Swept source/Fourier domain OCT is demonstrated for in vivo imaging of the rodent eye. Using commercial swept laser technology, we developed a prototype OCT imaging system for small animal ocular imaging operating in the 1050 nm wavelength range at an axial scan rate of 100 kHz with ~6 µm axial resolution. The high imaging speed enables volumetric imaging with high axial scan densities, measuring high flow velocities in vessels, and repeated volumetric imaging over time. The 1050 nm wavelength light provides increased penetration into tissue compared to standard commercial OCT systems at 850 nm. The long imaging range enables multiple operating modes for imaging the retina, posterior eye, as well as anterior eye and full eye length. A registration algorithm using orthogonally scanned OCT volumetric data sets which can correct motion on a per A-scan basis is applied to compensate motion and merge motion corrected volumetric data for enhanced OCT image quality. Ultrahigh speed swept source OCT is a promising technique for imaging the rodent eye, proving comprehensive information on the cornea, anterior segment, lens, vitreous, posterior segment, retina and choroid.
Swept source/Fourier domain OCT is demonstrated for in vivo imaging of the rodent eye. Using commercial swept laser technology, we developed a prototype OCT imaging system for small animal ocular imaging operating in the 1050 nm wavelength range at an axial scan rate of 100 kHz with ~6 µm axial resolution. The high imaging speed enables volumetric imaging with high axial scan densities, measuring high flow velocities in vessels, and repeated volumetric imaging over time. The 1050 nm wavelength light provides increased penetration into tissue compared to standard commercial OCT systems at 850 nm. The long imaging range enables multiple operating modes for imaging the retina, posterior eye, as well as anterior eye and full eye length. A registration algorithm using orthogonally scanned OCT volumetric data sets which can correct motion on a per A-scan basis is applied to compensate motion and merge motion corrected volumetric data for enhanced OCT image quality. Ultrahigh speed swept source OCT is a promising technique for imaging the rodent eye, proving comprehensive information on the cornea, anterior segment, lens, vitreous, posterior segment, retina and choroid.
Entities:
Keywords:
(170.3880) Medical and biological imaging; (170.4470) Ophthalmology; (170.4500) Optical coherence tomography
Rat and mouse models are essential for advancing our knowledge of eye development, evolution,
physiology, disease, and therapeutics. Rodent models have contributed to understanding the genetics
and mechanisms of diseases such as cataracts, glaucoma, age-related macular degeneration, diabetic
retinopathy, retinal degeneration, retinoblastoma, keratitis, and myopia [1]. Enucleation and histology are the gold standard for characterizing morphology
and anatomy of the small animal eye, but can only be performed after sacrificing the animal. For
disease progression studies using histological examinations, each animal represents only a single
time point. Noninvasive in vivo examination of the rodent eye enables monitoring disease progression
through its entire course in individual animals, which could provide more accurate monitoring of
disease onset, progression, and response to therapies.Current imaging methods for rodent eyes include fundus photography, fluorescein angiography,
slit-lamp imaging, scanning laser ophthalmoscope, magnetic resonance imaging and ultrasound
biomicroscopy [2-8]. Most of these methods only reveal two-dimensional information. Magnetic resonance
imaging and ultrasound can perform three-dimensional (3-D) imaging, but resolution is limited.
Optical coherence tomography (OCT) is a noninvasive imaging modality that can perform
high-resolution imaging of tissue morphology in situ and in real time [9]. OCT is analogous to ultrasound except that it measures the echo time delay and
magnitude of backreflected or backscattered light. By scanning an optical beam on tissue,
two-dimensional cross sectional and three-dimensional volumetric images can be generated.
Spectral/Fourier domain OCT (SD-OCT) imaging of the rodent retina was first reported by Srinivasan
et al. in 2006 [10]. SD-OCT offers excellent resolution, but
has limited imaging speed and imaging range due to spectrometer and camera performance limitations.
Current standard commercial OCT systems for small animal imaging are based on 850 nm SD-OCT
technology [10-12] and have been used in studies for imaging the retina and limited portions of the
anterior segment such as the cornea [13-15]. SD-OCT enables high resolution imaging of the rodent retina
using both 850 nm and 1050 nm wavelengths [10-16], but the limited imaging range is insufficient for imaging the
whole anterior segment of the rodent eye. Recent advances in swept source/Fourier domain OCT
(SS-OCT) enable in vivo ultrahigh speed imaging, offering a promising new technology for rodent eye
imaging [17,18].
