Total hip arthroplasty with the MS-30 polished surface cemented stem: a single surgeon consecutive series study at 10 year follow-up.

Gavin H O Brigstocke1, Philip A Mitchell, John W Rosson.   

Abstract

The Morscher-Sportorno femoral stem is a stainless steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. The clinical and radiographic results of this implant at a minimum of 5-year follow-up have been extremely encouraging. However, the results at 10 years have yet to be reported. Therefore, we wished to establish the clinical performance and survival of the polished surface MS-30 femoral stem at a minimum of 10-year follow-up. We report the results at a minimum follow-up of 10 years of a single surgeon series of 55 total hip replacements with the polished surface MS-30 cemented femoral component in 48 consecutive patients aged between 50 and 64. The median period of observation was 12 years (inter-quartile range, 10-13). Six patients died with their implant in situ, and one patient was unable to attend for follow-up. All the remaining 41 patients with 47 total hip replacements were reviewed. All patients were assessed clinically with the use of the Oxford hip score, and radiographs taken at the latest follow-up were evaluated by an independent specialist hip surgeon. To date, none of the implants have been revised. The median Oxford hip score was 13 (inter-quartile range, 12-14). Radiologically, no stems had evidence of loosening on assessment by the Harris criteria, and only 4 hips (8.5 %) had osteolysis adjacent to the stem, mostly in Gruen zone 7. The results of the polished surface MS-30 femoral stem were satisfactory with survivorship at a minimum of 10 years of 100 %. These findings support the continued use of this femoral component.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23412257     DOI: 10.1007/s00590-012-1132-7

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  10 in total

1.  The Exeter universal cemented femoral component at 8 to 12 years. A study of the first 325 hips.

Authors:  H D W Williams; G Browne; G A Gie; R S M Ling; A J Timperley; N A Wendover
Journal:  J Bone Joint Surg Br       Date:  2002-04

2.  The long-term results of the original Exeter polished cemented femoral component: a follow-up report.

Authors:  Robin S M Ling; John Charity; A J Clive Lee; Sarah L Whitehouse; A John Timperley; Graham A Gie
Journal:  J Arthroplasty       Date:  2009-03-17       Impact factor: 4.757

3.  Loosening of matt and polished cemented femoral stems.

Authors:  D W Howie; R G Middleton; K Costi
Journal:  J Bone Joint Surg Br       Date:  1998-07

4.  Effects of design changes on cemented tapered femoral stem fixation.

Authors:  R G Middleton; D W Howie; K Costi; P Sharpe
Journal:  Clin Orthop Relat Res       Date:  1998-10       Impact factor: 4.176

5.  Questionnaire on the perceptions of patients about total hip replacement.

Authors:  J Dawson; R Fitzpatrick; A Carr; D Murray
Journal:  J Bone Joint Surg Br       Date:  1996-03

6.  Prognosis of total hip replacement in Sweden. Follow-up of 92,675 operations performed 1978-1990.

Authors:  H Malchau; P Herberts; L Ahnfelt
Journal:  Acta Orthop Scand       Date:  1993-10

7.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

8.  Ten-year survival of the MS-30 matt-surfaced cemented stem.

Authors:  B J Berli; D Schäfer; E W Morscher
Journal:  J Bone Joint Surg Br       Date:  2005-07

9.  The Exeter Universal stem: a minimum ten-year review from an independent centre.

Authors:  S Hook; E Moulder; P J Yates; B J Burston; E Whitley; G C Bannister
Journal:  J Bone Joint Surg Br       Date:  2006-12

10.  Femoral component loosening using contemporary techniques of femoral cement fixation.

Authors:  W H Harris; J C McCarthy; D A O'Neill
Journal:  J Bone Joint Surg Am       Date:  1982-09       Impact factor: 5.284

  10 in total
  1 in total

1.  Implant survival of the most common cemented total hip devices from the Nordic Arthroplasty Register Association database.

Authors:  Mika Junnila; Inari Laaksonen; Antti Eskelinen; Pekka Pulkkinen; Leif Ivar Havelin; Ove Furnes; Anne Marie Fenstad; Alma B Pedersen; Søren Overgaard; Johan Kärrholm; Göran Garellick; Henrik Malchau; Keijo T Mäkelä
Journal:  Acta Orthop       Date:  2016-08-23       Impact factor: 3.717

  1 in total