Cone-beam computed tomography with a flat-panel imager: initial performance characterization.

D A Jaffray1, J H Siewerdsen.   

Abstract

The development and performance of a system for x-ray cone-beam computed tomography (CBCT) using an indirect-detection flat-panel imager (FPI) is presented. Developed as a bench-top prototype for initial investigation of FPI-based CBCT for bone and soft-tissue localization in radiotherapy, the system provides fully three-dimensional volumetric image data from projections acquired during a single rotation. The system employs a 512 x 512 active matrix of a-Si:H thin-film transistors and photodiodes in combination with a luminescent phosphor. Tomographic imaging performance is quantified in terms of response uniformity, response linearity, voxel noise, noise-power spectrum (NPS), and modulation transfer function (MTF), each in comparison to the performance measured on a conventional CT scanner. For the geometry employed and the objects considered, response is uniform to within 2% and linear within 1%. Voxel noise, at a level of approximately 20 HU, is comparable to the conventional CT scanner. NPS and MTF results highlight the frequency-dependent transfer characteristics, confirming that the CBCT system can provide high spatial resolution and does not suffer greatly from additive noise levels. For larger objects and/or low exposures, additive noise levels must be reduced to maintain high performance. Imaging studies of a low-contrast phantom and a small animal (a euthanized rat) qualitatively demonstrate excellent soft-tissue visibility and high spatial resolution. Image quality appears comparable or superior to that of the conventional scanner. These quantitative and qualitative results clearly demonstrate the potential of CBCT systems based upon flat-panel imagers. Advances in FPI technology (e.g., improved x-ray converters and enhanced electronics) are anticipated to allow high-performance FPI-based CBCT for medical imaging. General and specific requirements of kilovoltage CBCT systems are discussed, and the applicability of FPI-based CBCT systems to tomographic localization and image-guidance for radiotherapy is considered.

Entities:  

Mesh:

Year:  2000        PMID: 10902561     DOI: 10.1118/1.599009

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  119 in total

Review 1.  Imaging and image-guided radiation therapy in liver cancer.

Authors:  Kristy K Brock
Journal:  Semin Radiat Oncol       Date:  2011-10       Impact factor: 5.934

2.  Experimental flat-panel high-spatial-resolution volume CT of the temporal bone.

Authors:  Rajiv Gupta; Soenke H Bartling; Samit K Basu; William R Ross; Hartmut Becker; Armin Pfoh; Thomas Brady; Hugh D Curtin
Journal:  AJNR Am J Neuroradiol       Date:  2004-09       Impact factor: 3.825

3.  Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom.

Authors:  H-L Chao; W-L Chen; C-C Hu; J-K Wu; C-J Wu; J C-H Cheng
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

4.  Development, implementation and evaluation of a dedicated metal artefact reduction method for interventional flat-detector CT.

Authors:  D Prell; W A Kalender; Y Kyriakou
Journal:  Br J Radiol       Date:  2010-09-21       Impact factor: 3.039

5.  Assessing the intrinsic precision of 3D/3D rigid image registration results for patient setup in the absence of a ground truth.

Authors:  Jian Wu; Martin J Murphy
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

6.  Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT.

Authors:  Dimitrios Lazos; Jeffrey F Williamson
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

7.  Evaluating the image quality of cone beam CT acquired during rotational delivery.

Authors:  S A Yoganathan; K J Maria Das; K Maria Midunvaleja; D Gowtham Raj; Arpita Agarwal; J Velmurugan; Shaleen Kumar
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

8.  Investigation of Noise and Contrast Sensitivity of an Electron Multiplying Charge-Coupled Device (EMCCD) based Cone Beam Micro-CT System.

Authors:  Sumukh Bysani Krishnakumar; Alexander R Podgorsak; Sv Setlur Nagesh; Amit Jain; Stephen Rudin; Daniel R Bednarek; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-25

9.  Experimental realization of fluence field modulated CT using digital beam attenuation.

Authors:  T P Szczykutowicz; C A Mistretta
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

10.  High temporal resolution and streak-free four-dimensional cone-beam computed tomography.

Authors:  Shuai Leng; Jie Tang; Joseph Zambelli; Brian Nett; Ranjini Tolakanahalli; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2008-09-24       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.