SS-OCT offers several advantages over SD-OCT, including less sensitivity variation with imaging
depth, longer imaging range, higher detection efficiencies and reduced fringe washout. Furthermore,
increased penetration into the choroid and optic nerve head with reduced sensitivity to ocular
opacities has been reported for OCT imaging in the water absorption window at 1050 nm compared to
standard 850 nm wavelengths [19,20].Standard commercial SD-OCT systems for small animal imaging operate at the 850 nm wavelength
range with imaging speeds up to 40,000 axial scans per second, 3-7 µm resolution, and <2.5
mm imaging range in tissue. The systems have interchangeable sample arm optics that enables scanning
the small animal retina and anterior eye in different imaging modes. Several studies have been
performed using commercial SD-OCT systems [13-15]. However, the limited imaging range of the systems constrain
the imaging modes and require retinal, corneal, anterior lens and posterior lens imaging to be
performed separately, thereby making it difficult to perform simultaneous imaging of these ocular
structures.This study demonstrates OCT imaging in the rodent eye using a 1050 nm wavelength SS-OCT prototype
system at 100,000 axial scans per second with ~6 µm axial resolution and >5.3 mm imaging
range. In comparison to the human eye, the rodent eye has a large lens filling up most of the
vitreous cavity, making it challenging to image the anterior segment. The OCT system described here
is especially suited for rodent eye imaging because of the long imaging range which can accommodate
typical rodent eye lengths of ~6 mm in rats and ~3 mm in mice. OCT imaging protocols using
high-definition, high transverse pixel density three-dimensional imaging with dense raster scanning
are demonstrated. Long imaging ranges and improved sensitivity roll-off enable comprehensive imaging
of the anterior eye along with sections of the posterior lens surface and retina simultaneously.
Long wavelength OCT systems at 1050 nm improve visualization of deeper tissue structure compared to
850 nm where standard commercial OCT systems operate. The high speed and deep imaging penetration is
ideal for Doppler OCT of posterior eye blood vessels. High speed also enables four-dimensional (4-D)
time resolved volumetric imaging of dynamic responses of the eye.
2. Methods
2.1. Ultrahigh speed swept source/Fourier domain system
An ultrahigh speed swept source/Fourier domain OCT instrument was built for small animal imaging
(Fig. 1A
). A short external cavity, tunable light source (Axsun Technologies, Inc.) centered at 1044
nm which had a 3-dB bandwidth of 103 nm and a 10-dB bandwidth of 111 nm was used (Fig. 1B). The axial scan rate of the OCT system was 100 kHz, set by
the sweep rate of the laser. Light from the laser was split into a single pass reference arm and a
pre-objective scanning sample arm where the galvanometric scanners are placed in the back focal
plane of the objective lenses. Returning light from the reference and sample arms was combined in a
second (50:50) fiber coupler and the interferometric signal was detected using a low distortion 330
MHz dual balanced photodetector receiver (prototype; Thorlabs, Inc.). The signal from the
photodetector was digitized by a high-speed 8 bit analog-to-digital converter at 1 GSPS (ATS9870;
Alazar Technologies, Inc.). Different scan lenses in the sample arm were used to provide different
transverse spot sizes and depths of focus. A long working distance infrared microscope objective (M
Plan NIR 5X; Mitutoyo Corp.) was used for retinal imaging. The transverse spot size was ~12
µm full width at half maximum (FWHM) measured in air with a beam profiling camera and the
incident power was 1.6 mW. For anterior eye and full eye length imaging as well as for posterior eye
imaging, a 75 mm focal length achromatic lens was used. The transverse spot size was ~21 µm
FWHM in air and the incident power was 2.5 mW. Unlike human eye imaging, a telecentric scanning
interface was used for both anterior segment and retinal imaging of the rodent eye. A #1.5 coverslip
and index matching gel (Goniosol) was used over the corneal during the retinal and posterior eye
imaging procedures. An LED light stimulus was placed adjacent to the sample arm interface without
blocking the OCT beam. The measured axial resolution of the system was 6 µm in tissue. The
measured sensitivity of the system was 102 dB with a Nyquist limited depth range of 9.1 mm in tissue
where the −6-dB roll-off depth was at 2.5 mm in tissue and the −20-dB roll-off depth
was at 5.3mm in tissue. The sensitivity roll-off was limited by the bandwidth of the balanced
photodetector and finite light source coherence length (the roll-off from the coherence length
itself introduces a 6 dB sensitivity drop at 6 mm in air).
Fig. 1
(A) Schematic of ultrahigh speed swept source/Fourier domain OCT instrument. GS –
galvanometric scanners, RR – retroreflector, LS – LED light stimulus, DBP –
dual balanced photodetector, A/D – analog-to-digital converter, TRG – trigger signal.
(B) Spectrum of the light source. (C) Point spread function showing axial resolution. (D)
Sensitivity roll-off.
(A) Schematic of ultrahigh speed swept source/Fourier domain OCT instrument. GS –
galvanometric scanners, RR – retroreflector, LS – LED light stimulus, DBP –
dual balanced photodetector, A/D – analog-to-digital converter, TRG – trigger signal.
