimental Laboratory Examination in Fifth-Year Dental Students.

Authors:  Matthias C Schulz; Lena Rittmann; Ursula Range; Günter Lauer; Dominik Haim
Journal:  Dent J (Basel)       Date:  2018-09-01

9.  Inferior vena cava/aorta diameter index in the assessment of the body fluid status - a comparative study of measurements performed by experienced and inexperienced examiners in a group of young adults.

Authors:  Kaja Durajska; Emilia Januszkiewicz; Łukasz Szmygel; Wojciech Kosiak
Journal:  J Ultrason       Date:  2014-09-30

10.  Accuracy of implant surgery with surgical guide by inexperienced clinicians: an in vitro study.

Authors:  Takeshi Toyoshima; Hideaki Tanaka; Masanori Sasaki; Eiji Ichimaru; Yasushi Naito; Yasuyuki Matsushita; Kiyoshi Koyano; Seiji Nakamura
Journal:  Clin Exp Dent Res       Date:  2015-07-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.