X-ray phase contrast imaging of the breast: analysis of tissue simulating materials.

Srinivasan Vedantham1, Andrew Karellas.   

Abstract

PURPOSE: Phase contrast imaging, particularly of the breast, is being actively investigated. The purpose of this work is to investigate the x-ray phase contrast properties of breast tissues and commonly used breast tissue substitutes or phantom materials with an aim of determining the phantom materials best representative of breast tissues.
METHODS: Elemental compositions of breast tissues including adipose, fibroglandular, and skin were used to determine the refractive index, n = 1 - δ + i β. The real part of the refractive index, specifically the refractive index decrement (δ), over the energy range of 5-50 keV were determined using XOP software (version 2.3, European Synchrotron Radiation Facility, France). Calcium oxalate and calcium hydroxyapatite were considered to represent the material compositions of microcalcifications in vivo. Nineteen tissue substitutes were considered as possible candidates to represent adipose tissue, fibroglandular tissue and skin, and four phantom materials were considered as possible candidates to represent microcalcifications. For each material, either the molecular formula, if available, or the elemental composition based on weight fraction, was used to determine δ. At each x-ray photon energy, the absolute percent difference in δ between the breast tissue and the substitute material was determined, from which three candidates were selected. From these candidate tissue substitutes, the material that minimized the absolute percent difference in linear attenuation coefficient μ, and hence β, was considered to be best representative of that breast tissue.
RESULTS: Over the energy range of 5-50 keV, while the δ of CB3 and fibroglandular tissue-equivalent material were within 1% of that of fibroglandular tissue, the μ of fibroglandular tissue-equivalent material better approximated the fibroglandular tissue. While the δ of BR10 and adipose tissue-equivalent material were within 1% of that of adipose tissue, the tissue-equivalent material better approximated the adipose tissue in terms of μ. Polymethyl methacrylate, a commonly used tissue substitute, exhibited δ greater than fibroglandular tissue by ≈ 12%. The A-150 plastic closely approximated the skin. Several materials exhibited δ between that of adipose and fibroglandular tissue. However, there was an energy-dependent mismatch in terms of equivalent fibroglandular weight fraction between δ and μ for these materials. For microcalcifications, aluminum and calcium carbonate were observed to straddle the δ and μ of calcium oxalate and calcium hydroxyapatite. Aluminum oxide, commonly used to represent microcalcifications in the American College of Radiology recommended phantoms for accreditation exhibited δ greater than calcium hydroxyapatite by ≈ 23%.
CONCLUSIONS: A breast phantom comprising A-150 plastic to represent the skin, commercially available adipose and fibroglandular tissue-equivalent formulations to represent adipose and fibroglandular tissue, respectively, was found to be best suited for x-ray phase-sensitive imaging of the breast. Calcium carbonate or aluminum can be used to represent microcalcifications.

Entities:  

Mesh:

Year:  2013        PMID: 23556900      PMCID: PMC3612120          DOI: 10.1118/1.4794503

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  32 in total

1.  Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment.

Authors:  J M Boone
Journal:  Radiology       Date:  1999-10       Impact factor: 11.105

2.  X-ray scattering from human breast tissues and breast-equivalent materials.

Authors:  M E Poletti; D Gonçalves; I Mazzaro
Journal:  Phys Med Biol       Date:  2002-01-07       Impact factor: 3.609

3.  Flat-panel digital mammography system: contrast-detail comparison between screen-film radiographs and hard-copy images.

Authors:  Sankararaman Suryanarayanan; Andrew Karellas; Srinivasan Vedantham; Hetal Ved; Stephen P Baker; Carl J D'Orsi
Journal:  Radiology       Date:  2002-12       Impact factor: 11.105

4.  Multicontrast x-ray computed tomography imaging using Talbot-Lau interferometry without phase stepping.

Authors:  Nicholas Bevins; Joseph Zambelli; Ke Li; Zhihua Qi; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

5.  Monte Carlo and phantom study of the radiation dose to the body from dedicated CT of the breast.

Authors:  Ioannis Sechopoulos; Srinivasan Vedantham; Sankararaman Suryanarayanan; Carl J D'Orsi; Andrew Karellas
Journal:  Radiology       Date:  2008-02-21       Impact factor: 11.105