(B) Spectrum of the light source. (C) Point spread function showing axial resolution. (D)
Sensitivity roll-off.
2.2. Animal preparation
Sprague-Dawley rats and C57BL/6 mice were used to demonstrate the imaging capability of the OCT
system. Animals were anesthetized intraperitoneally with ketamine (40-80 mg/kg body weight) and
xylazine (5-10 mg/kg body weight) for all structural imaging studies. During anterior eye and full
eye length imaging, artificial tear drops were applied to prevent cornea dehydration. To perform
retinal and posterior eye imaging, eyes were dilated with topically applied tropicamide (1%) drops,
and a thin microscope coverslip was placed on the cornea with Hydroxypropyl methylcellulose
(Goniosol, 2.5%) to remove corneal refraction and preserve corneal hydration. To perform functional
imaging studies, imaging the pupillary reflex, animals were anesthetized intraperitoneally by a
cocktail containing ketamine (40 mg/kg body weight), xylazine (3 mg/kg body weight), and
acepromazine (1.5 mg/kg body weight). These studies were in compliance with the guidelines of the
ARVO Statement for the Use of Animals in Ophthalmic and Vision Research, and performed under a
protocol approved by the MIT Committee on Animal Care. After anesthetization, the animal was placed
in a comfortable mounting tube fixed on a height and tilt adjustable stage pivoted about the
animal's eye.
2.3. OCT image acquisition
OCT cross-sectional previews along with a real-time OCT en face fundus image were used for
alignment. For retinal imaging, six raster scans with orthogonal fast scan axis orientation
(horizontal and vertical) each consisting of 700 × 700 axial scans requiring an acquisition
time of ~5 seconds for each raster (~30 seconds total) were acquired. The six volumetric data sets
were then registered using motion correction software and merged. For anterior eye and full eye
length imaging, two raster scans with orthogonal fast scan axes (horizontal and vertical) consisting
of 500 × 500 axial scans requiring an acquisition time of <3 seconds each were acquired,
motion corrected and merged. For posterior eye and Doppler imaging, six orthogonal 500 × 500
axial scan raster scans were acquired, motion corrected, and merged, where each scan was acquired in
<3 seconds. For high-speed dynamic volumetric OCT imaging, repeated raster scans with 100
× 100 axial scans were taken in sequence to achieve ~10 volumes per second. The high speed
repeated volumes were not motion corrected.
2.4. Doppler OCT
Doppler OCT is a functional extension of OCT which provides velocity and flow information. In
Fourier domain OCT, the complex OCT spectral data is processed and the phase information is used for
velocity and flow calculations in Doppler OCT [21-25]. A simple phase subtraction of
neighboring oversampled OCT axial scans generates quantitative Doppler OCT measurements which was
used to visualize the vasculature in eyes:where λ0 is the center wavelength of the light,
n is the refractive index, f is the sweep rate, and
φ(z) is the phase profile of the
ith axial scan after Fourier transformation. The maximum detectable velocity before
phase wrapping, determined by the imaging speed, is ±20 mm/s in tissue. In theory, the
minimum detectable axial velocity is determined by the phase stability of the light source, and was
measured to be 1.8 mrad. In practice, the minimum measurable axial velocity in tissue is limited by
the phase decorrelation associated with trigger jitter and scanning the beam [25]. After applying a phase compensation algorithm where bulk motion was calculated
and removed using a histogram-based method [23], the standard
deviation of phase differences between successive sweeps over a mirror was measured to be 0.11 rad,
corresponding to a minimum measurable axial velocity of 0.7 mm/s in tissue.
2.5. OCT volumetric registration
Although the imaging was performed with the animals under anesthesia, motion artifacts persist
from breathing and heartbeat. Motion during the acquisition of volumetric OCT data distorts the data
and is a source of error in quantitative measurements. A motion correction algorithm was recently
developed for orthogonally scanned volumetric OCT data [26].
Motion correction is performed by estimating dense displacement fields, which describe the motion of
each A-scan, for each input volume and using the time structure of the acquisition process as a
constraint. After optimizing a global objective function, the displacement fields are estimated for
each volume to correct for motion, then a single volume is constructed by merging each motion
corrected volume. Motion corrected volumes do not show visible motion artifacts and a merged
registered volume has improved signal quality. The motion corrected, merged volumetric data more
accurately represents structure and morphology than individual volumes which can have motion
artifacts.
2.6. Light stimulation
For the pupil response experiments, a white light LED stimulus with ~800 cd/cm2
luminance was used. The LED light source was placed adjacent to the OCT beam without blocking the
OCT scan. Two stimulus protocols were used: a continuous >5 second stimulus and a short ~1 second
flash stimulus. The stimulus was activated during OCT data acquisition of repeated raster scans.
