Extra-dimensional Demons: a method for incorporating missing tissue in deformable image registration.

Sajendra Nithiananthan1, Sebastian Schafer, Daniel J Mirota, J Webster Stayman, Wojciech Zbijewski, Douglas D Reh, Gary L Gallia, Jeffrey H Siewerdsen.   

Abstract

PURPOSE: A deformable registration method capable of accounting for missing tissue (e.g., excision) is reported for application in cone-beam CT (CBCT)-guided surgical procedures. Excisions are identified by a segmentation step performed simultaneous to the registration process. Tissue excision is explicitly modeled by increasing the dimensionality of the deformation field to allow motion beyond the dimensionality of the image. The accuracy of the model is tested in phantom, simulations, and cadaver models.
METHODS: A variant of the Demons deformable registration algorithm is modified to include excision segmentation and modeling. Segmentation is performed iteratively during the registration process, with initial implementation using a threshold-based approach to identify voxels corresponding to "tissue" in the moving image and "air" in the fixed image. With each iteration of the Demons process, every voxel is assigned a probability of excision. Excisions are modeled explicitly during registration by increasing the dimensionality of the deformation field so that both deformations and excisions can be accounted for by in- and out-of-volume deformations, respectively. The out-of-volume (i.e., fourth) component of the deformation field at each voxel carries a magnitude proportional to the excision probability computed in the excision segmentation step. The registration accuracy of the proposed "extra-dimensional" Demons (XDD) and conventional Demons methods was tested in the presence of missing tissue in phantom models, simulations investigating the effect of excision size on registration accuracy, and cadaver studies emulating realistic deformations and tissue excisions imparted in CBCT-guided endoscopic skull base surgery.
RESULTS: Phantom experiments showed the normalized mutual information (NMI) in regions local to the excision to improve from 1.10 for the conventional Demons approach to 1.16 for XDD, and qualitative examination of the resulting images revealed major differences: the conventional Demons approach imparted unrealistic distortions in areas around tissue excision, whereas XDD provided accurate "ejection" of voxels within the excision site and maintained the registration accuracy throughout the rest of the image. Registration accuracy in areas far from the excision site (e.g., > ∼5 mm) was identical for the two approaches. Quantitation of the effect was consistent in analysis of NMI, normalized cross-correlation (NCC), target registration error (TRE), and accuracy of voxels ejected from the volume (true-positive and false-positive analysis). The registration accuracy for conventional Demons was found to degrade steeply as a function of excision size, whereas XDD was robust in this regard. Cadaver studies involving realistic excision of the clivus, vidian canal, and ethmoid sinuses demonstrated similar results, with unrealistic distortion of anatomy imparted by conventional Demons and accurate ejection and deformation for XDD.
CONCLUSIONS: Adaptation of the Demons deformable registration process to include segmentation (i.e., identification of excised tissue) and an extra dimension in the deformation field provided a means to accurately accommodate missing tissue between image acquisitions. The extra-dimensional approach yielded accurate "ejection" of voxels local to the excision site while preserving the registration accuracy (typically subvoxel) of the conventional Demons approach throughout the rest of the image. The ability to accommodate missing tissue volumes is important to application of CBCT for surgical guidance (e.g., skull base drillout) and may have application in other areas of CBCT guidance.

Entities:  

Mesh:

Year:  2012        PMID: 22957637      PMCID: PMC3443194          DOI: 10.1118/1.4747270

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Cone-beam computed tomography with a flat-panel imager: initial performance characterization.

Authors:  D A Jaffray; J H Siewerdsen
Journal:  Med Phys       Date:  2000-06       Impact factor: 4.071

2.  Medical image registration with partial data.

Authors:  Senthil Periaswamy; Hany Farid
Journal:  Med Image Anal       Date:  2005-06-23       Impact factor: 8.545

3.  Towards cardiac C-arm computed tomography.

Authors:  Günter Lauritsch; Jan Boese; Lars Wigström; Herbert Kemeth; Rebecca Fahrig
Journal:  IEEE Trans Med Imaging       Date:  2006-07       Impact factor: 10.048

Review 4.  Advances in image-guided radiation therapy.

Authors:  Laura A Dawson; David A Jaffray
Journal:  J Clin Oncol       Date:  2007-03-10       Impact factor: 44.544

5.  Image quality and localization accuracy in C-arm tomosynthesis-guided head and neck surgery.

Authors:  G Bachar; J H Siewerdsen; M J Daly; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

6.  Cone-beam computed tomography on a mobile C-arm: novel intraoperative imaging technology for guidance of head and neck surgery.