6.  Classification of breast computed tomography data.

Authors:  Thomas R Nelson; Laura I Cerviño; John M Boone; Karen K Lindfors
Journal:  Med Phys       Date:  2008-03       Impact factor: 4.071

7.  The myth of the 50-50 breast.

Authors:  M J Yaffe; J M Boone; N Packard; O Alonzo-Proulx; S Y Huang; C L Peressotti; A Al-Mayah; K Brock
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

8.  Two-dimensional grating-based X-ray phase-contrast imaging using Fourier transform phase retrieval.

Authors:  Hidenosuke Itoh; Kentaro Nagai; Genta Sato; Kimiaki Yamaguchi; Takashi Nakamura; Takeshi Kondoh; Chidane Ouchi; Takayuki Teshima; Yutaka Setomoto; Toru Den
Journal:  Opt Express       Date:  2011-02-14       Impact factor: 3.894

9.  Mammography with synchrotron radiation: first clinical experience with phase-detection technique.

Authors:  Edoardo Castelli; Maura Tonutti; Fulvia Arfelli; Renata Longo; Emilio Quaia; Luigi Rigon; Daniela Sanabor; Fabrizio Zanconati; Diego Dreossi; Alessando Abrami; Elisa Quai; Paola Bregant; Katia Casarin; Valentina Chenda; Ralf Hendrik Menk; Tatjana Rokvic; Alessandro Vascotto; Giuliana Tromba; Maria Assunta Cova
Journal:  Radiology       Date:  2011-03-24       Impact factor: 11.105

10.  Technical note: Skin thickness measurements using high-resolution flat-panel cone-beam dedicated breast CT.

Authors:  Linxi Shi; Srinivasan Vedantham; Andrew Karellas; Avice M O'Connell
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

View more
  8 in total

1.  Emerging Breast Imaging Technologies on the Horizon.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  Semin Ultrasound CT MR       Date:  2017-09-13       Impact factor: 1.875

2.  Towards standardization of x-ray beam filters in digital mammography and digital breast tomosynthesis: Monte Carlo simulations and analytical modelling.

Authors:  Suman Shrestha; Srinivasan Vedantham; Andrew Karellas
Journal:  Phys Med Biol       Date:  2017-01-11       Impact factor: 3.609

3.  Large-angle x-ray scatter in Talbot-Lau interferometry for breast imaging.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas
Journal:  Phys Med Biol       Date:  2014-10-08       Impact factor: 3.609

4.  A phase sensitive x-ray breast tomosynthesis system: Preliminary patient images with cancer lesions.

Authors:  Muhammad U Ghani; Laurie L Fajardo; Farid Omoumi; Aimin Yan; Peter Jenkins; Molly Wong; Yuhua Li; Michael E Peterson; Edward J Callahan; Stephen L Hillis; Bin Zheng; Xizeng Wu; Hong Liu
Journal:  Phys Med Biol       Date:  2021-10-29       Impact factor: 3.609

5.  Development and preclinical evaluation of a patient-specific high energy x-ray phase sensitive breast tomosynthesis system.

Authors:  Muhammad U Ghani; Xizeng Wu; Laurie L Fajardo; Zhengxue Jing; Molly D Wong; Bin Zheng; Farid Omoumi; Yuhua Li; Aimin Yan; Peter Jenkins; Stephen L Hillis; Laura Linstroth; Hong Liu
Journal:  Med Phys       Date:  2021-04-01       Impact factor: 4.071

6.  Phantom-Based Feasibility Studies on Phase-Contrast Mammography at Indian Synchrotron Facility Indus-2.

Authors:  Reena Sharma; S D Sharma; P S Sarkar; B Singh; A K Agrawal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar

7.  Exploring phase contrast imaging with a laser-based Kα x-ray source up to relativistic laser intensity.

Authors:  M Gambari; R Clady; A Stolidi; O Utéza; M Sentis; A Ferré
Journal:  Sci Rep       Date:  2020-04-21       Impact factor: 4.379

8.  High-resolution μ CT imaging for characterizing microcalcification detection performance in breast CT.

Authors:  Andrew M Hernandez; Amy E Becker; Su Hyun Lyu; Craig K Abbey; John M Boone
Journal:  J Med Imaging (Bellingham)       Date:  2021-07-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.