3. Results
3.1. 3-D retinal imaging in unpigmented rat eye and pigmented mouse eye
Three-dimensional volumetric OCT imaging of the unpigmented Sprague-Dawley rat retina and
pigmented C57BL/6 mouse retina is demonstrated in Fig. 2
and Fig. 3
, respectively. The animals were anesthetized so that measurement time is not limited by
blinking or motion, as is the case for human ophthalmic imaging. A coverslip was placed on the
cornea to remove the refractive power of the air-corneal interface, focusing and scanning the OCT
beam directly on the retina through the weaker refraction from the lens. Six orthogonally scanned
700 × 700 axial scan data sets were registered and merged.
Fig. 2
Imaging of the unpigmented Sprague-Dawley rat retina. Registered and merged OCT data set
generated from 6 orthogonally scanned OCT data sets. (700 × 700 axial scans over a 2.4 mm
× 2.4 mm region) (A) OCT fundus view. (B) Retinal layers visualized in the cropped, enlarged
OCT image. (C) 3-D rendering. (D, E, F) OCT images in the X direction. (G, H, I) OCT images in the Y
direction.
Fig. 3
Imaging of the pigmented C57BL/6 mouse retina. Registered and merged data set generated from 6
orthogonally scanned OCT data sets. (700 × 700 axial scans over a 1.2 mm × 1.2 mm
region) (A) OCT fundus view. (B) Retinal layers shown in the cropped, enlarged OCT image (C) 3-D
rendering. (D, E. F) OCT images in the X direction. (G, H, I) OCT images in the Y direction.
Imaging of the unpigmented Sprague-Dawley rat retina. Registered and merged OCT data set
generated from 6 orthogonally scanned OCT data sets. (700 × 700 axial scans over a 2.4 mm
× 2.4 mm region) (A) OCT fundus view. (B) Retinal layers visualized in the cropped, enlarged
OCT image. (C) 3-D rendering. (D, E, F) OCT images in the X direction. (G, H, I) OCT images in the Y
direction.Imaging of the pigmented C57BL/6 mouse retina. Registered and merged data set generated from 6
orthogonally scanned OCT data sets. (700 × 700 axial scans over a 1.2 mm × 1.2 mm
region) (A) OCT fundus view. (B) Retinal layers shown in the cropped, enlarged OCT image (C) 3-D
rendering. (D, E. F) OCT images in the X direction. (G, H, I) OCT images in the Y direction.Scans were performed over a 1.2 × 1.2 mm2 area of the mouse eye and a 2.4
× 2.4 mm2 area of the rat eye. Each data set is acquired in ~5 seconds. An OCT
fundus view is generated by axially summing the merged OCT data set (Fig. 2A and Fig. 3A). Cross-sectional images from the
merged data set enable visualization of major retinal layers including the inner plexiform layer
(IPL), inner nuclear layer (INL), outer plexiform layer (OPL), outer nuclear layer (ONL),
photoreceptor inner segment and outer segment (IS/OS) junction, retinal pigment epithelium (RPE),
choroid (CH), and sclera (SC) in both the Sprague-Dawley rat and C57BL/6 mouse eyes (Fig. 2B and Fig. 3B).In addition, the retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), and external
limiting membrane (ELM) are visible in the Sprague-Dawley rat (Fig.
2B). The smaller size of the mouse retina limited the visibility of smaller features.
Isotropic transverse sampling of the retina allows for volumetric rendering that shows the
structural details of the retina in 3-D (Fig. 2C and Fig. 3C). Since the 3-D OCT volumetric data set is motion
corrected, we can extract cross-sectional images in any position and direction without motion
artifacts (Figs. 2D–2I and Figs. 3D–3I). Notice that neither cross sectional images visualized in the horizontal and
vertical direction nor the 3-D volumetric rendering exhibit motion artifacts.
3.2. Anterior eye and full eye length imaging in rat and mouse eyes
The long imaging range of SS-OCT also enabled 3-D anterior eye and full eye length OCT imaging in
the rat eye and mouse eye using sample arm optics with a larger transverse spot size and longer
depth of field. Anterior eye and full eye length imaging provides 3-D information on the anterior
chamber and crystalline lens structure as well as biometric information on the full eye length. The
animal eye is placed directly in front of the scan lens under anesthesia and after dilation. Two
orthogonally scanned 500 × 500 axial scan data sets were registered and merged.The scans were performed over a 2.6 × 2.6 mm2 area of the mouse eye and a 7
× 7 mm2 area of the rat eye. Figure 4A
is an OCT en face view of a Sprague-Dawley rat eye. The Y-shaped shadow in the center of the
eye is the lens suture pattern on the anterior crystalline lens which can be visualized when an en
face cross section at appropriate depth is extracted from the volume (Fig. 4D). Cross sectional images in Fig. 4B
and Fig. 4C reveal detailed structures of the cornea, iris,
and lens, as well as visible signal from the vitreous and retina. A thin liquid film from the eye
drops is visible on the cornea of the rat eye. In addition to lens sutures, the nucleus is visible
in the cross sectional images.