Authors:  Yvonne Chan; Jeffrey H Siewerdsen; Mark A Rafferty; Douglas J Moseley; David A Jaffray; Jonathan C Irish
Journal:  J Otolaryngol Head Neck Surg       Date:  2008-02

7.  Image matching as a diffusion process: an analogy with Maxwell's demons.

Authors:  J P Thirion
Journal:  Med Image Anal       Date:  1998-09       Impact factor: 8.545

8.  Validation of an accelerated 'demons' algorithm for deformable image registration in radiation therapy.

Authors:  He Wang; Lei Dong; Jennifer O'Daniel; Radhe Mohan; Adam S Garden; K Kian Ang; Deborah A Kuban; Mark Bonnen; Joe Y Chang; Rex Cheung
Journal:  Phys Med Biol       Date:  2005-06-01       Impact factor: 3.609

9.  Evaluation of portable CT scanners for otologic image-guided surgery.

Authors:  Ramya Balachandran; Daniel Schurzig; J Michael Fitzpatrick; Robert F Labadie
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-07-21       Impact factor: 2.924

10.  Intraoperative cone-beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C-arm prototype.

Authors:  M J Daly; J H Siewerdsen; D J Moseley; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

View more
  14 in total

1.  Deformable image registration for cone-beam CT guided transoral robotic base-of-tongue surgery.

Authors:  S Reaungamornrat; W P Liu; A S Wang; Y Otake; S Nithiananthan; A Uneri; S Schafer; E Tryggestad; J Richmon; J M Sorger; J H Siewerdsen; R H Taylor
Journal:  Phys Med Biol       Date:  2013-06-27       Impact factor: 3.609

2.  The distance discordance metric-a novel approach to quantifying spatial uncertainties in intra- and inter-patient deformable image registration.

Authors:  Ziad H Saleh; Aditya P Apte; Gregory C Sharp; Nadezhda P Shusharina; Ya Wang; Harini Veeraraghavan; Maria Thor; Ludvig P Muren; Shyam S Rao; Nancy Y Lee; Joseph O Deasy
Journal:  Phys Med Biol       Date:  2014-01-20       Impact factor: 3.609

3.  Imaging Study of Pseudo-CT Synthesized From Cone-Beam CT Based on 3D CycleGAN in Radiotherapy.

Authors:  Hongfei Sun; Rongbo Fan; Chunying Li; Zhengda Lu; Kai Xie; Xinye Ni; Jianhua Yang
Journal:  Front Oncol       Date:  2021-03-12       Impact factor: 6.244

4.  dPIRPLE: a joint estimation framework for deformable registration and penalized-likelihood CT image reconstruction using prior images.

Authors:  H Dang; A S Wang; Marc S Sussman; J H Siewerdsen; J W Stayman
Journal:  Phys Med Biol       Date:  2014-08-06       Impact factor: 3.609

5.  Temporal subtraction contrast-enhanced dedicated breast CT.

Authors:  Peymon M Gazi; Shadi Aminololama-Shakeri; Kai Yang; John M Boone
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

6.  MIND Demons for MR-to-CT Deformable Image Registration In Image-Guided Spine Surgery.

Authors:  S Reaungamornrat; T De Silva; A Uneri; J-P Wolinsky; A J Khanna; G Kleinszig; S Vogt; J L Prince; J H Siewerdsen
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-18

7.  MIND Demons: Symmetric Diffeomorphic Deformable Registration of MR and CT for Image-Guided Spine Surgery.

Authors:  Sureerat Reaungamornrat; Tharindu De Silva; Ali Uneri; Sebastian Vogt; Gerhard Kleinszig; Akhil J Khanna; Jean-Paul Wolinsky; Jerry L Prince; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2016-06-02       Impact factor: 10.048

8.  Research on obtaining pseudo CT images based on stacked generative adversarial network.

Authors:  Hongfei Sun; Zhengda Lu; Rongbo Fan; Wenjun Xiong; Kai Xie; Xinye Ni; Jianhua Yang
Journal:  Quant Imaging Med Surg       Date:  2021-05

9.  Integration of free-hand 3D ultrasound and mobile C-arm cone-beam CT: Feasibility and characterization for real-time guidance of needle insertion.

Authors:  E Marinetto; A Uneri; T De Silva; S Reaungamornrat; W Zbijewski; A Sisniega; S Vogt; G Kleinszig; J Pascau; J H Siewerdsen
Journal:  Comput Med Imaging Graph       Date:  2017-04-03       Impact factor: 7.422

10.  Semi-Supervised Deep Learning-Based Image Registration Method with Volume Penalty for Real-Time Breast Tumor Bed Localization.

Authors:  Marek Wodzinski; Izabela Ciepiela; Tomasz Kuszewski; Piotr Kedzierawski; Andrzej Skalski
Journal:  Sensors (Basel)       Date:  2021-06-14       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.