Validating plastic scintillation detectors for photon dosimetry in the radiologic energy range.

Francois Lessard1, Louis Archambault, Mathieu Plamondon, Philippe Despres, Franccois Therriault-Proulx, Sam Beddar, Luc Beaulieu.   

Abstract

PURPOSE: Photon dosimetry in the kilovolt (kV) energy range represents a major challenge for diagnostic and interventional radiology and superficial therapy. Plastic scintillation detectors (PSDs) are potentially good candidates for this task. This study proposes a simple way to obtain accurate correction factors to compensate for the response of PSDs to photon energies between 80 and 150 kVp. The performance of PSDs is also investigated to determine their potential usefulness in the diagnostic energy range.
METHODS: A 1-mm-diameter, 10-mm-long PSD was irradiated by a Therapax SXT 150 unit using five different beam qualities made of tube potentials ranging from 80 to 150 kVp and filtration thickness ranging from 0.8 to 0.2 mmAl + 1.0 mmCu. The light emitted by the detector was collected using an 8-m-long optical fiber and a polychromatic photodiode, which converted the scintillation photons to an electrical current. The PSD response was compared with the reference free air dose rate measured with a calibrated Farmer NE2571 ionization chamber. PSD measurements were corrected using spectra-weighted corrections, accounting for mass energy-absorption coefficient differences between the sensitive volumes of the ionization chamber and the PSD, as suggested by large cavity theory (LCT). Beam spectra were obtained from x-ray simulation software and validated experimentally using a CdTe spectrometer. Correction factors were also obtained using Monte Carlo (MC) simulations. Percent depth dose (PDD) measurements were compensated for beam hardening using the LCT correction method. These PDD measurements were compared with uncorrected PSD data, PDD measurements obtained using Gafchromic films, Monte Carlo simulations, and previous data.
RESULTS: For each beam quality used, the authors observed an increase of the energy response with effective energy when no correction was applied to the PSD response. Using the LCT correction, the PSD response was almost energy independent, with a residual 2.1% coefficient of variation (COV) over the 80-150-kVp energy range. Monte Carlo corrections reduced the COV to 1.4% over this energy range. All PDD measurements were in good agreement with one another except for the uncorrected PSD data, in which an over-response was observed with depth (13% at 10 cm with a 100 kVp beam), showing that beam hardening had a non-negligible effect on the PSD response. A correction based on LCT compensated very well for this effect, reducing the over-response to 3%.
CONCLUSION: In the diagnostic energy range, PSDs show high-energy dependence, which can be corrected using spectra-weighted mass energy-absorption coefficients, showing no considerable sign of quenching between these energies. Correction factors obtained by Monte Carlo simulations confirm that the approximations made by LCT corrections are valid. Thus, PSDs could be useful for real-time dosimetry in radiology applications.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22957599      PMCID: PMC4108687          DOI: 10.1118/1.4738964

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  Pantak Therapax SXT 150: performance assessment and dose determination using IAEA TRS-398 protocol.

Authors:  D Jurado; T Eudaldo; P Carrasco; N Jornet; A Ruiz; M Ribas
Journal:  Br J Radiol       Date:  2005-08       Impact factor: 3.039

3.  Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry.

Authors:  Louis Archambault; A Sam Beddar; Luc Gingras; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

4.  Molybdenum target x-ray spectra: a semiempirical model.

Authors:  D M Tucker; G T Barnes; X Z Wu
Journal:  Med Phys       Date:  1991 May-Jun       Impact factor: 4.071

5.  Compton-scattering measurement of diagnostic x-ray spectrum using high-resolution Schottky CdTe detector.

Authors:  Koji Maeda; Masao Matsumoto; Akira Taniguchi
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

6.  Plastic scintillator response to low-energy photons.

Authors:  J F Williamson; J F Dempsey; A S Kirov; J I Monroe; W R Binns; H Hedtjärn
Journal:  Phys Med Biol       Date:  1999-04       Impact factor: 3.609

7.  Spectral discrimination of Cerenkov radiation in scintillating dosimeters.

Authors:  A M Frelin; J M Fontbonne; G Ban; J Colin; M Labalme; A Batalla; A Isambert; A Vela; T Leroux
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

8.  The use of cadmium telluride detectors for the qualitative analysis of diagnostic x-ray spectra.

Authors:  E Di Castro; R Pani; R Pellegrini; C Bacci
Journal:  Phys Med Biol       Date:  1984-09       Impact factor: 3.609

9.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

10.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

View more
  9 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Characterization of a scintillating fibre detector for small animal imaging and irradiation dosimetry.

Authors:  Coralie Le Deroff; Anne-Marie Frelin-Labalme; Xavier Ledoux
Journal:  Br J Radiol       Date:  2016-09-16       Impact factor: 3.039

3.  On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetry.

Authors:  François Therriault-Proulx; Luc Beaulieu; Louis Archambault; Sam Beddar
Journal:  Phys Med Biol       Date:  2013-03-08       Impact factor: 3.609

4.  Effects of Temperature and X-rays on Plastic Scintillating Fiber and Infrared Optical Fiber.

Authors:  Bongsoo Lee; Sang Hun Shin; Kyoung Won Jang; Wook Jae Yoo
Journal:  Sensors (Basel)       Date:  2015-05-11       Impact factor: 3.576

5.  Increased radiosensitivity of colorectal tumors with intra-tumoral injection of low dose of gold nanoparticles.

Authors:  Minghan Shi; Benoit Paquette; Thititip Thippayamontri; Louis Gendron; Brigitte Guérin; Léon Sanche
Journal:  Int J Nanomedicine       Date:  2016-10-12

6.  Characterization of an Innovative Detector Based on Scintillating Fiber for Personalized Computed Tomography Dosimetry.

Authors:  Clément Devic; Johann Plagnard; Mélodie Munier
Journal:  Sensors (Basel)       Date:  2021-12-23       Impact factor: 3.576

7.  Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector.

Authors:  Peter Georgi; Gustavo Kertzscher; Lars Nyvang; Jaroslav Šolc; Thorsten Schneider; Kari Tanderup; Jacob Graversen Johansen
Journal:  Med Phys       Date:  2022-03-03       Impact factor: 4.506

8.  Measurement of entrance surface dose on an anthropomorphic thorax phantom using a miniature fiber-optic dosimeter.

Authors:  Wook Jae Yoo; Sang Hun Shin; Dayeong Jeon; Seunghan Hong; Hyeok In Sim; Seon Geun Kim; Kyoung Won Jang; Seunghyun Cho; Won Sik Youn; Bongsoo Lee
Journal:  Sensors (Basel)       Date:  2014-04-01       Impact factor: 3.576

9.  Response to Re: Estimating and reducing dose received by cardiac devices for patients undergoing radiotherapy.

Authors:  Louis Archambault; Nicolas Varfalvy; Alexandra Bourgouin
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.