Fig. 4
Anterior eye and full eye length OCT imaging in the Sprague-Dawley rat and C57BL/6 mouse eye. (A,
E) En face view of the volumetric data set registered and merged from two orthogonally scanned 500
× 500 axial scan volumes. The rat eye was scanned over a 7 mm × 7 mm area. The mouse
eye was scanned over a 2.6 mm × 2.6 mm area. (B, C, F, G) Cross-sectional images in the
horizontal and vertical direction generated from averaging neighboring 5 frames. The red arrows
point to the rat lens nucleus. The green box is a region in the image where the contrast was
adjusted to better visualize signal from the rat retina. (D) Extracted en face cross section at the
red line position showing the Y-shaped suture pattern in the rat lens anterior. (H) Summed en face
view over the rat lens posterior as indicated in the red dotted box, which shows the pattern of
opacities in the mouse lens.
Anterior eye and full eye length OCT imaging in the Sprague-Dawley rat and C57BL/6 mouse eye. (A,
E) En face view of the volumetric data set registered and merged from two orthogonally scanned 500
× 500 axial scan volumes. The rat eye was scanned over a 7 mm × 7 mm area. The mouse
eye was scanned over a 2.6 mm × 2.6 mm area. (B, C, F, G) Cross-sectional images in the
horizontal and vertical direction generated from averaging neighboring 5 frames. The red arrows
point to the rat lens nucleus. The green box is a region in the image where the contrast was
adjusted to better visualize signal from the rat retina. (D) Extracted en face cross section at the
red line position showing the Y-shaped suture pattern in the rat lens anterior. (H) Summed en face
view over the rat lens posterior as indicated in the red dotted box, which shows the pattern of
opacities in the mouse lens.Figure 4E is an OCT en face view of a C57BL/6 mouse eye.
This particular mouse eye exhibits lens opacities which can be visualized by summing en face pixels
near the posterior lens surface, as shown in Fig. 4H. Cross
sectional images in Fig. 4F and Fig. 4G reveal detailed structures of the cornea, iris, lens, vitreous and retina.
The 3-D data set illustrates the larger lens volume compared to vitreous volume in the rat and mouse
eyes. The irises and pupils are dilated. Ocular biometry measurements of the eyes can be performed
after refraction correcting the OCT images in the axial direction.
3.3. 3-D posterior eye imaging in unpigmented rat eye
In order to better image the posterior eye of the unpigmented Sprague-Dawley rat, the focus was
advanced towards the posterior eye and a coverslip was placed on the eye to minimize corneal
refraction. Although the larger transverse spot size decreases the transverse resolution compared to
the retinal imaging mode described in section 3.1, the longer focal depth and long imaging range
enables visualization of the posterior lens, vitreous and retina. Figure 5
is a posterior eye data set generated from registering and merging six orthogonally scanned
volumes of 500 × 500 A-scans each. The scans were performed over a 2.6 × 2.6
mm2 area of the rat eye. The data set shows the posterior lens surface, where the
Y-shaped lens sutures can be visualized in an inverted orientation as seen on the en face image in
Fig. 5B. The hyaloid vessel as well as floaters can be seen
in the vitreous. Retinal layers can be clearly visualized in cross-sectional images (Figs. 5C–5D). The
deep penetration of 1050 nm wavelength light allows visualization of tissue structure in the choroid
and sclera. The fundus image shows that major retinal blood vessels radiate from the center of the
optic disk.
Fig. 5
Posterior eye imaging of the unpigmented Sprague-Dawley rat including retina, vitreous, and
posterior lens. (A) OCT en face view of the registered and merged data set from six orthogonally
scanned 500 × 500 axial scan volumes. Imaging of the rat posterior eye was performed over a
2.6 mm × 2.6 mm area. (B) En face cross section showing inverted Y-shaped posterior lens
sutures. (C, D) Cross-sections from the registered data set showing the retina, hyaloid vessel and
posterior part of the crystalline lens. The red line indicates the depth position of the en face
cross section. (E) 3-D rendering of the data set (Media
1).
Posterior eye imaging of the unpigmented Sprague-Dawley rat including retina, vitreous, and
posterior lens. (A) OCT en face view of the registered and merged data set from six orthogonally
scanned 500 × 500 axial scan volumes. Imaging of the rat posterior eye was performed over a
2.6 mm × 2.6 mm area. (B) En face cross section showing inverted Y-shaped posterior lens
sutures. (C, D) Cross-sections from the registered data set showing the retina, hyaloid vessel and
posterior part of the crystalline lens. The red line indicates the depth position of the en face
cross section. (E) 3-D rendering of the data set (Media
1).
3.4. 3-D Doppler imaging in unpigmented rat eye
Doppler OCT imaging of the Sprague-Dawley rat was performed using the phase difference of
neighboring axial scans in the six orthogonally scanned 500 x 500 axial scan data sets acquired in
the posterior eye imaging mode described in section 3.3. The larger spot size decreases the number
of samples required to obtain Doppler OCT information over the same scan area compared to the
retinal imaging mode in section 3.1. The Doppler OCT data sets were merged using motion correction
information (displacement fields) from the registered intensity images (Fig. 6
). Intensity and Doppler OCT cross-sections present complementary information on the structure
and function of the retina. When the Doppler image is overlaid on its structural counterpart (Figs. 6B–6D), the
combined images may provide insight into the relationship between structural and functional changes
of the retina in pathology and disease progression. A 3-D rendering of the vasculature in the retina
and choroid is shown in Fig. 6E. Two types of vessels can be
distinguished in the retina by looking at the direction of blood flow. When blood moves towards the
OCT beam, the Doppler shift is positive, as indicated with the red (warm) color. The retinal
arteries are oriented in this direction and are hence visualized in red. On the other hand, blood
returning from retinal tissue produces negative Doppler shifts. Consequently, retinal veins are
visualized in a blue (cold) color. In the rat retina, arterial and venous vascular systems can
overlap one another. Since Doppler OCT can only measure flow velocities in the axial direction, the
vessels appear to be discontinuous or disconnected when they are perpendicular to the OCT beam. In
addition to retinal vessels and the central retinal artery, some choroidal vessels, as well as long
posterior ciliary arteries are also visible owing to the deep penetration of 1050 nm wavelength
light.
Fig. 6
Doppler OCT imaging in a unpigmented Sprague-Dawley rat retina. Doppler OCT analysis was
performed using posterior eye OCT data (6 orthogonally scanned 500 × 500 axial scan OCT data
sets over a 2.6 mm × 2.6 mm region). The 6 Doppler OCT volumes were merged using the
displacement fields from registered structural OCT data. (A) OCT fundus image. (B, C, D) OCT color
Doppler images with blood flow information overlaid on structural images. (E) 3-D Doppler OCT
angiography (Media 2).
Doppler OCT imaging in a unpigmented Sprague-Dawley rat retina. Doppler OCT analysis was
performed using posterior eye OCT data (6 orthogonally scanned 500 × 500 axial scan OCT data
sets over a 2.6 mm × 2.6 mm region). The 6 Doppler OCT volumes were merged using the
displacement fields from registered structural OCT data. (A) OCT fundus image. (B, C, D) OCT color
Doppler images with blood flow information overlaid on structural images. (E) 3-D Doppler OCT
angiography (Media 2).
3.5. 4-D imaging of pupillary response in rat and mouse eyes
High speed OCT imaging enables time resolved volumetric 4-D imaging of the dynamic responses of
the eye to stimulus. A Sprague-Dawley rat and a C57BL/6 mouse were sedated lightly to preserve
pupillary responses. Figures 7
and 8
are demonstrations of 4-D imaging of the pupillary response in the rat and mouse eye. OCT
volumes of 100 × 100 axial scans were acquired at ~10 volumes per second for 5 seconds. Scans
were performed over a 7 × 7 mm2 area of the rat eye and a 3.5 × 3.5
mm2 area of the mouse eye. All volumes were scanned with a horizontal fast scan axis and
motion correction was not performed. A continuous stimulus and a short ~1 second flash stimulus were
applied to both the rat and mouse eyes. 3-D dynamics can be visualized and quantitative measurements
of the pupil area can be obtained. Despite the high volume rate of the acquisition, motion is
visible in the data sets predominantly due to breathing and heartbeat. As more light enters the eye,
the iris responds by rapid contraction, which leads to the decrease in pupil diameter. When the
stimulus is turned off, the iris muscles slowly relax. The pupil area from each acquired volume was
measured.
Fig. 7
Dynamic response of the Sprague-Dawley rat eye to long (Media
3) and short duration (Media
4) light stimulus. Volumetric imaging was performed at ~10 volumes
per second. Each volume consists of 100 × 100 axial scans over a 7 mm × 7 mm area.
Changes of the pupil area before, during, and after light stimulus are shown in plot. (A, B, C) (D,
E, F) Selected time points illustrate the structural changes in the eye over time. An OCT en face
view, central orthogonal cross-sections, as well as a 3-D rendering are shown for each time
point.
Fig. 8
Dynamic response of the C57BL/6 mouse eye to long (Media
5) and short duration (Media
6) light stimulus. Volumetric imaging was performed at ~10 volumes
per second. Each volume consists of 100 × 100 axial scans over a 3.5 mm × 3.5 mm area.
Changes of the pupil area before, during, and after light stimulus are shown in plot. (A, B, C) (D,
E, F) Selected time points illustrate the structural changes in the eye over time. An OCT en face
view, central orthogonal cross-sections, as well as a 3-D rendering are shown for each time
point.
Dynamic response of the Sprague-Dawley rat eye to long (Media
3) and short duration (Media
4) light stimulus. Volumetric imaging was performed at ~10 volumes
per second. Each volume consists of 100 × 100 axial scans over a 7 mm × 7 mm area.
Changes of the pupil area before, during, and after light stimulus are shown in plot. (A, B, C) (D,
E, F) Selected time points illustrate the structural changes in the eye over time. An OCT en face
view, central orthogonal cross-sections, as well as a 3-D rendering are shown for each time
point.Dynamic response of the C57BL/6 mouse eye to long (Media
5) and short duration (Media
6) light stimulus. Volumetric imaging was performed at ~10 volumes
per second. Each volume consists of 100 × 100 axial scans over a 3.5 mm × 3.5 mm area.
Changes of the pupil area before, during, and after light stimulus are shown in plot. (A, B, C) (D,
E, F) Selected time points illustrate the structural changes in the eye over time. An OCT en face
view, central orthogonal cross-sections, as well as a 3-D rendering are shown for each time
point.
4. Discussion and conclusion
Structural and Doppler angiographic imaging of the rodent eye was demonstrated using a novel
ultrahigh speed swept source/Fourier domain OCT instrument. The swept source OCT was based on a
recently developed instrument for high speed ophthalmic OCT imaging at long wavelengths [17]. The long imaging range and high imaging speed of SS-OCT, along
with the deep penetration into tissue of 1050 nm wavelength light enables 3-D volumetric imaging of
the retina and posterior eye, as well as the anterior eye and full eye length. A registration
algorithm was applied to remove motion artifacts and merge multiple data sets for enhanced
visualization.OCT has been widely used for in vivo retinal imaging of rodent models. This manuscript presents
the first motion corrected 3-D OCT images in rat and mouse eyes using SS-OCT. Although current 1050
nm wavelength SS-OCT technology has limited resolution compared to 850 nm SD-OCT, the improved
tissue penetration provides additional information in the choroid and sclera. Motion correction
enables large volumes to be acquired without motion artifacts and merging improves image quality.
Although visibility of features is poorer in the mouse compared with the rat, for retinal imaging,
major retinal layers including the retinal nerve fiber layer (RNFL), ganglion cell layer (GCL),
inner plexiform layer (IPL), inner nuclear layer (INL), outer plexiform layer (OPL), outer nuclear
layer (ONL), external limiting membrane (ELM), photoreceptor inner segment and outer segment (IS/OS)
junction, retinal pigment epithelium (RPE), choroid (CH), and sclera (SC) can be resolved. Doppler
OCT provides motion contrast highlighting the retinal vasculature and enables differentiating
arteries and veins. Quantitative information of blood flow velocity can also be obtained. No obvious
artifacts due to pulsatility remain in the motion corrected and merged Doppler OCT data which was
obtained from multiple orthogonally scanned volumes. The deep penetration into tissue allowed
Doppler OCT imaging of the long posterior ciliary arteries.The main advantages of SS-OCT compared to SD-OCT are increased sensitivities, imaging speed, and
imaging range. For rodent eye imaging, the high speeds and extended range enable 3-D imaging of the
retina, posterior eye, anterior eye and full eye length. Combined with image registration motion
correction, we demonstrated 3-D in vivo SS-OCT full eye length volumetric imaging while preserving
corneal, iris, lens, and retinal topography in both rats and mice. This promises to allow
quantitative measurements tracking the changes in morphogenesis and pathological processes in the
same animal eye over time.SS-OCT at 1050 nm may provide new means for in vivo rodent hyaloid vessel and lens development
studies. In our results, we were able to visualize the hyaloid vessel, lens nucleus, lens opacities,
and lens suture patterns. Comparing Fig. 4D with Fig. 5B, it can be seen that suture patterns in the anterior and
posterior pole of the crystalline lens have opposite orientations. Lens fibers make up the bulk of
the lens and form lens suture patterns which produce an upright 'Y' pattern anteriorly and
an inverted 'Y' pattern posteriorly. This is the first time that the lens suture pattern
in the rat eye has been visualized using OCT. In the mouse eye, we could observe lens opacities
possibly induced by ketamine-xylazine anesthesia (Fig. 4E)
suggesting that cataract characterization may be possible [27]. Meanwhile in the mouse full eye length image, the retina remains highly visible despite
the lens opacities. This also suggests that the reduced scattering of 1050 nm wavelength light will
be useful for OCT retinal imaging even in rodent eyes with cataracts.For mouse eye imaging and biometry measurements, Wang et al. [18] has described a 28 kHz SS-OCT imaging system with a center wavelength of 1056 nm and 70
nm scanning bandwidth. A subsequent paper demonstrated the ability to measure the heritability of
ocular component dimensions in mice. These papers reported the ability to acquire ~17.6 µm
axial resolution cross-sectional data in air using a fast Dispersion Encoded Full Range (DEFR)
algorithm over a depth range of ±5 mm in air with −18-dB roll-off at ±3 mm in
air where conjugate artifacts compromise the alignment of the imaging system. By taking advantage of
the longer imaging range with −20-dB roll-off at 5.3 mm in tissue, the higher 6 µm
axial resolution in tissue, the 3.6 times faster 100 kHz imaging speed, and registration-based
motion correction algorithm, our results provided motion-corrected 3-D OCT data of the full eye
length in not only the shorter mouse eye, but also the longer rat eye. Three-dimensional motion
corrected data can be refraction corrected and processed for more accurate biometry measurements
unaffected by alignment errors or motion artifacts when acquiring only a single cross-sectional OCT
image. Although the limited detector bandwidth along with the limited coherence length of the swept
source laser used in this paper affected signal in the rat retina during full eye length imaging
(Fig. 4B
Fig. 4C), next generation swept lasers with ultralong
coherence lengths will have virtually no signal roll-off within the range needed for rodent full eye
length imaging [28].We also performed the first demonstration of 4-D time resolved volumetric OCT imaging of the
mouse and rat full eye length. 4-D OCT dynamic imaging could provide spatial-temporal information of
3-D volumetric datasets for functional imaging studies. A simple demonstration of pupillometry is
shown in this paper. In addition to pupil area, 4-D OCT provides information of the iris structure.
The contraction of the iris was visualized in 3-D. Other 4-D OCT applications include imaging of
structural and blood flow responses to intraocular pressure [29,30] and the dynamics of neurovascular function in
the eye [31,32].To conclude, SS-OCT is a powerful imaging technique providing comprehensive 3-D information on
the rodent eye including motion corrected posterior eye, anterior eye and full eye length imaging,
retinal imaging with Doppler OCT angiography, as well as 4-D dynamic imaging of functional responses
of the eye. While standard excisional biopsy and histology require enucleating the eye, OCT can
perform repeated, noninvasive in situ imaging and quantitative measurements of the rodent retina.
Therefore, SS-OCT technology for rodent eye imaging is a potentially useful tool for in vivo imaging
of disease phenotypes such as corneal opacity, uveitis, keratitis, glaucoma, cataract,
retinoblastoma, retinal degeneration, retinal vascular disease and myopia.
Authors: Mathias W Seeliger; Susanne C Beck; Naira Pereyra-Muñoz; Susann Dangel; Jen-Yue Tsai; Ulrich F O Luhmann; Serge A van de Pavert; Jan Wijnholds; Marijana Samardzija; Andreas Wenzel; Eberhart Zrenner; Kristina Narfström; Edda Fahl; Naoyuki Tanimoto; Niyazi Acar; Felix Tonagel Journal: Vision Res Date: 2005-09-26 Impact factor: 1.886
Authors: Marco Ruggeri; Hassan Wehbe; Shuliang Jiao; Giovanni Gregori; Maria E Jockovich; Abigail Hackam; Yuanli Duan; Carmen A Puliafito Journal: Invest Ophthalmol Vis Sci Date: 2007-04 Impact factor: 4.799
Authors: Haiying Cheng; Govind Nair; Tiffany A Walker; Moon K Kim; Machelle T Pardue; Peter M Thulé; Darin E Olson; Timothy Q Duong Journal: Proc Natl Acad Sci U S A Date: 2006-11-06 Impact factor: 11.205
Authors: Vivek J Srinivasan; Tony H Ko; Maciej Wojtkowski; Mariana Carvalho; Allen Clermont; Sven-Erik Bursell; Qin Hui Song; Janis Lem; Jay S Duker; Joel S Schuman; James G Fujimoto Journal: Invest Ophthalmol Vis Sci Date: 2006-12 Impact factor: 4.799
Authors: Zhongwei Zhi; Xin Yin; Suzan Dziennis; Tomasz Wietecha; Kelly L Hudkins; Charles E Alpers; Ruikang K Wang Journal: Biomed Opt Express Date: 2012-10-24 Impact factor: 3.732
Authors: Raju Poddar; Poddar Raju; Dennis E Cortés; John S Werner; Mark J Mannis; Robert J Zawadzki Journal: J Biomed Opt Date: 2013-08 Impact factor: 3.170